Antimicrobial susceptibility testing determined by Alfred 60/AST (Alifax®) in a multi-sites lab: performance's evaluation and optimization of workflow

被引:10
|
作者
Cupaiolo, R. [1 ,8 ]
Cherkaoui, S. [1 ]
Serrano, G.
Dauby, N. [2 ]
Georgala, A. [3 ]
Blumental, S. [4 ]
Maillart, E. [5 ]
Hites, M.
Hallin, M. [6 ]
Martiny, D. [1 ,7 ]
机构
[1] Univ Libre Bruxelles, Microbiol Dept, Lab Hospitalier Univ Bruxelles Universitair Lab Br, Brussels, Belgium
[2] Univ Libre Bruxelles ULB, Ctr Hosp Univ CHU St Pierre, Infect Dis Dept, Brussels, Belgium
[3] Univ Libre Bruxelles ULB, Inst Jules Bordet, Infect Dis Dept, Brussels, Belgium
[4] Univ Libre Bruxelles ULB, Hop Univ Enfants Reine Fabiola, Paediat Infect Dis Dept, Brussels, Belgium
[5] Univ Libre Bruxelles ULB, CHU Brugmann, Infect Dis Dept, Brussels, Belgium
[6] Univ Libre Bruxelles ULB, Hop Erasme, Infect Dis Dept, Brussels, Belgium
[7] Univ Mons UMONS, Med & Pharm Dept, Mons, Belgium
[8] Dept Lab Med, 322,rue Haute, B-1000 Brussels, Belgium
关键词
Antimicrobial susceptibility testing; Automatization; Blood cultures; Turn-around-time optimization; DESORPTION-IONIZATION-TIME; BACTERIAL IDENTIFICATION; SEPTIC SHOCK; SEPSIS; MORTALITY;
D O I
10.1016/j.mimet.2022.106433
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Purpose: New techniques are needed to speed-up the identification and antimicrobial susceptibility testing (AST) of bacteria associated with bloodstream infections. Alfred 60/AST (Alifax (R), Polverara, Italy) performs AST by light scattering directly from positive blood cultures. Methods: We evaluated Alfred 60/AST performances for 4 months. Each new episode of bacteraemia was included and AST were compared to either our rapid automated AST (Vitek (R) 2) or disk diffusion method. The discrepancies were investigated using Etest (R). The time-to-result (TTR) was evaluated by comparing the blood volume inserted into Alfred 60/AST, i.e. 2 versus 7 blood drops. Taking into account the TTR, the workflow of positive blood cultures and the availability of AST results was studied in order to optimize the implementation of Alfred 60/AST. Results: A total of 249 samples and 1108 antibiotics for AST were tested. After exclusion of unavailable results, 1008 antibiotics were analysed. 94.9% (n = 957/1008) of the antibiotics showed categorical agreement. There were 14 very major errors (VME), 24 major errors (ME) and 13 minor errors (mE). The VME were mostly related to clindamycin (64.3%) whereas meropenem and piperacillin-tazobactam constituted the major part (37.5% and 61.5%) of ME and mE respectively. Results were highly reliable for Enterobacterales and enterococci. The mean TTR ranged between 4.3 and 6.3 h and was statistically 20 min faster when applying the 7 blood drops protocol. We showed that Alfred 60/AST could give relievable results within working hours for positive blood culture which are flagged the same day between 12:00 am and 12:00 pm. Conclusion: Our study confirmed that Alfred 60/AST gives reliable AST results in a short period of time, especially for Enterobacterales and enterococci. AST could thus be easily obtained the same day of a positive blood culture. Clinical impact studies are mandatory to validate a 24/24 working.
引用
收藏
页数:8
相关论文
共 7 条
  • [1] Performance evaluation of Alfred60AST rapid susceptibility testing directly from positive blood cultures in the routine laboratory workflow
    Mantzana, Paraskevi
    Netsika, Fotini
    Arhonti, Maria
    Meletis, Georgios
    Kandilioti, Eleni
    Kiriakopoulou, Maria
    Kagkalou, Georgia
    Vasilaki, Olga
    Tychala, Areti
    Protonotariou, Efthymia
    Skoura, Lemonia
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2021, 40 (07) : 1487 - 1494
  • [2] Performance evaluation of Alfred60AST rapid susceptibility testing directly from positive blood cultures in the routine laboratory workflow
    Paraskevi Mantzana
    Fotini Netsika
    Maria Arhonti
    Georgios Meletis
    Eleni Kandilioti
    Maria Kiriakopoulou
    Georgia Kagkalou
    Olga Vasilaki
    Areti Tychala
    Efthymia Protonotariou
    Lemonia Skoura
    European Journal of Clinical Microbiology & Infectious Diseases, 2021, 40 : 1487 - 1494
  • [3] Evaluation of the Alfred AST® system for rapid antimicrobial susceptibility testing directly from positive blood cultures
    Carlos Sánchez-Carrillo
    Paula Pescador
    Rosa Ricote
    Julia Fuentes
    Carmen Losada
    Ana Candela
    Emilia Cercenado
    European Journal of Clinical Microbiology & Infectious Diseases, 2019, 38 : 1665 - 1670
  • [4] Evaluation of the Alfred AST® system for rapid antimicrobial susceptibility testing directly from positive blood cultures
    Sanchez-Carrillo, Carlos
    Pescador, Paula
    Ricote, Rosa
    Fuentes, Julia
    Losada, Carmen
    Candela, Ana
    Cercenado, Emilia
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2019, 38 (09) : 1665 - 1670
  • [5] Evaluation of a new Rapid Antimicrobial Susceptibility system for Gram-negative and Gram-positive bloodstream infections: speed and accuracy of Alfred 60AST
    Anton-Vazquez, Vanesa
    Adjepong, Samuel
    Suarez, Cristina
    Planche, Timothy
    BMC MICROBIOLOGY, 2019, 19 (01)
  • [6] Applicability and performance of EUCAST’s rapid antimicrobial susceptibility testing (RAST) on primarily sterile body fluids in blood culture bottles in laboratory routine with total lab automation
    Jasmin Kaur Jasuja
    Stefan Zimmermann
    Irene Burckhardt
    European Journal of Clinical Microbiology & Infectious Diseases, 2021, 40 : 1217 - 1225
  • [7] Applicability and performance of EUCAST's rapid antimicrobial susceptibility testing (RAST) on primarily sterile body fluids in blood culture bottles in laboratory routine with total lab automation
    Jasuja, Jasmin Kaur
    Zimmermann, Stefan
    Burckhardt, Irene
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2021, 40 (06) : 1217 - 1225