Ultrahigh-Performance capillary liquid chromatography-mass spectrometry at 35 kpsi for separation of lipids

被引:18
作者
Sorensen, Matthew J. [1 ]
Miller, Kelsey E. [2 ]
Jorgenson, James W. [2 ]
Kennedy, Robert T. [1 ,3 ]
机构
[1] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
[2] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
[3] Univ Michigan, Dept Pharmacol, Ann Arbor, MI 48109 USA
关键词
UHPLC; LC-MS; Capillary chromatography; Lipidomics; Lipid isomers; PEAK-CAPACITY; SLURRY CONCENTRATION; HIGH-EFFICIENCY; PRESSURE; LIPIDOMICS; COLUMNS; MS; CHALLENGES; SYSTEMS; MS/MS;
D O I
10.1016/j.chroma.2019.460575
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Improvements in sample preparation, separation, and mass spectrometry continue to expand the coverage in LC-MS based lipidomics. While longer columns packed with smaller particles in theory give higher separation performance compared to shorter columns, the implementation of this technology above commercial limits has been sparse due to difficulties in packing long columns and successfully operating instruments at ultrahigh pressures. In this work, a liquid chromatograph that operates up to 35 kpsi was investigated for the separation and identification of lipid species from human plasma. Capillary columns between 15-50 cm long were packed with 1.7 mu m BEH C18 particles and evaluated for their ability to separate lipid isomers and complex lipid extracts from human plasma. Putative lipid class identifications were assigned using accurate mass and relative retention time data of the eluting peaks. Our findings indicate that longer columns packed and operated at 35 kpsi outperform shorter columns packed and run at lower pressures in terms of peak capacity and numbers of features identified. Packing columns with relatively high concentration slurries (200 mg/mL) while sonicating the column resulted in 6-34% increase in peak capacity for 50 cm columns compared to lower slurry concentrations and no sonication. For a given analysis time, 50 cm long columns operated at 35 kpsi provided a 20-95% increase in chromatographic peak capacity compared with 15 cm columns operated at 15 kpsi. Analysis times up to 4 h were evaluated, generating peak capacities up to 410 +/- 5 (n=3, measured at 4 sigma) and identifying 480 +/- 85 lipids (n =2). Importantly, the results also show a correlation between the peak capacity and the number of lipids identified from a human plasma extract. This correlation indicates that ionization suppression is a limiting factor in obtaining sufficient signal for identification by mass spectrometry. The result also shows that the higher resolution obtained by shallow gradients overcomes possible signal reduction due to broader, more dilute peaks in long gradients for improving detection of lipids in LC-MS. Lastly, longer columns operated at shallow gradients allowed for the best separation of both regional and geometrical isomers. These results demonstrate a system that enables the advantages of using longer columns packed and run at ultrahigh pressure for improving lipid separations and lipidome coverage. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:8
相关论文
共 44 条
[1]  
[Anonymous], 2010, Introduction to Modern Liquid Chromatography
[2]   Increasing Compound Identification Rates in Untargeted Lipidomics Research with Liquid Chromatography Drift Time-Ion Mobility Mass Spectrometry [J].
Blazenovic, Ivana ;
Shen, Tong ;
Mehta, Sajjan S. ;
Kind, Tobias ;
Ji, Jian ;
Piparo, Marco ;
Cacciola, Francesco ;
Mondello, Luigi ;
Fiehn, Oliver .
ANALYTICAL CHEMISTRY, 2018, 90 (18) :10758-10764
[3]  
BLIGH EG, 1959, CAN J BIOCHEM PHYS, V37, P911
[4]   Recent advances in capillary ultrahigh pressure liquid chromatography [J].
Blue, Laura E. ;
Franklin, Edward G. ;
Godinho, Justin M. ;
Grinias, James P. ;
Grinias, Kaitlin M. ;
Lunn, Daniel B. ;
Moore, Stephanie M. .
JOURNAL OF CHROMATOGRAPHY A, 2017, 1523 :17-39
[5]   Slurry concentration effects on the bed morphology and separation efficiency of capillaries packed with sub-2 μm particles [J].
Bruns, Stefan ;
Franklin, Edward G. ;
Grinias, James P. ;
Godinho, Justin M. ;
Jorgenson, James W. ;
Tallarek, Ulrich .
JOURNAL OF CHROMATOGRAPHY A, 2013, 1318 :189-197
[6]  
Cajka T, 2017, METHODS MOL BIOL, V1609, P149, DOI 10.1007/978-1-4939-6996-8_14
[7]   Hydrophilic interaction liquid chromatography-mass spectrometry of (lyso)phosphatidic acids, (lyso)phosphatidylserines and other lipid classes [J].
Cifkova, Eva ;
Hajek, Roman ;
Lisa, Miroslav ;
Holcapek, Michal .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1439 :65-73
[8]   Nano-LC/NSI MS Refines Lipidomics by Enhancing Lipid Coverage, Measurement Sensitivity, and Linear Dynamic Range [J].
Danne-Rasche, Niklas ;
Coman, Cristina ;
Ahrends, Robert .
ANALYTICAL CHEMISTRY, 2018, 90 (13) :8093-8101
[9]   On the feasibility to conduct gradient liquid chromatography separations in narrow-bore columns at pressures up to 2000 bar [J].
De Pauw, Ruben ;
Swier, Tim ;
Degreef, Bart ;
Desmet, Gert ;
Broeckhoven, Ken .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1473 :48-55
[10]   Advances in Ultrahigh-Pressure Liquid Chromatography Technology and System Design [J].
De Vos, Jelle ;
Broeckhoven, Ken ;
Eeltink, Sebastiaan .
ANALYTICAL CHEMISTRY, 2016, 88 (01) :262-278