Electrochemical POC device for fast malaria quantitative diagnosis in whole blood by using magnetic beads, Poly-HRP and microfluidic paper electrodes

被引:58
作者
Ruiz-Vega, Gisela [1 ]
Arias-Alpizar, Kevin [1 ]
de la Serna, Erica [1 ]
Neves Borgheti-Cardoso, Livia [2 ,3 ]
Sulleiro, Elena [4 ,5 ]
Molina, Israel [5 ,6 ]
Fernandez-Busquets, Xavier [2 ,3 ,7 ]
Sanchez-Montalva, Adrian [5 ,6 ]
Javier del Campo, F. [8 ]
Baldrich, Eva [1 ,9 ]
机构
[1] UAB, Vail dHebron Res Inst VHIR, CIBBIM Nanomed, Diagnost Nanotools Grp, Barcelona, Spain
[2] BIST, Inst Bioengn Catalonia IBEC, Nanomalaria Grp, Barcelona, Spain
[3] Univ Barcelona, Hosp Clin, ISGlobal, Barcelona Inst Global Hlth, Barcelona, Spain
[4] UAB, Vall dHebron Univ Hosp VHUH, Microbiol Dept, Barcelona, Spain
[5] PROSICS Catalan Int Hlth Program, Barcelona, Spain
[6] UAB, PROSICS Catalan Int Hlth Program, VHUH, Infect Dis Dept, Barcelona, Spain
[7] Nanosci & Nanotechnol Inst IN2UB, Barcelona, Spain
[8] CSIC, CNN, IMB, Ctr Nacl Microelect, Barcelona, Spain
[9] CIBER Bioingn Biomat & Nanomed CIBER BBN, Madrid, Spain
基金
欧盟地平线“2020”;
关键词
Electrochemical magneto-immunosensor; Paper microfluidic electrode; Malaria quantitative diagnosis; Plasmodium LDH; Point-of-care (POC) testing; CAPTURE; IMMUNOASSAYS; ANTIGEN; ASSAYS; MMP-9; PCR;
D O I
10.1016/j.bios.2019.111925
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Malaria, a parasitic infection caused by Plasmodium parasites and transmitted through the bite of infected female Anopheles mosquitos, is one of the main causes of mortality in many developing countries. Over 200 million new infections and nearly half a million deaths are reported each year, and more than three billion people are at risk of acquiring malaria worldwide. Nevertheless, most malaria cases could be cured if detected early. Malaria eradication is a top priority of the World Health Organisation. However, achieving this goal will require mass population screening and treatment, which will be hard to accomplish with current diagnostic tools. We report an electrochemical point-of-care device for the fast, simple and quantitative detection of Plasmodium falciparum lactate dehydrogenase (PfLDH) in whole blood samples. Sample analysis includes 5-min lysis to release intracellular parasites, and stirring for 5 more min with immuno-modified magnetic beads (MB) along with an immuno-modified signal amplifier. The rest of the magneto-immunoassay, including sample filtration, MB washing and electrochemical detection, is performed at a disposable paper electrode microfluidic device. The sensor provides PfLDH quantitation down to 2.47 ng mL(-1) in spiked samples and for 0.006-1.5% parasitemias in Plasmodium-infected cultured red blood cells, and discrimination between healthy individuals and malaria patients presenting parasitemias >0.3%. Quantitative malaria diagnosis is attained with little user intervention, which is not achieved by other diagnostic methods.
引用
收藏
页数:8
相关论文
共 43 条
[1]  
[Anonymous], HELP YOU BUILD SMART
[2]  
[Anonymous], 2015, Strategy for malaria elimination in the Greater Mekong Subregion: 2015-2030, DOI 9789241564991
[3]  
[Anonymous], BIOMALPAR
[4]  
[Anonymous], J CLIN MICROBIOL
[5]  
[Anonymous], 2015, Malaria rapid diagnostic test performance: results of WHO product testing of malaria RDTs: round 7 (2015-2016)
[6]   Malaria [J].
Ashley, Elizabeth A. ;
Phyo, Aung Pyae ;
Woodrow, Charles J. .
LANCET, 2018, 391 (10130) :1608-1621
[7]   Using magnetic beads and signal amplifiers to produce short and simple immunoassays: Application to MMP-9 detection in plasma samples [J].
Ben Ismail, Manel ;
de la Serna, Erica ;
Ruiz-Vega, Gisela ;
Garcia-Berrocoso, Teresa ;
Montaner, Joan ;
Zourob, Mohammed ;
Othmane, Ali ;
Baldrich, Eva .
ANALYTICA CHIMICA ACTA, 2018, 999 :144-154
[8]   Magneto Immunoassays for Plasmodium falciparum Histidine-Rich Protein 2 Related to Malaria based on Magnetic Nanoparticles [J].
de Souza Castilho, M. ;
Laube, T. ;
Yamanaka, H. ;
Alegret, S. ;
Pividori, M. I. .
ANALYTICAL CHEMISTRY, 2011, 83 (14) :5570-5577
[9]   Multiplexed, Patterned-Paper Immunoassay for Detection of Malaria and Dengue Fever [J].
Deraney, Rachel N. ;
Mace, Charles R. ;
Rolland, Jason P. ;
Schonhorn, Jeremy E. .
ANALYTICAL CHEMISTRY, 2016, 88 (12) :6161-6165
[10]   Disposable Sensors in Diagnostics, Food, and Environmental Monitoring [J].
Dincer, Can ;
Bruch, Richard ;
Costa-Rama, Estefania ;
Teresa Fernandez-Abedu, Maria ;
Merkoci, Arben ;
Manz, Andreas ;
Urban, Gerald Anton ;
Gueder, Firat .
ADVANCED MATERIALS, 2019, 31 (30)