A Blood Assist Index Control by Intraaorta Pump: A Control Strategy for Ventricular Recovery

被引:22
作者
Gao, Bin [1 ]
Gu, Kaiyun [1 ]
Zeng, Yi [1 ]
Liu, Youjun [1 ]
Chang, Yu [1 ]
机构
[1] Beijing Univ Technol, Sch Life Sci & BioEngn, Beijing 100124, Peoples R China
基金
中国国家自然科学基金;
关键词
PHYSIOLOGICAL CONTROL; CONTROL-SYSTEM; ROTARY; DEVICE; DURATION; SUPPORT; DESIGN;
D O I
10.1097/MAT.0b013e3182257fac
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Left ventricular assist devices (LVAD) are increasingly used for long-term support in heart failure patients. To promote ventricular reverse remodeling, a defined and adjustable energy distribution of LVAD and native heart is important. Therefore, a blood assist index (BAI), which is a ratio of power of LVAD and total power of the cardiovascular system, is defined to indicate the energy distribution of LVAD and native heart. Subsequently, an LVAD control algorithm that uses the BAI as control input is designed. The control strategy maintains the measured BAI tracking the desired BAI. A mathematic model of cardiovascular system is used to verify the feasibility of control strategy in the presence of left ventricular failure, physical active, and a recovery of cardiac function. The simulation results show that the control strategy automatically increases pump speed in response to the reduced peripheral systemic resistance (5,500 vs. 6,000 RPM). When E-max is increased from 0.6 to 1.8 mm Hg/ml to mimic left ventricular recovery, the blood flow is automatically increased from 5 to 8 L/min. As a key feature, the proposed control strategy provides a defined and adjustable energy distribution of LVAD and native heart by regulating the rotational speed of the pump, which is benefit to promote the left ventricular reverse remodeling. ASAIO Journal 2011; 57:358-362.
引用
收藏
页码:358 / 362
页数:5
相关论文
共 29 条
[1]   Physiological Control of a Rotary Blood Pump With Selectable Therapeutic Options: Control of Pulsatility Gradient [J].
Arndt, Andreas ;
Nuesser, Peter ;
Graichen, Kurt ;
Mueller, Johannes ;
Lampe, Bernhard .
ARTIFICIAL ORGANS, 2008, 32 (10) :761-771
[2]   Hierarchical control of heart-assist devices [J].
Boston, JR ;
Antaki, JF ;
Simaan, MA .
IEEE ROBOTICS & AUTOMATION MAGAZINE, 2003, 10 (01) :54-64
[3]   LVAD-Induced reverse remodeling: Basic and clinical implications for myocardial recovery [J].
Burkhoff, D ;
Klotz, S ;
Mancini, DM .
JOURNAL OF CARDIAC FAILURE, 2006, 12 (03) :227-239
[4]  
Chang Y, ASAIO J IN PRESS
[5]   Modeling and Identification of an Intra-Aorta Pump [J].
Chang, Yu ;
Gao, Bin .
ASAIO JOURNAL, 2010, 56 (06) :504-509
[6]   A Global Sliding Mode Controller Design for an Intra-Aorta Pump [J].
Chang, Yu ;
Gao, Bin .
ASAIO JOURNAL, 2010, 56 (06) :510-516
[7]   Myocardial Sympathetic Innervation and Long-Term Left Ventricular Mechanical Unloading [J].
Drakos, Stavros G. ;
Athanasoulis, Theodoros ;
Malliaras, Konstantinos G. ;
Terrovitis, John V. ;
Diakos, Nikolaos ;
Koudoumas, Dimitrios ;
Ntalianis, Argirios S. ;
Theodoropoulos, Stergios P. ;
Yacoub, Magdi H. ;
Nanas, John N. .
JACC-CARDIOVASCULAR IMAGING, 2010, 3 (01) :64-70
[8]   A Control System for Rotary Blood Pumps Based on Suction Detection [J].
Ferreira, Antonio ;
Boston, J. Robert ;
Antaki, James F. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (03) :656-665
[9]  
Gao B, ARTIF ORGAN IN PRESS
[10]   Physiological Control of Intraaorta Pump Based on Heart Rate [J].
Gao, Bin ;
Nie, Li Ya ;
Chang, Yu ;
Zeng, Yi .
ASAIO JOURNAL, 2011, 57 (03) :152-157