Imaging of Arthroplasties: Improved Image Quality and Lesion Detection With Iterative Metal Artifact Reduction, a New CT Metal Artifact Reduction Technique

被引:40
|
作者
Subhas, Naveen [1 ]
Polster, Joshua M. [1 ]
Obuchowski, Nancy A. [2 ]
Primak, Andrew N. [3 ]
Dong, Frank F. [4 ]
Herts, Brian R. [1 ]
Iannotti, Joseph P. [5 ]
机构
[1] Cleveland Clin, Dept Radiol, 9500 Euclid Ave,Ste A21, Cleveland, OH 44195 USA
[2] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44106 USA
[3] Siemens Healthcare, Malvern, PA USA
[4] Cleveland Clin, Sect Med Phys, Cleveland, OH 44106 USA
[5] Cleveland Clin, Dept Orthoped Surg, Cleveland, OH 44106 USA
关键词
artifacts; CT; metallic hardware; techniques; COMPUTED-TOMOGRAPHY; HIP PROSTHESES; RECONSTRUCTIONS; VISUALIZATION;
D O I
10.2214/AJR.15.15850
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare iterative metal artifact reduction (iMAR), a new single-energy metal artifact reduction technique, with filtered back projection (FBP) in terms of attenuation values, qualitative image quality, and streak artifacts near shoulder and hip arthroplasties and observer ability with these techniques to detect pathologic lesions near an arthroplasty in a phantom model. MATERIALS AND METHODS. Preoperative and postoperative CT scans of 40 shoulder and 21 hip arthroplasties were reviewed. All postoperative scans were obtained using the same technique (140 kVp, 300 quality reference mAs, 128 x 0.6 mm detector collimation) on one of three CT scanners and reconstructed with FBP and iMAR. The attenuation differences in bones and soft tissues between preoperative and postoperative scans at the same location were compared; image quality and streak artifact for both reconstructions were qualitatively graded by two blinded readers. Observer ability and confidence to detect lesions near an arthroplasty in a phantom model were graded. RESULTS. For both readers, iMAR had more accurate attenuation values (p < 0.001), qualitatively better image quality (p < 0.001), and less streak artifact (p < 0.001) in all locations near arthroplasties compared with FBP. Both readers detected more lesions (p <= 0.04) with higher confidence (p <= 0.01) with iMAR than with FBP in the phantom model. CONCLUSION. The iMAR technique provided more accurate attenuation values, better image quality, and less streak artifact near hip and shoulder arthroplasties than FBP; iMAR also increased observer ability and confidence to detect pathologic lesions near arthroplasties in a phantom model.
引用
收藏
页码:378 / 385
页数:8
相关论文
共 50 条
  • [1] Iterative deblurring for CT metal artifact reduction
    Wang, G
    Snyder, DL
    OSullivan, JA
    Vannier, MW
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) : 657 - 664
  • [2] Iterative metal artifact reduction: Evaluation and optimization of technique
    Naveen Subhas
    Andrew N. Primak
    Nancy A. Obuchowski
    Amit Gupta
    Joshua M. Polster
    Andreas Krauss
    Joseph P. Iannotti
    Skeletal Radiology, 2014, 43 : 1729 - 1735
  • [3] Iterative metal artifact reduction: Evaluation and optimization of technique
    Subhas, Naveen
    Primak, Andrew N.
    Obuchowski, Nancy A.
    Gupta, Amit
    Polster, Joshua M.
    Krauss, Andreas
    Iannotti, Joseph P.
    SKELETAL RADIOLOGY, 2014, 43 (12) : 1729 - 1735
  • [4] The λ-MLEM algorithm:: An iterative reconstruction technique for metal artifact reduction in CT images
    Oehler, May
    Buzug, Thorsten M.
    ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 42 - +
  • [5] Iterative CT reconstruction of real data with metal artifact reduction
    Hamelin, Benoit
    Goussard, Yves
    Gendron, David
    Dussault, Jean-Pierre
    Cloutier, Guy
    Beaudoin, Gilles
    Soulez, Gilles
    2008 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1-4, 2008, : 1453 - +
  • [6] A new metal artifact reduction algorithm based on a deteriorated CT image
    Kano, Toru
    Koseki, Michihiko
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (06) : 901 - 912
  • [7] CT Metal Artifact Reduction in the Spine: Can an Iterative Reconstruction Technique Improve Visualization?
    Kotsenas, A. L.
    Michalak, G. J.
    DeLone, D. R.
    Diehn, F. E.
    Grant, K.
    Halaweish, A. F.
    Krauss, A.
    Raupach, R.
    Schmidt, B.
    McCollough, C. H.
    Fletcher, J. G.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (11) : 2184 - 2190
  • [8] ITERATIVE DEBLURRING OF CT IMAGE-RESTORATION, METAL ARTIFACT REDUCTION AND LOCAL RECONSTRUCTION
    WANG, G
    SNYDER, DL
    VANNIER, MW
    RADIOLOGY, 1995, 197 : 291 - 291
  • [9] Metal Artifact Reduction in Tomosynthesis Imaging
    Zhang, Zhaoxia
    Yan, Ming
    Tao, Kun
    Xuan, Xiao
    Sabol, John M.
    Lai, Hao
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [10] Image quality and metal artifact reduction in total hip arthroplasty CT: deep learning-based algorithm versus virtual monoenergetic imaging and orthopedic metal artifact reduction
    Selles, Mark
    Wellenberg, Ruud H. H.
    Slotman, Derk J.
    Nijholt, Ingrid M.
    van Osch, Jochen A. C.
    van Dijke, Kees F.
    Maas, Mario
    Boomsma, Martijn F.
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)