Experimental pre-clinical assessment of the primary stability of two cementless femoral knee components

被引:11
作者
Berahmani, Sanaz [1 ]
Hendriks, Maartje [1 ]
Wolfson, David [2 ]
Wright, Abraham [3 ]
Janssen, Dennis [1 ]
Verdonschot, Nico [1 ,4 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Radboud Inst Hlth Sci, Orthopaed Res Lab, POB 9101, NL-6500 HB Nijmegen, Netherlands
[2] DePuy Synthes Joint Reconstruct, Leeds LS11 0BG, W Yorkshire, England
[3] DePuy Synthes Joint Reconstruct, Warsaw, IN USA
[4] Univ Twente, Lab Biomech Engn, Enschede, Netherlands
关键词
Uncemented total knee replacement; Micromotion; Primary stability; Digital image correlation; Loading condition; POROUS-SURFACED IMPLANTS; IN-VITRO; INITIAL STABILITY; INTERFERENCE FIT; BONE FIXATION; HIGH-FLEXION; REPLACEMENT; ARTHROPLASTY; VIVO; MICROMOTION;
D O I
10.1016/j.jmbbm.2017.07.043
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To achieve long-lasting fixation of cementless implants, an adequate primary stability is required. We aimed to compare primary stability of a new cementless femoral knee component (Attune (R)) against a conventional implant (LCS (R)) under different loading conditions. Six pairs of femora were prepared following the normal surgical procedure. Calibrated CT-scans and 3D-optical scans of the bones were obtained to measure bone mineral density (BMD) and cut accuracy, respectively. Micromotions were measured in nine regions of interest at the bone-implant interface using digital image correlation. The reconstructions were subjected to the implant-specific's peak tibiofemoral load of gait and a deep knee bend loading profiles. Afterwards, the implants were pushed-off at a flexion angle of 150 degrees. Micromotions of Attune were significantly lower than LCS under both loading conditions (P <= 0.001). Cut accuracy did not affect micromotions, and BMD was only a significant factor affecting the micromotions under simplified gait loading. No significant difference was found in high-flex push off force, but Attune required a significantly higher load to generate excessive micromotions during push-off. Parallel anterior and posterior bone cuts in the LCS versus the tapered bone cuts of the Attune may explain the difference between the two designs. Additionally, the rims at the borders of the LCS likely reduced the area of contact with the bone for the LCS, which may have affected the initial fixation.
引用
收藏
页码:322 / 329
页数:8
相关论文
共 31 条
[1]   Finite element modelling of primary hip stem stability: The effect of interference fit [J].
Abdul-Kadir, Mohammed Rafiq ;
Hansen, Ulrich ;
Klabunde, Ralf ;
Lucas, Duncan ;
Amis, Andrew .
JOURNAL OF BIOMECHANICS, 2008, 41 (03) :587-594
[2]   Does the Method of Component Fixation Influence Clinical Outcomes After Total Knee Replacement? A Systematic Literature Review [J].
Arnold, John B. ;
Walters, Julie L. ;
Solomon, Lucian B. ;
Thewlis, Dominic .
JOURNAL OF ARTHROPLASTY, 2013, 28 (05) :740-746
[3]   Low BMD affects initial stability and delays stem osseointegration in cementless total hip arthroplasty in women A 2-year RSA study of 39 patients [J].
Aro, Hannu T. ;
Alm, Jessica J. ;
Moritz, Niko ;
Makinen, Tatu J. ;
Lankinen, Petteri .
ACTA ORTHOPAEDICA, 2012, 83 (02) :107-114
[4]   Material tailoring of the femoral component in a total knee replacement to reduce the problem of aseptic loosening [J].
Bahraminasab, Marjan ;
Sahari, B. B. ;
Edwards, K. L. ;
Farahmand, Farzam ;
Hong, Tang Sai ;
Naghibi, Hamid .
MATERIALS & DESIGN, 2013, 52 :441-451
[5]   FE analysis of the effects of simplifications in experimental testing on micromotions of uncemented femoral knee implants [J].
Berahmani, Sanaz ;
Janssen, Dennis ;
Wolfson, David ;
Malefijt, Maarten de Waal ;
Fitzpatrick, Clare K. ;
Rullkoetter, Paul J. ;
Verdonschot, Nico .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2016, 34 (05) :812-819
[6]   The Effect of Surface Morphology on the Primary Fixation Strength of Uncemented Femoral Knee Prosthesis: A Cadaveric Study [J].
Berahmani, Sanaz ;
Janssen, Dennis ;
Wolfson, David ;
Rivard, Kori ;
Malefijt, Maarten de Waal ;
Verdonschot, Nico .
JOURNAL OF ARTHROPLASTY, 2015, 30 (02) :300-307
[7]   An experimental study to investigate biomechanical aspects of the initial stability of press-fit implants [J].
Berahmani, Sanaz ;
Janssen, Dennis ;
van Kessel, Sal ;
Wolfson, David ;
Malefijt, Maarten de Waal ;
Buma, Pieter ;
Verdonschot, Nico .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2015, 42 :177-185
[8]   Femoral component loosening in high-flexion total knee replacement AN IN VITRO COMPARISON OF HIGH-FLEXION VERSUS CONVENTIONAL DESIGNS [J].
Bollars, P. ;
Luyckx, J. -P. ;
Innocenti, B. ;
Labey, L. ;
Victor, J. ;
Bellemans, J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (10) :1355-1361
[9]   Biologic fixation of total hip implants - Insights gained from a series of canine studies [J].
Bragdon, CR ;
Jasty, M ;
Greene, M ;
Rubash, HE ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A :105-117
[10]   Analysis of bone-prosthesis interface micromotion for cementless tibial prosthesis fixation and the influence of loading conditions [J].
Chong, Desmond Y. R. ;
Hansen, Ulrich N. ;
Amis, Andrew A. .
JOURNAL OF BIOMECHANICS, 2010, 43 (06) :1074-1080