Regenos spacers are not suitable for open-door laminoplasty because of serious adverse events caused by their insufficient mechanical strength

被引:11
作者
Noguchi, Hiroshi [1 ]
Koda, Masao [1 ]
Funayama, Toru [1 ]
Kumagai, Hiroshi [1 ]
Saito, Junya [2 ]
Mannoji, Chikato [3 ]
Aramomi, Masaaki [4 ]
Abe, Tetsuya [1 ]
Nagashima, Katsuya [1 ]
Miura, Kousei [1 ]
Mataki, Kentaro [1 ]
Fuji, Kengo [1 ]
Furuya, Takeo [2 ]
Yamazaki, Masashi [1 ]
机构
[1] Univ Tsukuba, Fac Med, Dept Orthopaed Surg, 1-1-1 Tennodai, Tsukuba, Ibaraki 3058575, Japan
[2] Chiba Univ, Dept Orthoped Surg, Grad Sch Med, Chiba, Japan
[3] Narita Red Cross Hosp, Dept Orthoped Surg, Chiba, Japan
[4] Oyumino Cent Hosp, Dept Orthoped Surg, Chiba, Japan
关键词
Open-door laminoplasty; Hydroxyapatite; Spacer; Complication; Cervical spine; Bone bonding; Laminar reclosure; TECHNICAL IMPROVEMENTS; EXPANSIVE LAMINOPLASTY; CERVICAL MYELOPATHY; CLINICAL ARTICLE; HYDROXYAPATITE; LIGAMENT; MUSCLES; SPINE;
D O I
10.1016/j.jocn.2018.07.015
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We used a newly developed, high-porosity unidirectional porous hydroxyapatite spacer (Regenos spacer, not approved by the FDA). The aim of the present study was to elucidate the effectiveness of Regenos laminar spacers for open-door type laminoplasty. The present study included 39 patients who underwent open-door type laminoplasty with Regenos spacers from April 2015 to December 2016 and were followed up for at least 6 months after surgery. We grafted 68 Regenos spacers in 39 patients. Pre- and postoperative neurological status of patients were evaluated using JOA score and recovery rate. Breakage of Regenos spacers, laminar closure, and bone-hydroxyapatite spacer bonding were assessed using 12-month postoperative sagittal and axial CT images. The average preoperative JOA score was 9.5 +/- 3.2/17, and the average postoperative JOA score was 12.5 +/- 2.9/17. JOA score recovery rate was 34 41% at the latest follow-up visit. The bony fusion rate of the hinge sides was 87%. Breakage and deformity of implanted spacers was observed in 69% of patients and 59% of spacers with a CT sagittal view, and CT axial view at 12 months revealed fine cracks and collapse in 17 spacers in 14 patients. The average angle was -2.4 +/- 4.8 degrees, including 46 of 68 spacers showing a negative value, resulting in a rate of laminar reclosure of 35%. Postoperative CT demonstrated good bone bonding rate. Nevertheless, clinical results with low recovery rates were obtained with complications related to the use of Regenos spacers. (C) 2018 Elsevier Ltd. All rights reserved.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 14 条
[1]   Histological Analysis of Bone Bonding and Ingrowth into Connected Porous Hydroxyapatite Spacers in Spinal Surgery [J].
Funayama, Toru ;
Noguchi, Hiroshi ;
Tsukanishi, Toshinori ;
Sakane, Masataka .
BIOCERAMICS 24, 2013, 529-530 :309-+
[2]   OPERATIVE RESULTS AND POSTOPERATIVE PROGRESSION OF OSSIFICATION AMONG PATIENTS WITH OSSIFICATION OF CERVICAL POSTERIOR LONGITUDINAL LIGAMENT [J].
HIRABAYASHI, K ;
MIYAKAWA, J ;
SATOMI, K ;
MARUYAMA, T ;
WAKANO, K .
SPINE, 1981, 6 (04) :354-364
[3]   En bloc laminoplasty without dissection of paraspinal muscles [J].
Hosono, N ;
Sakaura, H ;
Mukai, Y ;
Ishii, T ;
Yoshikawa, H .
JOURNAL OF NEUROSURGERY-SPINE, 2005, 3 (01) :29-33
[4]   TECHNICAL IMPROVEMENTS AND RESULTS OF LAMINOPLASTY FOR COMPRESSIVE MYELOPATHY IN THE CERVICAL-SPINE [J].
ITOH, T ;
TSUJI, H .
SPINE, 1985, 10 (08) :729-736
[5]   Bone Regeneration and Remodeling within a Unidirectional Porous Hydroxyapatite Bone Substitute at a Cortical Bone Defect Site: Histological Analysis at One and Two Years after Implantation [J].
Iwasashi, Masashi ;
Funayama, Toru ;
Watanabe, Arata ;
Noguchi, Hiroshi ;
Tsukanishi, Toshinori ;
Suetsugu, Yasushi ;
Makihara, Takeshi ;
Ochiai, Naoyuki ;
Yamazaki, Masashi ;
Sakane, Masataka .
MATERIALS, 2015, 8 (08) :4884-4894
[6]   Postoperative displacement of hydroxyapatite spacers implanted during double-door laminoplasty Clinical article [J].
Kaito, Takashi ;
Hosono, Noboru ;
Makino, Takahiro ;
Kaneko, Noriyoshi ;
Namekata, Masato ;
Fuji, Takeshi .
JOURNAL OF NEUROSURGERY-SPINE, 2009, 10 (06) :551-556
[7]   Delayed dural laceration by hydroxyapatite spacer causing tetraparesis following double-door laminoplasty [J].
Kanemura, Aritetsu ;
Doita, Minoru ;
Iguchi, Tetsuhiro ;
Kasahara, Koichi ;
Kurosaka, Masahiro ;
Sumi, Masatoshi .
JOURNAL OF NEUROSURGERY-SPINE, 2008, 8 (02) :121-128
[8]   TECHNICAL IMPROVEMENTS AND RESULTS OF OPEN-DOOR EXPANSIVE LAMINOPLASTY WITH HYDROXYAPATITE IMPLANTS FOR CERVICAL MYELOPATHY [J].
Kihara, Shun-ichi ;
Umebayashi, Takeshi ;
Hoshimaru, Minoru .
NEUROSURGERY, 2005, 57 (04) :348-355
[9]   Biomechanical Evaluation of the Suture Anchors Used in Open-Door Laminoplasty [J].
Kurokawa, Yoshitaka ;
Yokoyama, Yasuharu ;
Kuroda, Koichi ;
Koruprolu, Sarath ;
Paller, David ;
Nakano, Atsushi ;
Baba, Ichiro ;
Neo, Masashi .
SPINE, 2014, 39 (21) :E1248-E1255
[10]   Laminar Closure After Classic Hirabayashi Open-Door Laminoplasty [J].
Lee, Dong-Ho ;
Park, Soo-An ;
Kim, Nam Heun ;
Hwang, Chang Ju ;
Kim, Yung-Tae ;
Lee, Choon Sung ;
Riew, K. Daniel .
SPINE, 2011, 36 (25) :E1634-E1640