Flow assessment as a function of pump timing of tubular pulsatile pump for use as a ventricular assist device in a left heart simulator

被引:0
作者
Sharifi, Alireza [1 ]
Bark, David, Jr. [2 ,3 ]
机构
[1] Colorado State Univ, Dept Mech Engn, Ft Collins, CO 80523 USA
[2] Washington Univ, Div Hematol & Oncol, Dept Pediat, 660 S Euclid Ave,Campus Box 8208,5th Floor MPRB, St Louis, MO 63108 USA
[3] Washington Univ, Dept Biomed Engn, St Louis, MO 63108 USA
关键词
heart tube; hemodynamics; left heart simulator; mechanical circulatory support; pneumatic pump; pulsatile; ventricular assist device; FAILURE; ATRIAL; COUNTERPULSATION; CANNULATION; EJECTION; PATIENT; SITES; STATE;
D O I
10.1111/aor.14196
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction Although mechanical circulatory support saved many lives during the last decade, clinical observations have shown that the continuous flow pumps are associated with a much higher incidence of gastrointestinal bleeding and kidney problems, among others, compared with the earlier generation pulsatile pumps. However, the presence of several moving mechanical components made pulsatile pumps less durable, bulky, and prone to malfunction, ultimately leading to bias in favor of continuous flow designs. Objective The aim of the current work is to create a prototype tubular pulsatile pump and to test the timing of the pump in a left heart simulator. Methods A left heart simulator to mimic pumping from a failing heart was created. This was used to experimentally test the output of a prototype ventricular assist device relative to a failing heart in the form of flow and pressure. The effect of pulsation timing was quantified. Results A failing heart was simulated with an average flow rate of 1.1 L/min and a systolic pressure of 47 mm Hg. With the pump, the flow rate increases to 4.8 L/min and a systolic pressure of 110 mm Hg, in a copulsation mode, while activating for 300-400 ms. If the activation time is reduced, or increased, the pump becomes less effective. Load on the heart is reduced when the pump operates in a counterpulsation mode. Conclusion A pulsatile pump, like the one proposed, provides adequate output for mechanical circulatory support, while minimizing the number of moving parts that could otherwise lead to tribological wear.
引用
收藏
页码:1294 / 1304
页数:11
相关论文
共 51 条
[1]  
Bachus L., 2003, Know and Understand Centrifugal Pumps
[2]   An unusual thrombus location in a Heartmate 3TM device with fatal outcome [J].
Banfi, Carlo ;
Rigamonti, Fabio ;
Ahmadov, Kamran ;
Meyer, Philippe ;
Hachulla, Anne-Lise ;
Craviari, Cecilia ;
Fontana, Pierre ;
Bendjelid, Karim ;
Giraud, Raphael .
PERFUSION-UK, 2020, 35 (05) :442-446
[3]   Hemodynamic Performance and Thrombogenic Properties of a Superhydrophobic Bileaflet Mechanical Heart Valve [J].
Bark, David L., Jr. ;
Vahabi, Hamed ;
Hieu Bui ;
Movafaghi, Sanli ;
Moore, Brandon ;
Kota, Arun K. ;
Popat, Ketul ;
Dasi, Lakshmi P. .
ANNALS OF BIOMEDICAL ENGINEERING, 2017, 45 (02) :452-463
[4]   Effect of Outflow Graft Size on Flow in the Aortic Arch and Cerebral Blood Flow in Continuous Flow Pumps: Possible Relevance to Strokes [J].
Bhat, Sindhor ;
Mathew, Jayakala ;
Balakrishnan, Komrakshi R. ;
Kumar, Ramarathnam Krishna .
ASAIO JOURNAL, 2017, 63 (02) :144-149
[5]   Fluid mechanics of arterial stenosis: Relationship to the development of mural thrombus [J].
Bluestein, D ;
Niu, LJ ;
Schoephoerster, RT ;
Dewanjee, MK .
ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (02) :344-356
[6]   Thrombotic Risk of Rotor Speed Modulation Regimes of Contemporary Centrifugal Continuous-flow Left Ventricular Assist Devices [J].
Boraschi, Andrea ;
Bozzi, Silvia ;
Thamsen, Bente ;
Granegger, Marcus ;
Wiegmann, Lena ;
Pappalardo, Federico ;
Slepian, Marvin J. ;
Kurtcuoglu, Vartan ;
Redaelli, Alberto ;
De Zelicourt, Diane ;
Consolo, Filippo .
ASAIO JOURNAL, 2021, 67 (07) :737-745
[7]   Turbulent Flow Promotes Cleavage of VWF (von Willebrand Factor) by ADAMTS13 (A Disintegrin and Metalloproteinase With a Thrombospondin Type-1 Motif, Member 13) [J].
Bortot, Maria ;
Ashworth, Katrina ;
Sharifi, Alireza ;
Walker, Faye ;
Crawford, Nathan C. ;
Neeves, Keith B. ;
Bark, David, Jr. ;
Di Paola, Jorge .
ARTERIOSCLEROSIS THROMBOSIS AND VASCULAR BIOLOGY, 2019, 39 (09) :1831-1842
[8]   Assessment and Management of Heart Failure After Left Ventricular Assist Device Implantation [J].
Burke, Michael A. ;
Givertz, Michael M. .
CIRCULATION, 2014, 129 (10) :1161-U197
[9]   Clinical Implications of the New York Heart Association Classification [J].
Caraballo, Cesar ;
Desai, Nihar R. ;
Mulder, Hillary ;
Alhanti, Brooke ;
Wilson, F. Perry ;
Fiuzat, Mona ;
Felker, G. Michael ;
Pina, Ileana L. ;
O'Connor, Christopher M. ;
Lindenfeld, Joanne ;
Januzzi, James L. ;
Cohen, Lawrence S. ;
Ahmad, Tariq .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2019, 8 (23)
[10]   Evolvement of left ventricular assist device: the implications on heart failure management [J].
Chair, Sek Ying ;
Yu, Doris S. F. ;
Ng, Michael Timothy ;
Wang, Qun ;
Cheng, Ho Yu ;
Wong, Eliza M. L. ;
Sit, Janet W. H. .
JOURNAL OF GERIATRIC CARDIOLOGY, 2016, 13 (05) :425-430