Prevalence and Significance of Patient Prosthesis Mismatch Following Edwards SAPIEN XT and SAPIEN 3 Transcatheter Aortic Valve Replacement

被引:2
|
作者
Vallabhaneni, Srilakshmi [1 ]
Matka, Marsel [1 ]
Modi, Vivek [1 ]
Carey, Matthew [2 ]
Sarnoski, Christopher [3 ]
Durkin, Raymond [3 ]
Olenchock, Stephen [4 ]
Razzaghi, Mehdi [5 ]
Shirani, Jamshid [1 ]
机构
[1] St Lukes Univ Hlth Network, Cardiol, Bethlehem, PA 18015 USA
[2] St Lukes Univ Hlth Network, Internal Med, Bethlehem, PA USA
[3] St Lukes Univ Hlth Network, Intervent Cardiol, Bethlehem, PA USA
[4] St Lukes Univ Hlth Network, Cardiac Surg, Bethlehem, PA USA
[5] Bloomsburg Univ Penn, Stat, Bloomsburg, PA 17815 USA
关键词
aortic stenosis; transcatheter aortic valve replacement; patient prosthesis mismatch; outcomes; edwards sapien valve; HIGH-RISK PATIENTS; CLINICAL IMPACT; STENOSIS; IMPLANTATION; PREDICTORS; SURVIVAL; OUTCOMES; TRIAL;
D O I
10.7759/cureus.18044
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Severe patient prosthesis mismatch (sPPM) after surgical aortic valve replacement is associated with worse outcomes. Limited data exists on the impact of sPPM on outcomes after transcatheter aortic valve replacement (TAVR), especially regarding the newer generation valves. The aim of this study was to evaluate the incidence, determinants, and outcomes of sPPM in patients undergoing TAVR with Edwards SAPIEN XT (ES XT) and Edwards SAPIEN 3 (ES3) valves (Edwards Lifesciences, Irvine, CA, USA). Methods We retrospectively reviewed 366 patients who underwent TAVR with ES XT (n = 114) or ES3 (n = 252) valves between July 2012 and June 2018. sPPM was defined as indexed effective orifice area (iEOA) <0.65 cm(2)/m(2). Kaplan-Meier survival estimates were used to determine outcomes. Results Multivariate linear regression analysis was utilized to determine potential independent effects of PPM on outcomes. sPPM was present in 40 (11%) of the patients [8 (7%) ES XT and 32 (13%) ES3] and was associated with female sex, smaller left ventricular outflow tract (LVOT) diameter and aortic valve annular area, absence of prior coronary artery bypass graft (CABG) surgery, shorter height, higher body mass index, and smaller pre-TAVR valve area (all p < 0.05). Among those with ES3 valves, the incidence of sPPM was inversely proportional to the valve size (50%, 25%, 5% and 3% for 20-, 23-, 26- and 29mm valve sizes, respectively; p < 0.001). At a mean follow-up period of 3.5 +/- 1.5 years, there was no difference in all-cause mortality (22.5% vs. 25.6%, p = 0.89) or a composite endpoint of heart failure, arrhythmias, stroke, and myocardial infarction (30% vs. 34%, p = 0.24) in those with or without sPPM. Conclusion ES3 was associated with a higher incidence of sPPM, particularly with smaller valve sizes. However, the presence of sPPM as defined by iEOA was not an independent predictor of adverse outcomes in patients undergoing TAVR within an intermediate follow-up period.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Prosthesis-Patient Mismatch after Transcatheter Aortic Valve Implantation Using the Edwards SAPIEN Prosthesis
    Van Linden, Arnaud
    Kempfert, Joerg
    Blumenstein, Johannes
    Rastan, Ardawan
    Holzhey, David
    Lehmann, Sven
    Mohr, Friedrich W.
    Walther, Thomas
    THORACIC AND CARDIOVASCULAR SURGEON, 2013, 61 (05): : 414 - 420
  • [2] Low Incidence of Prosthesis-Patient Mismatch Following Extremely Small Transcatheter Aortic Valve Replacement with 20-mm Sapien 3
    Okajima, Kazue
    Komatsu, Ikki
    Seto, Todd B.
    INTERNATIONAL JOURNAL OF ANGIOLOGY, 2021, : 120 - 125
  • [3] Edwards SAPIEN™ and Edwards SAPIEN XT™ transcatheter heart valves for the treatment of severe aortic stenosis
    Freeman, Melanie
    Webb, John G.
    EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (06) : 563 - 569
  • [4] Management of Paravalvular Regurgitation After Edwards SAPIEN Transcatheter Aortic Valve Replacement
    Martinez, Claudia A.
    Singh, Vikas
    O'Neill, Brian P.
    Alfonso, Carlos E.
    Bilsker, Martin S.
    Clark, Pedro Martinez
    Williams, Donald
    Cohen, Mauricio G.
    Heldman, Alan W.
    O'Neill, William W.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (02) : 300 - 311
  • [5] Evaluation of the Edwards SAPIEN 3 Transcatheter Valve For Aortic Stenosis
    Htun, Nay M.
    Webb, John G.
    EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (03) : 225 - 232
  • [6] Hemodynamics of the Edwards Sapien XT transcatheter heart valve in noncircular aortic annuli
    Scharfschwerdt, Michael
    Meyer-Saraei, Roza
    Schmidtke, Claudia
    Sievers, Hans-Hinrich
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 148 (01): : 126 - 132
  • [7] Feasibility and Safety of Early Discharge After Transfemoral Transcatheter Aortic Valve Implantation With the Edwards SAPIEN-XT Prosthesis
    Durand, Eric
    Eltchaninoff, Helene
    Canville, Alexandre
    Bouhzam, Najime
    Godin, Matthieu
    Tron, Christophe
    Rodriguez, Carlos
    Litzler, Pierre-Yves
    Bauer, Fabrice
    Cribier, Alain
    AMERICAN JOURNAL OF CARDIOLOGY, 2015, 115 (08): : 1116 - 1122
  • [8] First-in-Man Transfemoral Transcatheter Aortic Valve Replacement With the 29 mm Edwards SAPIEN XT Valve
    Freeman, Melanie
    Rodes-Cabau, Josep
    Urena, Marina
    DeLarochelliere, Robert
    Dumont, Eric
    Masson, Jean-Bernard
    Willson, Alexander B.
    Binder, Ronald K.
    Toggweiler, Stefan
    Leipsic, Jonathon
    Wood, David A.
    Webb, John G.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (04) : 664 - 670
  • [9] Comparison of Edwards SAPIEN 3 versus SAPIEN XT in transfemoral transcatheter aortic valve implantation: Difference of valve selection in the real world
    Arai, Takahide
    Lefevre, Thierry
    Hovasse, Thomas
    Morice, Marie-Claude
    Garot, Philippe
    Benamer, Hakim
    Unterseeh, Thierry
    Hayashida, Kentaro
    Watanabe, Yusuke
    Bouvier, Erik
    Cormier, Bertrand
    Chevalier, Bernard
    JOURNAL OF CARDIOLOGY, 2017, 69 (3-4) : 565 - 569
  • [10] Conduction Abnormalities May Be Transient Following Transcatheter Aortic Valve Replacement with the Edwards-Sapien Prosthesis
    Baltodano, Pablo
    Sager, Solomon I.
    O'Neill, Brian P.
    Shah, Sachil
    Martinez, Claudia A.
    Alfonso, Carlos E.
    Martinez-Clark, Pedro
    Cohen, Mauricio G.
    Heldman, Alan W.
    O'Neill, William W.
    Williams, Donald
    Carrillo, Roger G.
    CIRCULATION, 2012, 126 (21)