Fluoxetine and norfluoxetine analysis by direct injection of human plasma in a column switching liquid chromatographic system

被引:24
作者
Santos-Neto, Alvaro J. [1 ]
Fernandes, Christian [1 ]
Rodrigues, Jose C. [1 ]
Lancas, Fernando M. [1 ]
机构
[1] Univ Sao Paulo, Inst Quim Sao Carlos, BR-13566590 Sao Carlos, SP, Brazil
关键词
column switching; direct injection; fluoxetine; online sample preparation; restricted access media;
D O I
10.1002/jssc.200700153
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A column switching LC method is presented for the analysis of fluoxetine (FLU) and norfluoxetine (NFLU) by direct injection of human plasma using a lab-made restricted access media (RAM) column. A RAM-BSA-octadecyl silica (C-18) column (40 min x 4.6 mm, 10 mu m) is evaluated in both backflush and foreflush elution modes and coupled with a C-18 lab-made (50 mm x 4.6 mm, 3 pm) analytical column in order to perform online sample preparation. Direct injection of 100 mu L, of plasma samples is possible with the developed approach. In addition, reduction of sample handling is obtained when compared with traditional liquid-liquid extraction (LLE) and SPE. The total analysis time is around 20 min. A LOQ of 15 ng/mL is achieved in a concentration range of 15-500 ng/mL, allowing the therapeutic drug monitoring of clinical samples. The precision values achieved are lower than 15% for all the evaluated points with adequate recovery and accuracy. Furthermore, no matrix interferences are found in the analysis and the proposed method shows to be an adequate alternative for analysis of FLU in plasma.
引用
收藏
页码:78 / 85
页数:8
相关论文
共 50 条