Wear of metal-on-metal hip bearings: metallurgical considerations after hip simulator studies

被引:10
作者
Affatato, Saverio [1 ]
Traina, Francesco [1 ]
Ruggeri, Oddone [2 ]
Toni, Aldo [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, IT-40136 Bologna, Italy
[2] Fac Chim Ind, Ist Met, Bologna, Italy
关键词
Cast alloy; High carbon; Ions release; Microstructure; Wear; Wrought components; JOINT SIMULATOR; FEMORAL-HEAD; IMPLANTS; LUBRICATION; ARTHROPLASTY; REPLACEMENTS; PROSTHESES; DIAMETER; ALLOYS; MICROSTRUCTURE;
D O I
10.5301/ijao.5000065
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Metal-on-metal hip-bearing components with different percentages of carbon content (low and high carbon) were tested in 6 different configurations using a hip joint simulator. The aim of this study was to characterize metallurgical and tribological events occurring at the articulating surfaces of these articulations. Also, ion release was evaluated and correlated with wear. After the test, for the high-carbon components, carbides were observed below the matrix surface. In the low-carbon content components, most carbides were "carbide-free", while a minority were worn below the matrix surface with increased test cycles. In the cast alloy components, some carbides were pulled out resulting in micropits. Scanning microscope electron characterization of the tested specimens showed scratches and holes. The surface showed a dominant severe wear mechanism due to third-body particles. A greater amount of ions was released in the lubricant used during the wear test for the smaller diameter compared with the bigger one. This study showed that the metallurgical and tribological events taking place at the articulating surfaces of metal-metal hip implants are numerous and complex. The surface morphology after the test showed the effect of more critical working conditions with smaller diameters.
引用
收藏
页码:1155 / 1164
页数:10
相关论文
共 54 条
  • [1] A. ASTM, 2001, E301 A ASTM
  • [2] Predictive role of the Λ ratio in the evaluation of metal-on-metal total hip replacement
    Affatato, S.
    Spinelli, M.
    Zavalloni, M.
    Leardini, W.
    Viceconti, M.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) : 617 - 628
  • [3] Larger diameter bearings reduce wear in metal-on-metal hlp hnplants
    Affatato, S.
    Leardini, W.
    Jedenmalm, A.
    Ruggeri, O.
    Toni, A.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (456) : 153 - 158
  • [4] Amstutz H.C., 1998, HIP Int, V8, P187, DOI [10.1177/112070009800800401, DOI 10.1177/112070009800800401]
  • [5] Amstutz HC, 1996, CLIN ORTHOP RELAT R, pS11
  • [6] CONTACT AND RUBBING OF FLAT SURFACES
    ARCHARD, JF
    [J]. JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) : 981 - 988
  • [7] ARMITAGE P, 1994, STAT METHODS MED RES, P237
  • [8] Asphahani AI, 1987, CORROSION, P658
  • [9] Black J., 1996, CORR, V329S, P244, DOI DOI 10.1097/00003086-199608001-00021
  • [10] Biological effects of clinically relevant wear particles from metal-on-metal hip prostheses
    Brown, C
    Fisher, J
    Ingham, E
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 355 - 369