Optimising diffusion-weighted imaging in the abdomen and pelvis: comparison of image quality between monopolar and bipolar single-shot spin-echo echo-planar sequences

被引:22
|
作者
Kyriazi, Stavroula [1 ,2 ,3 ]
Blackledge, Matthew [1 ,2 ,3 ]
Collins, David J. [1 ,2 ,3 ]
deSouza, Nandita M. [1 ,2 ,3 ]
机构
[1] Royal Marsden NHS Fdn Trust, CR UK, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Fdn Trust, EPSRC Canc Imaging Ctr, Sutton SM2 5PT, Surrey, England
[3] Inst Canc Res, Sutton SM2 5PT, Surrey, England
关键词
Diffusion-weighted imaging DWI; Image quality; Artefact; Eddy currents; Distortion; EDDY-CURRENT ARTIFACTS; COEFFICIENT MEASUREMENTS; MRI; REDUCTION; GRADIENT;
D O I
10.1007/s00330-010-1826-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare geometric distortion, signal-to-noise ratio (SNR), apparent diffusion coefficient (ADC), efficacy of fat suppression and presence of artefact between monopolar (Stejskal and Tanner) and bipolar (twice-refocused, eddy-current-compensating) diffusion-weighted imaging (DWI) sequences in the abdomen and pelvis. A semiquantitative distortion index (DI) was derived from the subtraction images with b = 0 and 1,000 s/mm(2) in a phantom and compared between the two sequences. Seven subjects were imaged with both sequences using four b values (0, 600, 900 and 1,050 s/mm(2)) and SNR, ADC for different organs and fat-to-muscle signal ratio (FMR) were compared. Image quality was evaluated by two radiologists on a 5-point scale. DI was improved in the bipolar sequence, indicating less geometric distortion. SNR was significantly lower for all tissues and b values in the bipolar images compared with the monopolar (p < 0.05), whereas FMR was not statistically different. ADC in liver, kidney and sacrum was higher in the bipolar scheme compared to the monopolar (p < 0.03), whereas in muscle it was lower (p = 0.018). Image quality scores were higher for the bipolar sequence (p a parts per thousand currency signaEuro parts per thousand 0.025). Artefact reduction makes the bipolar DWI sequence preferable in abdominopelvic applications, although the trade-off in SNR may compromise ADC measurements in muscle.
引用
收藏
页码:2422 / 2431
页数:10
相关论文
共 50 条
  • [21] Diffusion-weighted imaging of the brain at 7 T with echo-planar and turbo spin echo sequences: preliminary results
    Sigmund, Eric E.
    Gutman, David
    MAGNETIC RESONANCE IMAGING, 2011, 29 (06) : 752 - 765
  • [22] Comparison of Readout-Segmented and Conventional Single-Shot for Echo-Planar Diffusion-Weighted Imaging in the Assessment of Kidney Interstitial Fibrosis
    Friedli, Iris
    Crowe, Lindsey Alexandra
    de Perrot, Thomas
    Berchtold, Lena
    Martin, Pierre-Yves
    de Seigneux, Sophie
    Vallee, Jean-Paul
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 46 (06) : 1631 - 1640
  • [23] Prospective Comparison of FOCUS MUSE and Single-Shot Echo-Planar Imaging for Diffusion-Weighted Imaging in Evaluating Thyroid-Associated Ophthalmopathy
    Wang, YunMeng
    Cui, YuanYuan
    Dai, JianKun
    Ni, ShuangShuang
    Zhang, TianRan
    Chen, Xin
    Jiang, QinLing
    Cheng, YuXin
    Ma, YiChuan
    Li, Tuo
    Xiao, Yi
    KOREAN JOURNAL OF RADIOLOGY, 2024, 25 (10) : 913 - 923
  • [24] Image quality assessment of single-shot turbo spin echo diffusion-weighted imaging with parallel imaging technique: a phantom study
    Yoshida, Tsukasa
    Urikura, Atsushi
    Shirata, Kensei
    Nakaya, Yoshihiro
    Terashima, Shingo
    Hosokawa, Yoichiro
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1065):
  • [25] Comparison between MRI with spin-echo echo-planar diffusion-weighted sequence (DWI) and histology in the diagnosis of soft-tissue tumours
    Genovese, E.
    Cani, A.
    Rizzo, S.
    Angeretti, M. G.
    Leonardi, A.
    Fugazzola, C.
    RADIOLOGIA MEDICA, 2011, 116 (04): : 644 - 656
  • [26] Comparison of Readout-Segmented Echo-Planar Imaging (EPI) and Single-Shot EPI in Clinical Application of Diffusion-Weighted Imaging of the Pediatric Brain
    Yeom, Kristen W.
    Holdsworth, Samantha J.
    Van, Anh T.
    Iv, Michael
    Skare, Stefan
    Lober, Robert M.
    Bammer, Roland
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 200 (05) : W437 - W443
  • [27] Differences in apparent diffusion coefficients between normal brain echo-planar images and turbo spin-echo diffusion-weighted images with distortion correction
    Takatsu, Yasuo
    Nakamura, Masafumi
    Sagawa, Hajime
    Suzuki, Yuichi
    Mori, Nobuyuki
    Motegi, Shunichi
    Miyati, Tosiaki
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 149
  • [28] Mean diffusivity, fractional anisotropy maps, and three-dimensional white-matter tractography by diffusion tensor imaging. Comparison between single-shot fast spin-echo and single-shot echo-planar sequences at 1.5 Tesla
    Masaaki Hori
    Keiichi Ishigame
    Nobuyuki Shiraga
    Hiroshi Kumagai
    Shigeki Aoki
    Tsutomu Araki
    European Radiology, 2008, 18 : 830 - 834
  • [29] Mean diffusivity, fractional anisotropy maps, and three-dimensional white-matter tractography by diffusion tensor imaging. Comparison between single-shot fast spin-echo and single-shot echo-planar sequences at 1.5 Tesla
    Hori, Masaaki
    Ishigame, Keiichi
    Shiraga, Nobuyuki
    Kumagai, Hiroshi
    Aoki, Shigeki
    Araki, Tsutomu
    EUROPEAN RADIOLOGY, 2008, 18 (04) : 830 - 834
  • [30] Ultrafast T2-weighted imaging of the abdomen and pelvis: Use of single shot fast spin-echo imaging
    Tang, Y
    Yamashita, Y
    Takahashi, M
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (02): : 384 - 390