Comparison of two time-resolved detectors for diffuse optical tomography: photomultiplier tube-time-correlated single photon counting and multi-channel streak camera

被引:8
|
作者
Poulet, P [1 ]
Zint, CV [1 ]
Torregrossa, M [1 ]
Uhring, W [1 ]
Cunin, B [1 ]
机构
[1] ULP, UMR 7004 CNRS, Inst Biol Phys, F-67085 Strasbourg, France
来源
OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE V | 2003年 / 4955卷
关键词
diffuse optical tomography; time-resolved experiments; instrumentation; streak camera; time-correlated single photon counting;
D O I
10.1117/12.476841
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Two experimental setups for time-resolved diffuse optical tomography (DOT) are described. The first involves a titanium-sapphire laser and a streak camera in a synchroscan mode. A multiple-arm light guide allows to measure simultaneously the temporal profiles of photons re-emitted at different boundary sites of the objects studied. The second one uses picosecond laser diodes and a multiple-anode micro-channel plate photomultiplier tube followed by parallel time-correlated single photon counting channels. The two instruments are described and their main specifications compared: instrument response and time resolution, sweep time, temporal stability, sensitivity and its inter-channel variation, cross-talk between adjacent channels. We will also discuss detector-specific data analysis prior to image reconstruction. Both instruments have been tested with phantoms simulating tissues and the absorption and reduced scattering images obtained have been compared.
引用
收藏
页码:154 / 163
页数:10
相关论文
共 50 条
  • [1] Time-resolved fluorescence measurements for diffuse optical tomography using ultrafast time-correlated single photon counting
    Berube-Lauziere, Yves
    Robichaud, Vincent
    ADVANCED PHOTON COUNTING TECHNIQUES, 2006, 6372
  • [2] USE OF STREAK CAMERA FOR TIME-RESOLVED PHOTON-COUNTING FLUOROMETRY
    DAVIS, LM
    PARIGGER, C
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (01) : 85 - 90
  • [3] Diffuse optical tomography by using time-resolved single pixel camera
    Farina, A.
    Lepore, M.
    Di Sieno, L.
    Dalla Mora, A.
    Ducros, N.
    Pifferi, A.
    Valentini, G.
    Arridge, S.
    D'Andrea, C.
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE XI, 2015, 9319
  • [4] Optical tomography of the breast using a multi-channel time-resolved imager
    Yates, T
    Hebden, JC
    Gibson, A
    Everdell, N
    Arridge, SR
    Douek, M
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (11): : 2503 - 2517
  • [5] Multi-dimensional time-correlated single photon counting applied to diffuse optical tomography
    Becker, W
    Bergmann, A
    Gibson, A
    Everdell, N
    Jennions, D
    Schweige, M
    Arridge, SR
    Hebden, JC
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE VI, 2005, 5693 : 34 - 42
  • [6] Time-resolved non-contact fluorescence diffuse optical tomography measurements with ultra-fast time-correlated single photon counting avalanche photodiodes
    Berube-Lauziere, Yves
    Robichaud, Vincent
    Lapointe, Eric
    DIFFUSE OPTICAL IMAGING OF TISSUE, 2007, 6629
  • [7] PICOSECOND TIME-RESOLVED PHOTOELECTRIC CORRELATION-MEASUREMENT WITH A PHOTON-COUNTING STREAK CAMERA
    UEDA, M
    KUWATA, M
    NAGASAWA, N
    URAKAMI, T
    TAKIGUCHI, Y
    TSUCHIYA, Y
    OPTICS COMMUNICATIONS, 1988, 65 (05) : 315 - 318
  • [8] Diffuse Optical Tomography using Time-resolved Photon Path Distribution
    Yukio Ueda
    Kazuyoshi Ohta
    Yutaka Yamashita
    Optical Review, 2005, 12 : 334 - 337
  • [9] Image quality in time-resolved laser transillumination by time-correlated single photon counting
    Daniele, A.
    Mormile, M.
    Esposito, R.
    Lepore, A.
    Indovina, P. L.
    BIOPHOTONICS AND NEW THERAPY FRONTIERS, 2006, 6191
  • [10] Time-resolved two-photon excitation fluorescence spectroscopy using a streak camera
    Liu, Lixin
    Qu, Junle
    Lin, Ziyang
    Guo, Baciping
    Liu, Wenqing
    Niu, Hanben
    2007 IEEE/ICME INTERNATIONAL CONFERENCE ON COMPLEX MEDICAL ENGINEERING, VOLS 1-4, 2007, : 964 - 968