Construction and validation of a low cost paediatric pelvis phantom

被引:20
作者
Ali, Ali Mohammed [1 ]
Hogg, Peter [1 ]
Johansen, Safora [2 ,3 ]
England, Andrew [1 ]
机构
[1] Univ Salford, Sch Hlth Sci, Salford M6 6PU, Lancs, England
[2] Oslo Metropolitan Univ, Fac Hlth Sci, Oslo, Norway
[3] Oslo Univ Hosp, Radiumhosp, Dept Oncol, Div Canc Med Surg & Transplantat, Oslo, Norway
关键词
Pelvis phantom; Low-cost; Dose optimisation; Validation; Development; IMAGE QUALITY; COMPUTED RADIOGRAPHY; CHEST; OPTIMIZATION; DOSIMETRY; EXPOSURE; SKULL;
D O I
10.1016/j.ejrad.2018.09.015
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Imaging phantoms can be cost prohibitive, therefore a need exists to produce low cost alternatives which are fit for purpose. This paper describes the development and validation of a low cost paediatric pelvis phantom based on the anatomy of a 5-year-old child. Methods: Tissue equivalent materials representing paediatric bone (Plaster of Paris; PoP) and soft tissue (Poly methyl methacrylate; PMMA) were used. PMMA was machined to match the bony anatomy identified from a CT scan of a 5-year-old child and cavities were created for infusing the PoP. Phantom validation comprised physical and visual measures. Physical included CT density comparison between a CT scan of a 5-year-old child and the phantom and Signal to Noise Ratio (SNR) comparative analysis of anteroposterior phantom X-ray images against a commercial anthropomorphic phantom. Visual analysis using a psychometric image quality scale (face validity). Results: CT density, the percentage difference between cortical bone, soft tissue and their equivalent tissue substitutes were -4.7 to -4.1% and -23.4%, respectively. For SNR, (mAs response) there was a strong positive correlation between the two phantoms (r > 0.95 for all kVps). For kVp response, there was a strong positive correlation between 1 and 8 mAs (r = 0.85), this then decreased as mAs increased (r = -0.21 at 20 mAs). Psychometric scale results produced a Cronbach's Alpha of almost 0.8. Conclusions: Physical and visual measures suggest our low-cost phantom has suitable anatomical characteristics for X-ray imaging. Our phantom could have utility in dose and image quality optimisation studies.
引用
收藏
页码:84 / 91
页数:8
相关论文
共 42 条
[1]  
[Anonymous], 1989, TISS SUBST RAD DOS M
[2]  
Bourne R, 2010, FUNDAMENTALS OF DIGITAL IMAGING IN MEDICINE, P173, DOI 10.1007/978-1-84882-087-6_8
[3]  
Bower M. W., 1997, PHYS ANTHROPOMORPHIC
[4]   Dose optimization in pelvic radiography by air gap method on CR and DR systems - A phantom study [J].
Chan, C. T. P. ;
Fung, K. K. L. .
RADIOGRAPHY, 2015, 21 (03) :214-223
[5]   BEAM QUALITY INDEPENDENT ATTENUATION PHANTOM FOR ESTIMATING PATIENT EXPOSURE FROM X-RAY AUTOMATIC EXPOSURE CONTROLLED CHEST EXAMINATIONS [J].
CONWAY, BJ ;
BUTLER, PF ;
DUFF, JE ;
FEWELL, TR ;
GROSS, RE ;
JENNINGS, RJ ;
KOUSTENIS, GH ;
MCCROHAN, JL ;
RUETER, FG ;
SHOWALTER, CK .
MEDICAL PHYSICS, 1984, 11 (06) :827-832
[6]   Investigation of dose reduction in neonatal radiography using specially designed phantoms and LiF:Mg,Cu,P TLDs [J].
Duggan, L ;
Warren-Forward, H ;
Smith, T ;
Kron, T .
BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (904) :232-237
[7]   Construction of pediatric homogeneous phantoms for optimization of chest and skull radiographs [J].
Fattori Alves, Allan Felipe ;
de Arruda Miranda, Jose Ricardo ;
Bacchim Neto, Fernando Antonio ;
Duarte, Sergio Barbosa ;
de Pina, Diana Rodrigues .
EUROPEAN JOURNAL OF RADIOLOGY, 2015, 84 (08) :1579-1585
[8]  
Fauber T., 2013, Radiographic imaging and exposure, V4th
[9]   On the use of pediatric phantoms in the dose evaluation during computed tomography (CT) thorax examinations [J].
Giacco, G ;
Cannata, V ;
Furetta, C ;
Santopietro, F ;
Fariello, G .
MEDICAL PHYSICS, 2001, 28 (02) :199-204
[10]  
Goitein M., 1982, PROGR MED RAD PHYS, V1, P195