Unexpected abnormal coagulation test results in a 2-year-old child: A case report

被引:2
作者
Radovanovic, Patricija Bankovic [1 ]
Mikulcic, Tanja Zivkovic [1 ]
Medica, Jasmina Simovic [1 ]
机构
[1] Gen Hosp Pula, Dept Transfus Med, Pula, Croatia
关键词
case report; preanalytical phase; sample collection; coagulation; INCORRECT ORDER; BLOOD-SAMPLES; DRAW; BIOCHEMISTRY; PHLEBOTOMY; MYTH;
D O I
10.11613/BM.2020.011002
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Rejection of the sample with repeated blood withdrawal is always an unwanted consequence of sample nonconformity and preanalytical errors, especially in the most vulnerable population - children. Here is presented a case with unexpected abnormal coagulation test results in a 2-year-old child with no previously documented coagulation disorder. Child is planned for tympanostomy tubes removal under the anaesthesia driven procedure, and preoperative coagulation tests revealed prolonged prothrombin time, activated partial thromboplastin time and thrombin time, with fibrinogen and antithrombin within reference intervals. From the anamnestic and clinical data, congenital coagulation disorder was excluded, and with further investigation, sample mismatch, clot presence and accidental ingestion of oral anticoagulant, heparin contamination or vitamin K deficiency were excluded too. Due to suspected EDTA carryover during blood sampling another sample was taken the same day and all tests were performed again. The results for all tests were within reference intervals confirming EDTA effect on falsely prolongation of the coagulation times in the first sample. This case can serve as alert to avoid unnecessary loss in terms of blood withdrawal repetitions and discomfort of the patients and their relatives, tests repeating, prolonging medical procedures, and probably delaying diagnosis or proper medical treatment. It is the responsibility of the laboratory specialists to continuously educate laboratory staff and other phlebotomists on the correct blood collection as well as on its importance for the patient's safety.
引用
收藏
页数:6
相关论文
共 22 条
  • [1] Clinical biochemistry laboratory rejection rates due to various types of preanalytical errors
    Atay, Aysenur
    Demir, Leyla
    Cuhadar, Serap
    Saglam, Gulcan
    Unal, Hulya
    Aksun, Saliha
    Arslan, Banu
    Ozkan, Asuman
    Sutcu, Recep
    [J]. BIOCHEMIA MEDICA, 2014, 24 (03) : 376 - 382
  • [2] Burtis C., 2008, TIETZ Fundamentals of Clinical Chemistry, V6th
  • [3] Relevance of EDTA carryover during blood collection
    Cadamuro, Janne
    Felder, Thomas Klaus
    Oberkofler, Hannes
    Mrazek, Cornelia
    Wiedemann, Helmut
    Haschke-Becher, Elisabeth
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (08) : 1271 - 1278
  • [4] CLSI, 2007, H3A6 CLSI
  • [5] Incorrect order of draw of blood samples does not cause potassium EDTA sample contamination
    Cornes, M. P.
    Sulaiman, R. A.
    Whitehead, S. J.
    Othonos, N.
    Ford, C.
    Gama, R.
    [J]. BRITISH JOURNAL OF BIOMEDICAL SCIENCE, 2012, 69 (03) : 136 - 138
  • [6] Order of blood draw: Opinion Paper by the European Federation for Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase (WG-PRE)
    Cornes, Michael
    van Dongen-Lases, Edmee
    Grankvist, Kjell
    Ibarz, Mercedes
    Kristensen, Gunn
    Lippi, Giuseppe
    Nybo, Mads
    Simundic, Ana-Maria
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (01) : 27 - 31
  • [7] The order of draw, myth or science
    Cornes, Michael P.
    Ford, Clare
    Gama, Rousseau
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (12) : E285 - E285
  • [8] Underestimation of plasma level of factor V coagulant activity and fibrinogen concentration together with prolonged prothrombin time, activated partial thromboplastin time and thrombin time can result from pre-analytical very low calcium level in citrated sample tube
    d'Audigier, C.
    Delassasseigne, C.
    Robert, A.
    Eschwege, V.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 (01) : 50 - 53
  • [9] Pre-analytical Variables in Coagulation Testing Associated With Diagnostic Errors in Hemostasis
    Favaloro, Emmanuel J.
    Funk, Dorothy M.
    Lippi, Giuseppe
    [J]. LABMEDICINE, 2012, 43 (02): : 54 - 60
  • [10] Funk D.M., 2008, Collection, transport, and processing of blood specimens for testing plasma-based coagulation assays and molecular hemostasis assays: approved guideline, V5th