Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure

被引:59
作者
Mase, Michela
Mattei, Walter
Cucino, Roberta
Faes, Luca
Nollo, Giandomenico
机构
[1] Univ Trent, Dept Phys, I-38123 Trento, Italy
[2] Univ Trent, BIOtech, I-38123 Trento, Italy
关键词
Noninvasive measurement; Blood pressure monitoring; Pulse transit time; Validation; Accuracy; HYPERTENSION-INTERNATIONAL-PROTOCOL; PULSE-WAVE VELOCITY; MEASURING DEVICES; EUROPEAN-SOCIETY; TRANSIT-TIME; ARRIVAL-TIME; VALIDATION; MARKER; ADULTS;
D O I
10.1016/j.jelectrocard.2010.11.019
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We validated a prototype cuff-free device for noninvasive estimation of blood pressure (BP). The system assumed a linear relation between BP values and the inverse of arterial blood pulse transit time, measured as time interval between the R wave on the electrocardiograph and the onset of the peripheral pulse wave on a finger plethysmogram. Thirty-three healthy subjects were analyzed at rest and during increasing stress exercise. To estimate subject-specific linear model parameters, the system was calibrated ad personam with reference to BP measures obtained by a cuff sphygmomanometer. High correlation values (R-2 = 0.89 and 0.78 for systolic and diastolic BP, respectively) and differences consistent with clinical requirements (mean discrepancy of -0.058 and -0.25 mm Hg; 95% confidence interval of -13.0 to +12.9 mm Hg and -11.3 to +10.8 mm Hg, for systolic and diastolic BP, respectively) were observed between device and reference measurements. Calibration parameter stability and accuracy level were confirmed in a midterm evaluation, 30 days after calibration. These findings suggest the suitability of the device for noninvasive BP monitoring and its potentiality for clinical applications. Improvements can be achieved by further investigation of the calibration procedure and sensor placement. (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:201 / 207
页数:7
相关论文
共 15 条
[1]   Arterial blood pressure measurement and pulse wave analysis-their role in enhancing cardiovascular assessment [J].
Avolio, Alberto P. ;
Butlin, Mark ;
Walsh, Andrew .
PHYSIOLOGICAL MEASUREMENT, 2010, 31 (01) :R1-R47
[2]   Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration [J].
Chen, W ;
Kobayashi, T ;
Ichikawa, S ;
Takeuchi, Y ;
Togawa, T .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2000, 38 (05) :569-574
[3]   PULSE-WAVE VELOCITY AS A MEASURE OF BLOOD-PRESSURE CHANGE [J].
GRIBBIN, B ;
STEPTOE, A ;
SLEIGHT, P .
PSYCHOPHYSIOLOGY, 1976, 13 (01) :86-90
[4]   An evaluation of a new continuous blood pressure monitoring system in critically ill patients [J].
Heard, SO ;
Lisbon, A ;
Toth, I ;
Ramasubramanian, R .
JOURNAL OF CLINICAL ANESTHESIA, 2000, 12 (07) :509-518
[5]  
Ngim CA, 1999, ACTA CARDIOL, V54, P277
[6]   SPONTANEOUS BEAT-TO-BEAT VARIABILITY OF THE VENTRICULAR REPOLARIZATION DURATION [J].
NOLLO, G ;
SPERANZA, G ;
GRASSO, R ;
BONAMINI, R ;
MANGIARDI, L ;
ANTOLINI, R .
JOURNAL OF ELECTROCARDIOLOGY, 1992, 25 (01) :9-17
[7]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[8]   European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults [J].
O'Brien, Eoin ;
Atkins, Neil ;
Stergiou, George ;
Karpettas, Nikos ;
Parati, Gianfranco ;
Asmar, Roland ;
Imai, Yutaka ;
Wang, Jiguang ;
Mengden, Thomas ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :23-38
[9]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[10]   Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure [J].
Payne, RA ;
Symeonides, CN ;
Webb, DJ ;
Maxwell, SRJ .
JOURNAL OF APPLIED PHYSIOLOGY, 2006, 100 (01) :136-141