Effect of frequency on pressure cost of ventilation and gas exchange in newborns receiving high-frequency oscillatory ventilation

被引:14
|
作者
Zannin, Emanuela [1 ]
Dellaca', Raffaele L. [1 ]
Dognini, Giulia [2 ]
Marconi, Lara [1 ]
Perego, Martina [1 ]
Pillow, Jane J. [3 ]
Tagliabue, Paolo E. [2 ]
Ventura, Maria Luisa [2 ]
机构
[1] Politecn Milano Univ, Dipartimento Elettron Informaz & Bioingn, Milan, Italy
[2] Fdn MBBM ASST, Obstetr & Pediat Dept, Neonatal Intens Care Unit, Monza, Italy
[3] Univ Western Australia, Sch Anat Physiol & Human Biol, Ctr Neonatal Res & Educ, Perth, WA, Australia
关键词
RESPIRATORY-DISTRESS-SYNDROME; ALVEOLAR PRESSURE; RESONANCE FREQUENCY; ENDOTRACHEAL-TUBE; SYSTEM; FLOW; DEPENDENCE; MECHANICS; INFANTS; MODEL;
D O I
10.1038/pr.2017.151
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
BACKGROUND: We hypothesized that ventilating at the resonant frequency of the respiratory system optimizes gas exchange while limiting the mechanical stress to the lung in newborns receiving high-frequency oscillatory ventilation (HFOV). We characterized the frequency dependence of oscillatory mechanics, gas exchange, and pressure transmission during HFOV. METHODS: We studied 13 newborn infants with a median (interquartile range) gestational age of 29.3 (26.4-30.4) weeks and body weight of 1.00 (0.84-1.43) kg. Different frequencies (5, 8, 10, 12, and 15 Hz) were tested, keeping carbon dioxide diffusion coefficient (DCO2) constant. Oscillatory mechanics and transcutaneous blood gas were measured at each frequency. The attenuation of pressure swings (Delta P) from the airways opening to the distal end of the tracheal tube (TT) and to the alveolar compartment was mathematically estimated. RESULTS: Blood gases were unaffected by frequency. The mean (SD) resonant frequency was 16.6 (3.5) Hz. Damping of Delta P increased with frequency and with lung compliance. Delta P at the distal end of the TT was insensitive to frequency, whereas Delta P at the peripheral level decreased with frequency. CONCLUSION: There is no optimal frequency for gas exchange when DCO2 is held constant. Greater attenuation of oscillatory pressure at higher frequencies offers more protection from barotrauma, especially in patients with poor compliance.
引用
收藏
页码:994 / 999
页数:6
相关论文
共 50 条
  • [1] Effect of frequency on pressure cost of ventilation and gas exchange in newborns receiving high-frequency oscillatory ventilation
    Emanuela Zannin
    Raffaele L Dellaca'
    Giulia Dognini
    Lara Marconi
    Martina Perego
    Jane J Pillow
    Paolo E Tagliabue
    Maria Luisa Ventura
    Pediatric Research, 2017, 82 : 994 - 999
  • [2] EFFECT OF MEAN AIRWAY PRESSURE ON GAS-EXCHANGE DURING HIGH-FREQUENCY OSCILLATORY VENTILATION
    BOYNTON, BR
    VILLANUEVA, D
    HAMMOND, MD
    VREELAND, PN
    BUCKLEY, B
    FRANTZ, ID
    JOURNAL OF APPLIED PHYSIOLOGY, 1991, 70 (02) : 701 - 707
  • [3] COMPARISON OF HIGH-FREQUENCY OSCILLATORY VENTILATION AND HIGH-FREQUENCY JET VENTILATION
    BOROS, SJ
    MAMMEL, MC
    HORCHER, P
    GORDON, MJ
    CLINICAL RESEARCH, 1987, 35 (01): : A240 - A240
  • [4] Specific compliance and gas exchange during high-frequency oscillatory ventilation
    Wood, B
    Karna, P
    Adams, A
    CRITICAL CARE MEDICINE, 2002, 30 (07) : 1523 - 1527
  • [5] Pendelluft effect on gas exchange in high-frequency ventilation
    Tanida, Y
    JSME INTERNATIONAL JOURNAL SERIES B-FLUIDS AND THERMAL ENGINEERING, 1996, 39 (02) : 368 - 375
  • [6] HIGH-FREQUENCY OSCILLATORY VENTILATION
    SALLE, BL
    CLARIS, O
    PUTET, G
    PEDIATRIE, 1993, 48 (12): : 861 - 863
  • [7] High-frequency oscillatory ventilation
    Greer, Sarah E.
    McCunn, Maureen
    CURRENT PROBLEMS IN SURGERY, 2013, 50 (10) : 471 - 478
  • [8] GAS TRAPPING WITH HIGH-FREQUENCY VENTILATION - JET VERSUS OSCILLATORY VENTILATION
    BANCALARI, A
    GERHARDT, T
    BANCALARI, E
    SUGUIHARA, C
    HEHRE, D
    REIFENBERG, L
    GOLDBERG, RN
    JOURNAL OF PEDIATRICS, 1987, 110 (04): : 617 - 622
  • [9] Gas exchange during high frequency oscillatory ventilation
    Dembinski, R
    Max, M
    Bensberg, R
    Bickenbach, J
    Kuhlen, R
    Rossaint, R
    INTENSIVE CARE MEDICINE, 2001, 27 : S157 - S157
  • [10] HIGH-FREQUENCY JET VENTILATION IMPROVES GAS-EXCHANGE WHEN COMPARED TO HIGH-FREQUENCY OSCILLATORY VENTILATION AT EQUIVALENT AIRWAY PRESSURES
    BOROS, SJ
    MAMMEL, MC
    HORCHER, P
    GORDON, MJ
    PEDIATRIC RESEARCH, 1987, 21 (04) : A445 - A445