High-Temperature (>500°C) Wall Thickness Monitoring Using Dry-Coupled Ultrasonic Waveguide Transducers

被引:126
|
作者
Cegla, Frederic B. [1 ]
Cawley, Peter [1 ]
Allin, Jonathan [2 ]
Davies, Jacob [2 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Mech Engn, London, England
[2] Permasense Ltd, Horsham, W Sussex, England
基金
英国工程与自然科学研究理事会;
关键词
PROPAGATION; SENSORS;
D O I
10.1109/TUFFC.2011.1782
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Conventional ultrasonic transducers cannot withstand high temperatures for two main reasons: the piezoelectric elements within them depolarize, and differential thermal expansion of the different materials within a transducer causes them to fail. In this paper, the design of a high-temperature ultrasonic thickness gauge that bypasses these problems is described. The system uses a waveguide to isolate the vulnerable transducer and piezoelectric elements from the high-temperature measurement zone. Use of thin and long waveguides of rectangular cross section allows large temperature gradients to be sustained over short distances without the need for additional cooling equipment. The main problems that had to be addressed were the transmission and reception of ultrasonic waves into and from the testpiece that the waveguides are coupled to, and optimization of the wave propagation along the waveguide itself. It was found that anti-plane shear loading performs best at transmitting and receiving from the surface of a component that is to be inspected. Therefore, a non-dispersive guided wave mode in large-aspect-ratio rectangular strips was employed to transmit the anti-plane shear loading from the transducer to the measurement zone. Different joining methods to attach the waveguides to the component were investigated and experiments showed that clamping the waveguides to the component surface gave the best results. The thickness of different plate samples was consistently measured to within less than 0.1 mm. Performance at high temperatures was tested in a furnace at 730 degrees C for 4 weeks without signal degradation. Thicknesses in the range of 3 to 25 mm could be monitored using Hanning windowed tonebursts with 2 MHz center frequency.
引用
收藏
页码:156 / 167
页数:12
相关论文
共 27 条
  • [21] Monitoring of CoS2 reactions using high-temperature XRD coupled with gas chromatography (GC)
    Rodriguez, Mark A.
    Coker, Eric N.
    Griego, James J. M.
    Mowry, Curtis D.
    Pimentel, Adam S.
    Anderson, Travis M.
    POWDER DIFFRACTION, 2016, 31 (02) : 90 - 96
  • [22] Fabrication of 4H-SiC DIMOSFETs by high-temperature (>1400°C) rapid thermal oxidation and nitridation using cold-wall oxidation furnace
    Kosugi, R.
    Suzuki, K.
    Takao, K.
    Hayashi, Y.
    Yatsuo, T.
    Fukuda, K.
    Ohashi, H.
    Arai, K.
    SILICON CARBIDE AND RELATED MATERIALS 2005, PTS 1 AND 2, 2006, 527-529 : 1309 - +
  • [23] Automated In-Process Cure Monitoring of Composite Laminates Using a Guided Wave-Based System With High-Temperature Piezoelectric Transducers
    Hudson, Tyler B.
    Yuan, Fuh-Gwo
    JOURNAL OF NONDESTRUCTIVE EVALUATION, DIAGNOSTICS AND PROGNOSTICS OF ENGINEERING SYSTEMS, 2018, 1 (02):
  • [24] High-temperature monitoring of refractory wall recession using frequency-modulated continuouswave (FM-CW) radar techniques
    Varghese, B
    DeConick, C
    Cartee, G
    Zoughi, R
    Velez, M
    Moore, R
    Review of Progress in Quantitative Nondestructive Evaluation, Vols 24A and 24B, 2005, 760 : 1714 - 1720
  • [25] High-temperature tensile strength of alumina-spinel refractory by Brazilian test at 1200°C coupled with DIC in-situ monitoring
    Darban, Sina
    Jastrzebska, Ilona
    Gillibert, Jean
    Prorok, Ryszard
    Sayet, Thomas
    Blond, Eric
    Szczerba, Jacek
    JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2024, 44 (11) : 6743 - 6758
  • [26] Optimizing contact thermometry high-temperature fixed-point cells (> 1,100°C) using finite-element analysis
    Pearce, J. V.
    Lowe, D. H.
    Head, D. I.
    Machin, G.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (01) : 250 - 260
  • [27] Compact Triple-Band High-Temperature Superconducting Filter Using Coupled-Line Stepped Impedance Resonator(C-SIR)
    Liu, Hai Wen
    Peng, Yang
    Guan, Xue Hui
    Lei, Jiu Huai
    Ren, Bao Ping
    Qin, Feng
    Wen, Pin
    Liu, Yan De
    2015 IEEE International Conference on Applied Superconductivity and Electromagnetic Devices (ASEMD), 2015, : 163 - 164