The use of a distal occlusion balloon protection device in acute coronary syndrome

被引:9
作者
Li, SSL [1 ]
Lam, CW [1 ]
So, YC [1 ]
Wong, CH [1 ]
Yiu, SF [1 ]
Ho, D [1 ]
Chan, KW [1 ]
Chan, KT [1 ]
机构
[1] Queen Elizabeth Hosp, Dept Med, Div Cardiol, Kowloon, Hong Kong, Peoples R China
关键词
coronary angioplasty; distal protection; acute coronary syndrome; acute myocardial infarction;
D O I
10.1016/S0167-5273(03)00097-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report our early experience in using the PercuSurge GuardWire Plus system as a distal protection device in patients with acute coronary syndrome and acute myocardial infarction. Forty-three patients received percutaneous coronary intervention with the GuardWire Plus system. Thirteen had unstable angina, five had non-Q myocardial infarction and 25 had ST segment elevation myocardial infarction. Forty-one target lesions were in native coronary vessels and two were in saphenous vein grafts. Total occlusion occurred in 18 patients. The mean occlusion time by the distal protective balloon was 262.8 +/- 114.1 s. Preoperatively, TIMI 0 flow was present in 18, TIMI II flow in two and TIMI III flow in 23 patients. Post-operatively, TIMI II and TIMI III flow were established in two and 41 patients, respectively. All procedures were successful and the GuardWire Plus system was successfully deployed in all but two patients. There was no procedure-related major adverse clinical event. There was no major adverse clinical event at 30 days. There was no device-related complication. We believe that the GuardWire Plus system is safe and feasible in patients with acute coronary syndrome and acute myocardial infarction. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:281 / 284
页数:4
相关论文
共 9 条
[1]   Evaluation of a balloon occlusion and aspiration system for protection from distal embolization during stenting in saphenous vein grafts [J].
Grube, E ;
Schofer, J ;
Webb, J ;
Schuler, G ;
Colombo, A ;
Sievert, H ;
Gerckens, U ;
Stone, GW .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 89 (08) :941-945
[2]  
OESTERLE S, 1998, J AM COLL CARDIO S2A, V31, P326
[3]   INCIDENCE AND TREATMENT OF NO-REFLOW AFTER PERCUTANEOUS CORONARY INTERVENTION [J].
PIANA, RN ;
PAIK, GY ;
MOSCUCCI, M ;
COHEN, DJ ;
GIBSON, CM ;
KUGELMASS, AD ;
CARROZZA, JP ;
KUNTZ, RE ;
BAIM, DS .
CIRCULATION, 1994, 89 (06) :2514-2518
[4]  
THERON J, 1990, AM J NEURORADIOL, V11, P869
[5]   Carotid artery stenosis: Treatment with protected balloon angioplasty and stent placement [J].
Theron, JG ;
Payelle, GG ;
Coskun, O ;
Huet, HF ;
Guimaraens, L .
RADIOLOGY, 1996, 201 (03) :627-636
[6]   Recognition of the importance of embolization in atherosclerotic vascular disease [J].
Topol, EJ ;
Yadav, JS .
CIRCULATION, 2000, 101 (05) :570-580
[7]   Reperfusion for ST-segment elevation myocardial infarction - An overview of current treatment options [J].
Van de Werf, F ;
Baim, DS .
CIRCULATION, 2002, 105 (24) :2813-2816
[8]   Retrieval and analysis of particulate debris after saphenous vein graft intervention [J].
Webb, JG ;
Carere, RG ;
Virmani, R ;
Baim, D ;
Teirstein, PS ;
Whitlow, P ;
McQueen, C ;
Kolodgie, FD ;
Buller, E ;
Dodek, A ;
Mancini, GBJ ;
Oesterle, S .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 34 (02) :468-475
[9]   An emboli containment system for saphenous vein graft angioplasty [J].
Webb, JG ;
Carere, RG ;
Lo, K ;
Li, C ;
McQueen, C ;
Dodek, A ;
Virmani, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) :236A-236A