Biomechanical evaluation of a new pedicle screw-based posterior dynamic stabilization device (Awesome Rod System) - a finite element analysis

被引:14
作者
Chen, Chen-Sheng [1 ]
Huang, Chang-Hung [2 ]
Shih, Shih-Liang [3 ,4 ]
机构
[1] Natl Yang Ming Univ, Dept Phys Therapy & Assist Technol, Taipei 112, Taiwan
[2] Mackay Mem Hosp, Biomed Res, New Taipei City, Taiwan
[3] Taipei City Hosp, Zhong Xing Branch, Dept Orthopaed Surg, Taipei 10341, Taiwan
[4] Natl Chengchi Univ, Inst Neurosci, Taipei 11623, Taiwan
来源
BMC MUSCULOSKELETAL DISORDERS | 2015年 / 16卷
关键词
Posterior dynamic stabilization; Adjunct to fusion; Finite element model; Lumbar spine; IN-VITRO; LUMBAR SPINE; FOLLOW-UP; FUSION; STIFFNESS; SEGMENTS; SPACER;
D O I
10.1186/s12891-015-0538-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Pedicle-screw-based posterior dynamic stabilization devices are designed to alleviate the rate of accelerated degeneration of the vertebral level adjacent to the level of spinal fusion. A new pedicle-screw-based posterior dynamic stabilization device-the Awesome Dynamic Rod System was designed with curve cuts on the rods to provide flexibility. The current study was conducted to evaluate the biomechanical properties of this new device. Methods: Finite element models were developed for the intact spine (INT), the Awesome Dynamic Rod Implanted at L4-L5 (AWE), a traditional rigid rod system implanted at L4-L5 along with an interbody cage (FUS), and the Awesome Dynamic Rod System implanted at L4-L5 along with an interbody cage as an adjunct to fusion procedures and extension of dynamic fixation to L3-L4 (AWEFUS). The models were subjected to axial loads and pure moments and evaluated by a hybrid method on range of motion (ROM)s, disc stresses, pedicle screws stresses, and facet joint contact forces. Results: FUS sustained the lowest L4-L5 ROM decrement in flexion and torsion. AWE demonstrated the lowest adjacent level ROM increment in all moments except for extension at L3-L4, and AWEFUS showed the greatest ROM increment at L2-L3. AWE demonstrated lowest adjacent segment disc stress in flexion, lateral bending and torsion at L3-L4. AWEFUS showed the highest disc stress increment in flexion, extension, and lateral bending, and the lowest disc stress decrement in torsion at L2-L3. AWE sustained greater adjacent facet joint contact forces than did FUS in extension and lateral bending at L3-L4, and AWEFUS demonstrated the greatest contact forces concentrating at L2-L3. Conclusion: The results demonstrate that the Awesome Dynamic Rod System preserved more bridged segment motion than did the traditional rigid rod fixation system except in extension. However, the Awesome Dynamic Rod System bore a greater facet joint contact force in extension. The Awesome Dynamic Rod System did protect the adjacent level of fusion segments, but led to much greater ROM, disc stresses, and facet joint contact forces increasing at the adjacent level of instrumented segments.
引用
收藏
页数:8
相关论文
共 30 条
  • [1] [Anonymous], [No title captured]
  • [2] Awasthi D, 2005, JUXTAFUSIONAL OUTCOM
  • [3] Posterior Pedicle Fixation-based Dynamic Stabilization Devices for the Treatment of Degenerative Diseases of the Lumbar Spine
    Bono, Christopher M.
    Kadaba, Murali
    Vaccaro, Alexander R.
    [J]. JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (05): : 376 - 383
  • [4] Pedicle screw-based posterior dynamic stabilizers for degenerative spine: In vitro biomechanical testing and clinical outcomes
    Chamoli, Uphar
    Diwan, Ashish D.
    Tsafnat, Naomi
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2014, 102 (09) : 3324 - 3340
  • [5] Immediate biomechanical effects of lumbar posterior dynamic stabilization above a circumferential fusion
    Cheng, Boyle C.
    Gordon, Jeff
    Cheng, Joseph
    Welch, William C.
    [J]. SPINE, 2007, 32 (23) : 2551 - 2557
  • [6] The effect of spinal destabilization and instrumentation on lumbar intradiscal pressure an in vitro biomechanical analysis
    Cunningham, BW
    Kotani, Y
    McNulty, PS
    Cappuccino, A
    McAfee, PC
    [J]. SPINE, 1997, 22 (22) : 2655 - 2663
  • [7] Posterior lumbar fusion by peek rods in degenerative spine: preliminary report on 30 cases
    De Iure, F.
    Bosco, G.
    Cappuccio, M.
    Paderni, S.
    Amendola, L.
    [J]. EUROPEAN SPINE JOURNAL, 2012, 21 : S50 - S54
  • [8] A Computational Biomechanical Investigation of Posterior Dynamic Instrumentation: Combination of Dynamic Rod and Hinged (Dynamic) Screw
    Erbulut, Deniz U.
    Kiapour, Ali
    Oktenoglu, Tunc
    Ozer, Ali F.
    Goel, Vijay K.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (05):
  • [9] Frost HM, 2004, ANGLE ORTHOD, V74, P3
  • [10] Adjacent segment degeneration in the lumbar spine
    Ghiselli, G
    Wang, JC
    Bhatia, NN
    Hsu, WK
    Dawson, EG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (07) : 1497 - 1503