Objective: To study the feasibility of percutaneous cryoablation of hepatic tumours monitored by magnetic resonance imaging (MRI). Design: Prospective study Setting: University hospital, Norway Patients: Six patients with hepatic metastases from colorectal cancer. Interventions: Percutaneous cryoprobe positioning under general anaesthesia. Positioning and freezing monitored by near-real-time MRI using an open 0.5 Tesla MM configuration system. Main outcome measures: Safety and feasibility of the procedure. Measurement of volumes of cryolesions. Results: One patient developed a biliary leakage that had to be drained. Four patients developed pleural fluid. Two small tumours were adequately cryoablated. In the remaining 4 patients with large (>4 cm) tumours, an adequate cryolesion could not be formed. Cryolesion volumes larger than 105 cm(3). were not produced even using 3-4 probes. MRI visualised the growing cryolesion well, but positioning of the cryoprobes was time-consuming. Conclusion: MR guided cryoablation is clinically feasible and gives good visualisation of the procedure, Patients with small tumours (<3 cm) seem to be best suited to this percutaneous approach as cryolesion volumes claimed to be adequate for tumour destruction can be produced. Measurement of tumour volume preoperatively may help to select patients who win respond.
引用
收藏
页码:610 / 617
页数:8
相关论文
共 50 条
  • [1] Miniaturizing Floating Traps to Increase RF Safety of Magnetic-Resonance-Guided Percutaneous Procedures
    Griffin, Gregory H.
    Anderson, Kevan J. T.
    Wright, Graham A.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (02) : 329 - 340
  • [3] Magnetic resonance imaging-guided percutaneous cryoablation of hepatocellular carcinoma in special regions
    Wu, Bin
    Xiao, Yue-Yong
    Zhang, Xiao
    Zhang, Ai-Lian
    Li, Hong-Jun
    Gao, Deng-Fa
    HEPATOBILIARY & PANCREATIC DISEASES INTERNATIONAL, 2010, 9 (04) : 384 - 392
  • [4] Magnetic-resonance-guided biopsy of focal liver lesions
    Smith, Ethan A.
    Grove, Jason J.
    Van der Spek, Abraham F. L.
    Jarboe, Marcus D.
    PEDIATRIC RADIOLOGY, 2017, 47 (06) : 750 - 754
  • [5] Magnetic-resonance-guided biopsy of focal liver lesions
    Ethan A. Smith
    Jason J. Grove
    Abraham F. L. Van Der Spek
    Marcus D. Jarboe
    Pediatric Radiology, 2017, 47 : 750 - 754
  • [6] Percutaneous cryoablation of porcine kidneys with magnetic resonance imaging monitoring
    Shingleton, WB
    Farabaugh, P
    Hughson, M
    Sewell, PE
    JOURNAL OF UROLOGY, 2001, 166 (01) : 289 - 291
  • [7] Percutaneous renal tumor cryoablation with magnetic resonance imaging guidance
    Shingleton, WB
    Sewell, PE
    JOURNAL OF UROLOGY, 2001, 165 (03) : 773 - 776
  • [8] Magnetic-resonance-guided radiotherapy The beginning of a new era in radiation oncology?
    Hoegen, P.
    Spindeldreier, C. K.
    Buchele, C.
    Rippke, C.
    Regnery, S.
    Weykamp, F.
    Klueter, S.
    Debus, J.
    Hoerner-Rieber, J.
    RADIOLOGE, 2021, 61 (01): : 13 - 20
  • [9] Identifying the predictors of acute urinary retention following magnetic-resonance-guided prostate brachytherapy
    Thomas, MD
    Cormack, R
    Tempany, CM
    Kumar, S
    Manola, J
    Schneider, L
    Hurwitz, MD
    D'Amico, AV
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04): : 905 - 908
  • [10] Percutaneous cryoablation of benign bony tumours of the mandible
    May, L.
    Blatter, J.
    Bize, P.
    Tsoumakidou, G.
    Denys, A.
    Broome, M.
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2020, 58 (01) : 75 - 78