Technical report: Reliability and validity of the iSAM 9000 isokinetic dynamometer

被引:20
作者
Orri, Julia C. [1 ]
Darden, Gibson F. [2 ]
机构
[1] Univ San Francisco, Dept Exercise & Sport Sci, San Francisco, CA 94117 USA
[2] Coastal Carolina Univ, Dept Hlth Phys Educ & Recreat, Conway, SC USA
关键词
muscular strength; peak torque; angular velocity; industry; Cybex;
D O I
10.1519/JSC.0b013e31815fa2c8
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
This study assessed the mechanical reliability and validity of the INRTEK iSAM 9000 isokinetic dynamometer, and compared the obtained torque values of the prototype device with those from a traditional device. Sixty volunteers (40 men and 20 women) were tested at 60 degrees per second for shoulder, knee, and trunk flexion, and extension on both the Cybex 6000 and a new isokinetic dynamometer (iSAM 9000). Intraclass correlation coefficients (ICC) and standard errors of measurement (SEM) revealed a high level of reproducibility and precision in the device's torque measurements (ICC range = 0.94-0.98; SEM range = 5.2-29.7). Pearson r values revealed very high relationships between the two instruments (set 1 : r = 0.84-0.93; set 2: r = 0.87-0.93; P < 0.05). Significantly higher peak torque for both sets of left and right knee flexion and extension, right shoulder extension and trunk extension was found for the iSAM 9000 compared to the Cybex 6000 (P < 0.05). The strong ICCs and small SEMs support the device's mechanical reliability and validity. The high correlation coefficients between the prototype dynamometer and the Cybex 6000 support the new device's validity in the measurement of isokinetic torque, The findings of this study will be used to refine the next generation of the INRTEK isokinetic device with respect to test protocols and the reliability of measuring human muscle performance.
引用
收藏
页码:310 / 317
页数:8
相关论文
共 16 条
[1]  
Almekinders, 1994, J Am Acad Orthop Surg, V2, P221
[2]   INTRAMACHINE AND INTERMACHINE RELIABILITY FOR SELECTED DYNAMIC MUSCLE PERFORMANCE TESTS [J].
BANDY, WD ;
MCLAUGHLIN, S .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 1993, 18 (05) :609-613
[3]   Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements [J].
Drouin, JM ;
Valovich-mcLeod, TC ;
Shultz, SJ ;
Gansneder, BM ;
Perrin, DH .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2004, 91 (01) :22-29
[4]  
DUEKER JA, 1994, J OCCUP ENVIRON MED, V36, P42
[5]  
FARRELL M, 1986, MED SCI SPORT EXER, V18, P44
[6]  
FEIRING D C, 1990, Journal of Orthopaedic and Sports Physical Therapy, V11, P298
[7]   Clinical assessment of the low-cost VariCom isokinetic knee exerciser [J].
Greenblatt, D ;
Diesel, W ;
Noakes, TD .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (03) :273-278
[8]  
Gulch R.W., 1994, INT J SPORTS MED, V15, P2
[9]   ISOKINETIC LIFTING STRENGTH AND OCCUPATIONAL INJURY - A PROSPECTIVE-STUDY [J].
MOSTARDI, RA ;
NOE, DA ;
KOVACIK, MW ;
PORTERFIELD, JA .
SPINE, 1992, 17 (02) :189-193
[10]  
Newman MA, 2004, J STRENGTH COND RES, V18, P867