Efficiency of a cubic NaI(Tl) detector with rectangular cavity using standard radioactive point sources placed at non-axial position

被引:34
作者
Abbas, Mahmoud, I [1 ]
Badawi, Mohamed S. [2 ]
Thabet, Abouzeid A. [3 ]
Kopatch, Yu N. [4 ]
Ruskov, Ivan N. [4 ,5 ]
Grozdanov, D. N. [4 ,5 ]
Noureddine, Salam [6 ]
Fedorov, N. A. [4 ,7 ]
Gouda, Mona M. [1 ]
Hramco, C. [4 ,8 ]
Abd-Elzaher, Mohamed [9 ]
Hamzawy, A. [10 ]
Elsafi, Mohamed [1 ]
El-Khatib, Ahmed M. [1 ]
机构
[1] Alexandria Univ, Fac Sci, Phys Dept, Alexandria 21511, Egypt
[2] Beirut Arab Univ, Fac Sci, Phys Dept, Beirut, Lebanon
[3] Pharos Univ Alexandria, Fac Allied Med Sci, Dept Med Equipment Technol, Alexandria, Egypt
[4] Joint Inst Nucl Res JINR, Dubna, Russia
[5] BAS, Inst Nucl Res & Nucl Energy INRNE, Sofia, Bulgaria
[6] Lebanese Univ, Fac Sci, Phys Dept, Beirut, Lebanon
[7] Lomonosov Moscow State Univ MSU, Fac Phys, Moscow, Russia
[8] Moldavian Acad Sci, Inst Chem, Kishinev, Moldova
[9] Arab Acad Sci Technol & Maritime Transport, Fac Engn, Dept Basic & Appl Sci, Alexandria, Egypt
[10] Umm Al Qura Univ, Al Jamoum Univ Coll, Phys Dept, Al Jamoum, Saudi Arabia
关键词
Cubic detector; Geant4; simulation; Efficiency transfer; Effective solid angle and full-energy peak efficiency; ENERGY PEAK EFFICIENCY; MATHEMATICAL-METHOD; CALCULATE;
D O I
10.1016/j.apradiso.2020.109139
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Low cost scintillation detectors as compared with HPGe detectors are considered to be one of most important radiation detection tools. Therefore, these detectors can be manufactured in different shapes and work at room temperature without any cooling systems, which added an extra advantage to it. This work presents a study of a cubic detector with a rectangular cavity in different experimental setup geometries, using standard point-like gamma-ray sources, where the efficiency of the detector in these geometries was the target to be studied. According to this aim, the data from the experimental measurements was used to determine the detector efficiency. An analytical calculation of the detector efficiency was done by using a new mathematical expression, this mathematical expression depends on the efficiency transfer technique and effective solid angle calculations. To support the mathematical model, the source-to-detector arrangement was simulated by Geant4 Monte Carlo code. All the compared efficiency results were found to be promising and trusted based on the calculated deviation percentages.
引用
收藏
页数:10
相关论文
共 20 条
[1]   Analytical calculations of the solid angles subtended by a well-type detector at point and extended circular sources [J].
Abbas, Mahmoud I. .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (09) :1048-1056
[2]   A new analytical method to calibrate cylindrical phoswich and LaBr3(Ce) scintillation detectors [J].
Abbas, Mahmoud I. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) :413-418
[3]   A simple mathematical method to determine the efficiencies of log-conical detectors [J].
Abbas, MI ;
Nafee, S ;
Selim, YS .
RADIATION PHYSICS AND CHEMISTRY, 2006, 75 (07) :729-736
[4]   Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software [J].
Badawi, Mohamed S. ;
Jovanovic, Slobodan I. ;
Thabet, Abouzeid A. ;
El-Khatib, Ahmed M. ;
Dlabac, Aleksandar D. ;
Salem, Bohaysa A. ;
Gouda, Mona M. ;
Mihaljevic, Nikola N. ;
Almugren, Kholud S. ;
Abbas, Mahmoud I. .
AIP ADVANCES, 2017, 7 (03)
[5]   An empirical formula to calculate the full energy peak efficiency of scintillation detectors [J].
Badawi, Mohamed S. ;
Abd-Elzaher, Mohamed ;
Thabet, Abouzeid A. ;
El-khatib, Ahmed M. .
APPLIED RADIATION AND ISOTOPES, 2013, 74 :46-49
[6]  
Briesmeister J.F., 2010, MCNP GEN MONTE CARLO, V456
[7]   ROOT - An object oriented data analysis framework [J].
Brun, R ;
Rademakers, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 389 (1-2) :81-86
[8]   Full-Energy peak efficiency of an NaI(Tl) detector with coincidence summing correction showing the effect of the source-to-detector distance [J].
El-Khatib, Ahmed M. ;
Salem, Bohaysa A. ;
Badawi, Mohamed S. ;
Gouda, Mona M. ;
Thabet, Abouzeid A. ;
Abbas, Mahmoud I. .
CHINESE JOURNAL OF PHYSICS, 2017, 55 (02) :478-489
[9]   Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity [J].
El-Khatib, Ahmed M. ;
Badawi, Mohamed S. ;
Thabet, Abouzeid A. ;
Jovanovic, Slobodan I. ;
Gouda, Mona M. ;
Mohamed, Moustafa M. ;
Dlabac, Aleksandar D. ;
Abd-Elzaher, Mohamed ;
Mihaljevic, Nikola N. ;
Abbas, Mahmoud I. .
CHINESE JOURNAL OF PHYSICS, 2016, 54 (03) :338-346
[10]   STUDY ON THE EFFECT OF THE SELF-ATTENUATION COEFFICIENT ON γ-RAY DETECTOR EFFICIENCY CALCULATED AT LOW AND HIGH ENERGY REGIONS [J].
El-Khatib, Ahmed. M. ;
Thabet, Abouzeid A. ;
Elzaher, Mohamed A. ;
Badawi, Mohamed S. ;
Salem, Bohaysa A. .
NUCLEAR ENGINEERING AND TECHNOLOGY, 2014, 46 (02) :217-224