A suction-type microfluidic immunosensing chip for rapid detection of the dengue virus

被引:22
作者
Weng, Chen-Hsun [4 ]
Huang, Tze-Bin [2 ]
Huang, Chih-Chia [3 ]
Yeh, Chen-Sheng [3 ]
Lei, Huan-Yao [4 ]
Lee, Gwo-Bin [1 ]
机构
[1] Natl Tsing Hua Univ, Dept Power Mech Engn, Hsinchu 30013, Taiwan
[2] Natl Cheng Kung Univ, Dept Engn Sci, Tainan 701, Taiwan
[3] Natl Cheng Kung Univ, Dept Chem, Tainan 701, Taiwan
[4] Natl Cheng Kung Univ, Dept Microbiol & Immunol, Tainan 701, Taiwan
关键词
Microfluidics; Suction; Immunosensing; ELISA; Dengue virus; Surface modification; MICROCHIP ELECTROPHORESIS; SURFACE MODIFICATION; PROTEIN NS1; ASSAY; CHANNELS; ANTIBODY; ELISA; FLOW;
D O I
10.1007/s10544-011-9529-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The enzyme-linked immunosorbent assay (ELISA) is widely used in medical diagnostics. In order to reduce the diagnosis time and to lower the consumption of sample/reagents in an ELISA assay, a suction-type, automatic, pneumatically-driven microfluidic chip has been designed and fabricated in this study. The microfluidic chip integrates a multi-functional micro-transport/mixing unit, for transporting metering and mixing of samples and reagents in order to automatically perform the entire ELISA protocol. A new surface modification has been adopted which allows for a high processing capacity. The detection sensitivity for the dengue virus is found to be 10(1) PFU/ml, which is much better than a conventional ELISA assay (10(3) PFU/ml). The entire assay time is only 30 min, which is much faster than with 96-well microtiter plates (4 h). The consumed sample and reagent volume is only 12 mu l, which is less than a conventional assay (100 mu l). The development of this microfluidic chip may be promising for other immunosensing applications.
引用
收藏
页码:585 / 595
页数:11
相关论文
共 24 条
[1]   Microfluidic immunosensor systems [J].
Bange, A ;
Halsall, HB ;
Heineman, WR .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (12) :2488-2503
[2]   Surface modification in microchip electrophoresis [J].
Belder, D ;
Ludwig, M .
ELECTROPHORESIS, 2003, 24 (21) :3595-3606
[3]   Rapid detection and quantification of RNA of Ebola and Marburg viruses, Lassa virus, Crimean-Congo hemorrhagic fever virus, Rift Valley fever virus, Dengue virus, and Yellow fever virus by real-time reverse transcription-PCR [J].
Drosten, C ;
Göttig, S ;
Schilling, S ;
Asper, M ;
Panning, M ;
Schmitz, H ;
Günther, S .
JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (07) :2323-2330
[4]   Development and characterization of an ELISA assay in PDMS microfluidic channels [J].
Eteshola, E ;
Leckband, D .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 72 (02) :129-133
[5]   RAPID IDENTIFICATION OF DENGUE VIRUS ISOLATES BY USING MONOCLONAL-ANTIBODIES IN AN INDIRECT IMMUNOFLUORESCENCE ASSAY [J].
HENCHAL, EA ;
MCCOWN, JM ;
SEGUIN, MC ;
GENTRY, MK ;
BRANDT, WE .
AMERICAN JOURNAL OF TROPICAL MEDICINE AND HYGIENE, 1983, 32 (01) :164-169
[6]   Pneumatic micropumps with serially connected actuation chambers [J].
Huang, Chun-Wei ;
Huang, Song-Bin ;
Lee, Gwo-Bin .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2006, 16 (11) :2265-2272
[7]   The dual-specific binding of dengue virus and target cells for the antibody-dependent enhancement of dengue virus infection [J].
Huang, Kao-Jean ;
Yang, Yu-Ching ;
Lin, Yee-Shin ;
Huang, Jyh-Hsiung ;
Liu, Hsiao-Sheng ;
Yeh, Trai-Ming ;
Chen, Shun-Hua ;
Liu, Ching-Chuan ;
Lei, Huan-Yao .
JOURNAL OF IMMUNOLOGY, 2006, 176 (05) :2825-2832
[8]   STUDY OF ANTI-DENGUE NS1 ANTIBODY BY WESTERN-BLOT [J].
KUNO, G ;
VORNDAM, AV ;
GUBLER, DJ ;
GOMEZ, I .
JOURNAL OF MEDICAL VIROLOGY, 1990, 32 (02) :102-108
[9]   An integrated microfluidic system using magnetic beads for virus detection [J].
Lee, Wan-Chi ;
Lien, Kang-Yi ;
Lee, Gwo-Bin ;
Lei, Huan-Yao .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2008, 60 (01) :51-58
[10]   An integrated microfluidic system for rapid diagnosis of dengue virus infection [J].
Lee, Yu-Fang ;
Lien, Kang-Yi ;
Lei, Huan-Yao ;
Lee, Gwo-Bin .
BIOSENSORS & BIOELECTRONICS, 2009, 25 (04) :745-752