Examination of failed ex vivo metal-on-metal metatarsophalangeal prosthesis and comparison with theoretically determined lubrication regimes

被引:38
作者
Joyce, T. J. [1 ]
机构
[1] Newcastle Univ, Sch Mech & Syst Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
metatarsophalangeal; metal-on-metal; lubrication regimes; diamond-like carbon;
D O I
10.1016/j.wear.2006.11.045
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Replacement of the first metatarsophalangeal (NITP) joint is a relatively uncommon procedure compared with hip and knee arthroplasty. A cobalt chrome-on -cobalt chrome MTP prosthesis, which had a diamond-like carbon (DLC) coating applied to its articulating faces, was obtained for ex vivo analysis. By modelling the ball and socket implant as an equivalent ball-on-plane model and employing elastohydrodynamic theory, the predicted lubrication regimes applicable to this implant design were determined. These calculations were undertaken for a 10-1500 N range of loading values and a 0-30 mm/s range of entraining velocities, for both worn and unworn situations. Calculations showed that the implant would almost always operate in the boundary lubrication regime. The presence of scratches on the articulating faces of the ex vivo sample further implied boundary lubrication. The DLC coating had been removed from the entire face of the phalangeal component and from most of the face of the metatarsal component. From the latter it appeared as if the coating had been scratched and then flaked away parallel to the scratches. In turn this suggested a corrosion based failure of the interface between the DLC coating and the cobalt chrome subsurface. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1050 / 1054
页数:5
相关论文
共 29 条
[1]  
Bankes MJK, 1999, FOOT ANKLE SURG, V5, P235
[2]  
Bommireddy RS, 2003, FOOT, V13, P151
[3]   THE EFFECT OF EXPOSURE TO BIOLOGICAL-FLUIDS ON THE SPALLATION RESISTANCE OF DIAMOND-LIKE CARBON COATINGS ON METALLIC SUBSTRATES [J].
CHANDRA, L ;
ALLEN, M ;
BUTTER, R ;
RUSHTON, N ;
LETTINGTON, AH ;
CLYNE, TW .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1995, 6 (10) :581-589
[4]   Metal-on-metal total hip replacement - What does the literature say? [J].
Dumbleton, JH ;
Manley, MT .
JOURNAL OF ARTHROPLASTY, 2005, 20 (02) :174-188
[5]  
Ess P, 2002, Scand J Surg, V91, P202
[6]   USE OF A HINGED SILICONE PROSTHESIS FOR REPLACEMENT ARTHROPLASTY OF THE 1ST METATARSOPHALANGEAL JOINT [J].
GRANBERRY, WM ;
NOBLE, PC ;
BISHOP, JO ;
TULLOS, HS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (10) :1453-1459
[7]   ELASTOHYDRODYNAMIC LUBRICATION OF ELLIPTICAL CONTACTS FOR MATERIALS OF LOW ELASTIC-MODULUS I - FULLY FLOODED CONJUNCTION [J].
HAMROCK, BJ ;
DOWSON, D .
JOURNAL OF LUBRICATION TECHNOLOGY-TRANSACTIONS OF THE ASME, 1978, 100 (02) :236-245
[8]  
Hanyu T, 2001, J Orthop Sci, V6, P141, DOI 10.1007/s007760100062
[9]  
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46
[10]   An overview on the tribological behavior of diamond-like carbon in technical and medical applications [J].
Hauert, R .
TRIBOLOGY INTERNATIONAL, 2004, 37 (11-12) :991-1003