Thoracic aorta at multi-detector row CT: Motion artifact with various reconstruction windows
被引:23
|作者:
Morgan-Hughes, GJ
Owens, PE
Marshall, AJ
Roobottom, CA
机构:
[1] Plymouth NHS Trust, SW Cardiothorac Ctr, Dept Cardiol, Plymouth PL6 8DH, Devon, England
[2] Plymouth NHS Trust, SW Cardiothorac Ctr, Dept Radiol, Plymouth PL6 8DH, Devon, England
来源:
关键词:
aneurysm;
dissecting;
aorta;
CT;
computed tomography (CT);
artifact;
multidetector row;
D O I:
10.1148/radiol.2282020873
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The authors assessed motion artifact of the thoracic aorta in 25 patients who underwent multi-detector row computed tomography (CT) with retrospective electrocardiographic (ECG) gating. CT reconstructions centered at four phases of diastole were compared for five different levels of the thoracic aorta. A significant positive correlation was observed between heart rate and motion artifact (r = 0.72, P < .001). The optimal reconstruction phase varied between patients, and this was directly related to heart rate. For patients with a heart rate of 70 beats per minute, the reconstruction phase centered at 75% of the R-R interval had the significantly least motion artifact (P = .004). Conversely, the optimal reconstruction phase for patients with heart rates above 70 beats per minute was centered at 50% of the R-R interval (P = .09). (C) RSNA, 2003.