High-Resolution Mapping of Scar-Related Atrial Arrhythmias Using Smaller Electrodes With Closer Interelectrode Spacing

被引:180
作者
Anter, Elad
Tschabrunn, Cory M.
Josephson, Mark E.
机构
[1] Beth Israel Deaconess Med Ctr, Harvard Thorndike Electrophysiol Inst, Dept Med, Cardiovasc Div, Boston, MA 02215 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
关键词
arrhythmias; cardiac; catheter ablation; electrophysiology; CATHETER ABLATION; FIBRILLATION; TACHYCARDIA; RHYTHM;
D O I
10.1161/CIRCEP.114.002737
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The resolution of mapping is influenced by electrode size and interelectrode spacing. Smaller electrodes with closer interelectrode spacing may improve mapping resolution, particularly in scar. The aims of this study were to establish normal electrogram criteria in the atria for both 3.5-mm electrode tip linear catheters (Thermocool) and 1-mm multielectrode-mapping catheters (Pentaray) and to compare their mapping resolution in scar-related atrial arrhythmias. Methods and Results Normal voltage amplitude cutoffs for both catheters were validated in 10 patients with structurally normal atria. In 20 additional patients with scar-related atrial arrhythmias, similar sequential mapping with both catheters was performed. Normal bipolar voltage amplitude was similar between 3.5- and 1-mm electrode catheters with a fifth percentile of 0.48 and 0.52 mV, respectively (P=0.65). In patients with scar-related atrial arrhythmias, the total area of bipolar voltage <0.5 mV measured using 1-mm electrode catheters was smaller than that measured using 3.5-mm catheter (14.7 versus 20.4 cm2; P=0.02). The mean bipolar voltage amplitude in this area of low voltage was significantly higher with 1-mm electrode catheters (0.28 and 0.17 mV; P=0.01). Importantly, 54.4% of all low voltage data points recorded with 1-mm electrode catheter had distinct electrograms that allowed annotation of local activation time compared with only 21.4% with 3.5-mm electrode tip catheters (P=0.01). Overdrive pacing with capture of the tachycardia from within the area of low voltage was more frequent with 1-mm electrode catheters (66.7 versus 33.4; P=0.01). Conclusions Mapping with small closely spaced electrode catheters can improve mapping resolution within areas of low voltage.
引用
收藏
页码:537 / 545
页数:9
相关论文
共 12 条
[11]   Atrial Tachycardia After Ablation of Persistent Atrial Fibrillation Identification of the Critical Isthmus With a Combination of Multielectrode Activation Mapping and Targeted Entrainment Mapping [J].
Patel, Anshul M. ;
d'Avila, Andre ;
Neuzil, Petr ;
Kim, Steven J. ;
Mela, Theofanie ;
Singh, Jagmeet P. ;
Ruskin, Jeremy N. ;
Reddy, Vivek Y. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2008, 1 (01) :14-22
[12]   Long-Term Outcome After Successful Catheter Ablation of Atrial Fibrillation [J].
Tzou, Wendy S. ;
Marchlinski, Francis E. ;
Zado, Erica S. ;
Lin, David ;
Dixit, Sanjay ;
Callans, David J. ;
Cooper, Joshua M. ;
Bala, Rupa ;
Garcia, Fermin ;
Hutchinson, Mathew D. ;
Riley, Michael P. ;
Verdino, Ralph ;
Gerstenfeld, Edward P. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (03) :237-242