Evaluation of 3 rapid influenza diagnostic tests during the 2012-2013 epidemic: influences of subtype and viral load

被引:15
作者
Busson, Laurent [1 ]
Hallin, Marie [2 ]
Thomas, Isabelle [3 ]
De Foor, Marc [1 ]
Vandenberg, Olivier [1 ,4 ]
机构
[1] Dept Microbiol, Iris Lab, Brussels, Belgium
[2] Dept Mol Diag, Iris Lab, Brussels, Belgium
[3] Natl Influenza Ctr, Sci Inst Publ Hlth, Brussels, Belgium
[4] Univ Libre Bruxelles, Sch Publ Hlth, Infect Dis Epidemiol Unit, Brussels, Belgium
关键词
Influenza; Binax; Directigen; bioNexia; RIDT; Immunochromatography; PERFORMANCE;
D O I
10.1016/j.diagmicrobio.2014.08.015
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
This article evaluates the performance of 3 rapid influenza diagnostic tests (RIDTs), in correlation with the influenza subtypes and the viral load. A total of 236 samples were prospectively analyzed with BinaxNOW Influenza A/B, Directigen EZ Flu A and B, and bioNexia Influenza A+B. The results were compared to cell cultures and real-time polymerase chain reaction. Positive samples were further subtyped. Thirty-seven samples were positive for influenza A, and 57, for influenza B. For A(H1N1), the sensitivities were 71.42% for BinaxNOW, 78.57% for Directigen, and 67.85% for bioNexia. Eight samples were positive for A(H3N2), and only the bioNexia test had 1 false-negative result. Lowest sensitivities were observed for influenza B/Yamagata, (56.86% for BinaxNOW and Directigen and 39.21% for bioNexia). The 3 evaluated RIDTs were more efficient at detecting influenza A (H3N2) than for A(H1N1) and B/Yamagata. Highest viral loads in the samples were associated with better rate of detection. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:287 / 291
页数:5
相关论文
共 18 条
  • [1] Influenza virus shedding-Excretion patterns and effects of antiviral treatment
    Aoki, Fred Y.
    Boivin, Guy
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2009, 44 (04) : 255 - 261
  • [2] Beck Eric, 2012, Morbidity and Mortality Weekly Report, V61, P873
  • [3] Centers for Disease Control and Prevention (CDC), 2009, CDC PROT REAL TIM PC
  • [4] Accuracy of Rapid Influenza Diagnostic Tests A Meta-analysis
    Chartrand, Caroline
    Leeflang, Mariska M. G.
    Minion, Jessica
    Brewer, Timothy
    Pai, Madhukar
    [J]. ANNALS OF INTERNAL MEDICINE, 2012, 156 (07) : 500 - U80
  • [5] Daniels R, 2013, INFLUENZA VIRUS CHAR
  • [6] ECDC, 2005, INFL CAS DEF
  • [7] COMPARISON OF DIFFERENT TISSUE-CULTURES FOR ISOLATION AND QUANTITATION OF INFLUENZA AND PARA-INFLUENZA VIRUSES
    FRANK, AL
    COUCH, RB
    GRIFFIS, CA
    BAXTER, BD
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1979, 10 (01) : 32 - 36
  • [8] Ginocchio C, 2011, MANUAL CLIN MICROBIO, P1289
  • [9] Hungnes O, 2011, WHO INFORM MOL DIAGN
  • [10] Performance of six influenza rapid tests in detecting human influenza in clinical specimens
    Hurt, Aeron C.
    Alexander, Robert
    Hibbert, Jan
    Deed, Nicola
    Barr, Ian G.
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2007, 39 (02) : 132 - 135