Evaluation of two point-of-care meters and a portable chemistry analyzer for measurement of blood glucose concentrations in juvenile white-tailed deer (Odocoileus virginianus)

被引:14
作者
Burdick, Stacy [1 ]
Mitchell, Mark A. [1 ]
Neil, Johanna [1 ]
Heggem, Brittany [1 ]
Whittington, Julia [1 ]
Acierno, Mark J. [2 ]
机构
[1] Univ Illinois, Dept Vet Clin Med, Coll Vet Med, Urbana, IL 61802 USA
[2] Louisiana State Univ, Sch Vet Med, Dept Vet Clin Sci, Baton Rouge, LA 70810 USA
来源
JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION | 2012年 / 240卷 / 05期
关键词
AGREEMENT; ACCURACY;
D O I
10.2460/javma.240.5.596
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To evaluate agreement of blood glucose concentrations measured in juvenile white-tailed deer (Odocoileus virginianus) by use of 2 point-of-care (POC) blood glucose meters and 1 portable chemistry analyzer with values obtained in serum by use of a standard laboratory chemistry analyzer, and to evaluate agreement between results obtained with the 2 POC meters. Design-Prospective evaluation study. Sample-14 venous blood samples from 14 healthy white-tailed deer fawns. Procedures-Blood glucose concentration was measured with each of 2 POC meters. The remainder of the sample was divided into 2 tubes (1 that contained lithium heparin and 1 with no anticoagulant). Glucose concentration in anticoagulated whole blood was measured with the portable analyzer. Serum was collected from the remaining sample for measurement of glucose concentrations with the laboratory analyzer. Bland-Altman plots were used to assess agreement. Results-Agreement between POC blood glucose meters and the laboratory analyzer was poor; mean values for bias were 2.9 mg/dL (95% limits of agreement [LOA], -70.2 to 76.0 mg/dL) and -30.8 mg/dL (95% LOA, -111.6 to 49.9 mg/dL), respectively. Agreement between the 2 POC meters was also poor (bias, 31.0 mg/dL; 95% LOA, -47.2 to 109.2 mg/dL). Agreement between the portable analyzer and the laboratory analyzer was good (bias, -1.6 mg/dL; 95% LOA, -15.3 to 12.1 mg/dL). Conclusions and Clinical Relevance-Results suggested that the POC blood glucose meters used in this study are not appropriate for measurement of blood glucose concentrations in juvenile white-tailed deer. (J Am Vet Med Assoc 2012;240:596-599)
引用
收藏
页码:596 / 599
页数:4
相关论文
共 12 条
[1]   Evaluation of the agreement among three handheld blood glucose meters and a laboratory blood analyzer for measurement of blood glucose cocentration Hispaniolan Amazon parrots (Amazona ventralis) [J].
Acierno, Mark J. ;
Mitchell, Mark A. ;
Schuster, Patricia J. ;
Freeman, Diana ;
Guzman, David Sanchez-Migallon ;
Tully, Thomas N., Jr. .
AMERICAN JOURNAL OF VETERINARY RESEARCH, 2009, 70 (02) :172-175
[2]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Accuracy of devices used for self-monitoring of blood glucose [J].
Johnson, RN ;
Baker, JR .
ANNALS OF CLINICAL BIOCHEMISTRY, 1998, 35 :68-74
[5]  
Lieske CL, 2002, J AVIAN MED SURG, V16, P277, DOI 10.1647/1082-6742(2002)016[0277:EOHBGM]2.0.CO
[6]  
2
[7]  
MOLITCH ME, 1994, DIABETES CARE, V17, P81
[8]  
Reusch CE, 2001, COMP CONT EDUC PRACT, V23, P544
[9]  
Tang ZP, 2000, ARCH PATHOL LAB MED, V124, P1135
[10]   Accuracy of home blood glucose meters during hypoglycemia [J].
Trajanoski, Z ;
Brunner, GA ;
Gfrerer, RJ ;
Wach, P ;
Pieber, TR .
DIABETES CARE, 1996, 19 (12) :1412-1415