Noninvasive monitoring of hemoglobin - The effects of WBC counts on measurement

被引:16
作者
Saigo, K
Imoto, S
Hashimoto, M
Mito, H
Moriya, J
Chinzei, T
Kubota, Y
Numada, S
Ozawa, T
Kumagai, S
机构
[1] Kobe Univ Hosp, Div Blood Transfus, Chuo Ku, Kobe, Hyogo 6500017, Japan
[2] Kakogawa Municipal Hosp, Dept Med, Kakogawa, Japan
[3] Hyogo Red Cross Blood Ctr, Kobe, Hyogo, Japan
[4] Kagawa Med Univ Hosp, Dept Transfus Med, Kagawa, Japan
[5] Sysmex, Kobe, Hyogo, Japan
关键词
noninvasive monitoring; hemoglobin; astrim; near-infrared radiation; leukocyte; WBC;
D O I
10.1309/61ANJ1XV3MW878YP
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The efficacy of a noninvasive hemoglobin monitoring device (Astrim, Sysmex, Kobe, Japan) was evaluated for healthy volunteers and for patients with hematologic disorders. At the same time, the effects of WBC counts on noninvasive monitoring were studied by clinical evaluation and in ex vivo experiments. The hemoglobin levels determined by the device (Ast-Hb) and a conventional analyzer (T-Hb) were compared. The coefficient of correlation between findings with the Ast-Hb and the T-Hb for healthy volunteers was r = 0.626, whereas that for patients with hematologic disorders was r = 0.762. A comparison of the ratios of measurement errors in hemoglobin levels by Ast-Hb and T-Hb indicated that the number of WBCs had no effect on hemoglobin monitoring. Moreover, ex vivo studies using isolated WBCs and an optical model that imitates blood vessels and tissue in human fingers confirmed these results. Therefore, this new hemoglobin monitoring device can be expected to be useful for continuous hemoglobin monitoring.
引用
收藏
页码:51 / 55
页数:5
相关论文
共 9 条
  • [1] HALL JW, 1992, CLIN CHEM, V38, P1623
  • [2] The definition of anemia in older persons
    Izaks, GJ
    Westendorp, RGJ
    Knook, DL
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 281 (18): : 1714 - 1717
  • [3] NONINVASIVE, INFRARED MONITORING OF CEREBRAL AND MYOCARDIAL OXYGEN SUFFICIENCY AND CIRCULATORY PARAMETERS
    JOBSIS, FF
    [J]. SCIENCE, 1977, 198 (4323) : 1264 - 1267
  • [4] Measurement of hemoglobin concentrations using the Astrim™ noninvasive blood vessel monitoring apparatus
    Kinoshita, Y
    Yamane, T
    Takubo, T
    Kanashima, H
    Kamitani, T
    Tatsumi, N
    Hino, M
    [J]. ACTA HAEMATOLOGICA, 2002, 108 (02) : 109 - 110
  • [5] Lev A, 2002, ACT NEUR S, V81, P295
  • [6] ACCURACY OF PULSE OXIMETRY AT VARIOUS HEMATOCRITS AND DURING HEMOLYSIS IN AN INVITRO MODEL
    VEGFORS, M
    LINDBERG, LG
    OBERG, PA
    LENNMARKEN, C
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1993, 31 (02) : 135 - 141
  • [7] WADA K, 1998, SYSMEX J, V21, P171
  • [8] Near-infrared spectroscopy: Theory and applications
    Wahr, JA
    Tremper, KK
    Samra, S
    Delpy, DT
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1996, 10 (03) : 406 - 418
  • [9] Yonemura M., 1999, SYSMEX J, V22, P249