Multivariate analysis applied to the study of spatial distributions found in drug-eluting stent, coatings by confocal Raman microscopy

被引:18
作者
Balss, Karin M. [1 ]
Long, Frederick H. [2 ]
Veselov, Vladimir [1 ]
Orana, Argjenta [1 ]
Akerman-Revis, Eugena [1 ]
Papandreou, George [1 ]
Maryanoff, Cynthia A. [1 ]
机构
[1] Johnson & Johnson Co, Cordis Corp, Spring House, PA 19477 USA
[2] Spectroscop Solut, Randolph, NJ 07869 USA
关键词
D O I
10.1021/ac7025767
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Multivariate data analysis was applied to confocal Raman measurements on stents coated with the polymers and drug used in the CYPHER Sirolimus-eluting Coronary Stents. Partial least-squares (PLS) regression was used to establish three independent calibration curves for the coating constituents: sirolimus, poly(n-butyl methacrylate) [PBMA], and poly(ethylene-co-vinyl acetate) [PEVA]. The PLS calibrations were based on average spectra generated from each spatial location profiled. The PLS models were tested on six unknown stent samples to assess accuracy and precision. The wt % difference between PLS predictions and laboratory assay values for sirolimus was less than 1 wt % for the composite of the six unknowns, while the polymer models were estimated to be less than 0.5 wt % difference for the combined samples. The linearity and specificity of the three PLS models were also demonstrated with the three PLS models. In contrast to earlier univaiiate models, the PLS models achieved mass balance with better accuracy. This analysis was extended to evaluate the spatial distribution of the three constituents. Quantitative bitmap images of drug-eluting stent coatings are presented for the first time to assess the local distribution of components.
引用
收藏
页码:4853 / 4859
页数:7
相关论文
共 35 条
[1]   Fourier transform Raman microscopic study of drug distribution in a transdermal drug delivery device [J].
Armstrong, GL ;
Edwards, HGM ;
Farwell, DW ;
Williams, AC .
VIBRATIONAL SPECTROSCOPY, 1996, 11 (02) :105-113
[2]   Quantitative spatial distribution of sirolimus and polymers in drug-eluting stents using confocal Raman microscopy [J].
Balss, K. M. ;
LIanos, G. ;
Papandreou, G. ;
Maryanoff, C. A. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2008, 85A (01) :258-270
[3]   Chemical imaging of drug eluting coatings: Combining surface analysis and confocal Raman microscopy [J].
Belu, Anna ;
Mahoney, Christine ;
Wormuth, Klaus .
JOURNAL OF CONTROLLED RELEASE, 2008, 126 (02) :111-121
[4]   Confocal Raman-spectroscopy: Analytical approach to solid dispersions and mapping of drugs [J].
Breitenbach, J ;
Schrof, W ;
Neumann, J .
PHARMACEUTICAL RESEARCH, 1999, 16 (07) :1109-1113
[5]   Monitoring the penetration enhancer dimethyl sulfoxide in human stratum corneum in vivo by confocal Raman spectroscopy [J].
Caspers, PJ ;
Williams, AC ;
Carter, EA ;
Edwards, HGM ;
Barry, BW ;
Bruining, HA ;
Puppels, GJ .
PHARMACEUTICAL RESEARCH, 2002, 19 (10) :1577-1580
[6]   FTIR, FT-Raman and chemometrics: Applications to the analysis and characterisation of polymers [J].
Chalmers, JM ;
Everall, NJ .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (01) :18-25
[7]   A novel drug-eluting stent spray-coated with multi-layers of collagen and sirolimus [J].
Chen, MC ;
Liang, HF ;
Chiu, YL ;
Chang, Y ;
Wei, HJ ;
Sung, HW .
JOURNAL OF CONTROLLED RELEASE, 2005, 108 (01) :178-189
[8]  
*CORD CORP, 2007, UNPUB
[9]   FT RAMAN-SPECTROSCOPY OF DRUGS IN POLYMERS [J].
DAVIES, MC ;
BINNS, JS ;
MELIA, CD ;
HENDRA, PJ ;
BOURGEOIS, D ;
CHURCH, SP ;
STEPHENSON, PJ .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1990, 66 (1-3) :223-232
[10]   Nanoparticles of poly(D,L-lactide)/methoxy poly(ethylene glycol)-poly(D,L-lactide) blends for controlled release of paclitaxel [J].
Dong, Yuancai ;
Feng, Si-Shen .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2006, 78A (01) :12-19