Objective To assess the value of applying rigid threshold values in interpreting prostate specific antigen (PSA) results, by selecting and comparing five current methods for measuring free and total PSA. Materials and methods Samples taken from an ongoing screening study for prostate cancer (total PSA by Tandem-E assay, 17 334 participants; biopsy criterion a PSA of 3.0 mug/L, 4 464 men) from men with a total PSA of 1.0-6.0 mug/L were measured for free and total PSA using the Access, Immulite, Elecsys and Prostatus analysis kits, in two patient groups, i.e. with prostate cancer or no evidence of disease. Results Both patient groups had equal means for total PSA but not for free PSA. In all, 360 samples from men with cancer and 96 from men with no evidence of disease were analysed. All methods applied to both groups deviated statistically significantly from the Tandem-E result for total PSA, except for the Access kit. There was a close correlation among all the methods (correlation coefficients of 0.89-0.97). There were very discordant results for the combination of the Tandem-E vs Prostatus (8% difference), representing 315 participants at a threshold of 3.0 mug/L. For free PSA (free/total PSA) the situation was worse, with extreme differences of 32% and 36% for both patient groups (Elecsys vs Access). Conclusions Depending on the threshold value applied as an indication for biopsy, when using the total PSA alone or combined with the free/total PSA, care is needed in interpreting patient groups because of the discordance among PSA assays.
引用
收藏
页码:545 / 550
页数:6
相关论文
共 50 条
  • [1] Free prostate-specific antigen outperforms total prostate-specific antigen as a predictor of prostate volume in patients without prostate cancer
    Avci, Sinan
    Onen, Efe
    Caglayan, Volkan
    Kilic, Metin
    Sambel, Murat
    Oner, Sedat
    ARCHIVIO ITALIANO DI UROLOGIA E ANDROLOGIA, 2020, 92 (01) : 1 - 6
  • [2] Consistency and diagnostic accuracy of 4 assays in the detection of the total and free prostate-specific antigen
    Deng, Lingyan
    Yue, Daoyuan
    Wang, Xu
    Li, Huijun
    TRANSLATIONAL ANDROLOGY AND UROLOGY, 2023, 12 (02) : 261 - 270
  • [3] Stability of total prostate-specific antigen and free prostate-specific antigen after 10 years' storage
    Simanek, Vaclav
    Topolcan, Ondrej
    Karlikova, Marie
    Dolejsova, Olga
    Fuchsova, Radka
    Kinkorova, Judita
    Slouka, David
    Kucera, Radek
    INTERNATIONAL JOURNAL OF BIOLOGICAL MARKERS, 2018, 33 (04): : 463 - 466
  • [4] A comparative performance analysis of total prostate-specific antigen, percentage free prostate-specific antigen, prostate-specific antigen velocity and urinary prostate cancer gene 3 in the first, second and third repeat prostate biopsy
    Auprich, Marco
    Augustin, Herbert
    Budaeus, Lars
    Kluth, Luis
    Mannweiler, Sebastian
    Shariat, Shahrokh F.
    Fisch, Margit
    Graefen, Markus
    Pummer, Karl
    Chun, Felix K. -H.
    BJU INTERNATIONAL, 2012, 109 (11) : 1627 - 1635
  • [5] Complexed prostate-specific antigen for the detection of prostate cancer
    Filella, X
    Truan, D
    Alcover, J
    Gutierrez, R
    Molina, R
    Coca, F
    Ballesta, AM
    ANTICANCER RESEARCH, 2004, 24 (06) : 4181 - 4185
  • [6] Using the free-to-total prostate-specific antigen ratio to detect prostate cancer in men with nonspecific elevations of prostate-specific antigen levels
    Richard M. Hoffman
    David L. Clanon
    Benjamin Littenberg
    Joseph J. Frank
    John C. Peirce
    Journal of General Internal Medicine, 2000, 15 : 739 - 748
  • [7] Effect of transrectal ultrasonography of the prostate on serum prostate-specific antigen levels and free/total prostate-specific antigen ratio
    Serel, TA
    Çetin, M
    Delibas, N
    Çelik, E
    Tahoglu, M
    UROLOGIA INTERNATIONALIS, 2000, 64 (01) : 24 - 26
  • [8] Using the free-to-total prostate-specific antigen ratio to detect prostate cancer in men with nonspecific elevations of prostate-specific antigen levels
    Hoffman, RM
    Clanon, DL
    Littenberg, B
    Frank, JJ
    Peirce, JC
    JOURNAL OF GENERAL INTERNAL MEDICINE, 2000, 15 (10) : 739 - 748
  • [9] Value of Prostate-Specific Antigen and Prostate-Specific Antigen Density in Detection of Prostate Cancer in an Iranian Population of Men
    Ghafoori, Mahyar
    Varedi, Peyman
    Hosseini, Seyed Jalil
    Asgari, Mojgan
    Shakiba, Madjid
    UROLOGY JOURNAL, 2009, 6 (03) : 182 - 188
  • [10] Serum free prostate-specific antigen in the diagnosis of prostate cancer
    Leung, HY
    Lai, LC
    Day, J
    Thomson, J
    Neal, DE
    Hamdy, FC
    BRITISH JOURNAL OF UROLOGY, 1997, 80 (02): : 256 - 259