Portable combination of Fourier transform infrared spectroscopy and differential mobility spectrometry for advanced vapor phase analysis

被引:13
作者
Hagemann, L. Tamina [1 ,2 ]
McCartney, Mitchell M. [1 ]
Fung, Alexander G. [1 ]
Peirano, Daniel J. [1 ]
Davis, Cristina E. [1 ]
Mizaikoff, Boris [2 ]
机构
[1] Univ Calif Davis, Dept Mech & Aerosp Engn, Davis, CA 95616 USA
[2] Ulm Univ, Inst Analyt & Bioanalyt Chem, D-89081 Ulm, Germany
基金
美国国家科学基金会;
关键词
GAS-PHASE; ION; HYDROGEN; CHROMATOGRAPHY; CONFORMATIONS; DISSOCIATION; EVOLUTION; METHANE; BREATH; FAIMS;
D O I
10.1039/c8an01192c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Designing mobile devices for the analysis of complex sample mixtures containing a variety of analytes at different concentrations across a large dynamic range remains a challenging task in many analytical scenarios. To meet this challenge, a compact hybrid analytical platform has been developed combining Fourier transform infrared spectroscopy based on substrate-integrated hollow waveguides (iHWG-FTIR) with gas chromatography coupled differential mobility spectrometry (GC-DMS). Due to the complementarity of these techniques regarding analyte type and concentration, their combination provides a promising tool for the detection of complex samples containing a broad range of molecules at different concentrations. To date, the combination of infrared spectroscopy and ion mobility techniques remains expensive and bound to a laboratory utilizing e.g. IMS as prefilter or IR as ionization source. In the present study, a cost-efficient and portable solution has been developed and characterized representing the first truly hyphenated IR-DMS system. As a model analyte mixture, 5 ppm isopropylmercaptan (IPM) in methane (CH4) were diluted, and the concentration-dependent DMS signal of IPM along with the concentration-dependent IR signal of CH4 were recorded for all three hybrid IR-DMS systems. While guiding the sample through the iHWG-FTIR or the GC-DMS first did not affect the obtained signals, optimizing the IR data acquisition parameters did benefit the analytical results.
引用
收藏
页码:5683 / 5691
页数:9
相关论文
共 48 条
[1]   Nonlinearity of Beer's law in gas-phase FT-IR spectroscopy [J].
Ahro, M ;
Kauppinen, J .
APPLIED SPECTROSCOPY, 2001, 55 (01) :50-54
[2]   Detection of Huanglongbing Disease Using Differential Mobility Spectrometry [J].
Aksenov, Alexander A. ;
Pasamontes, Alberto ;
Peirano, Daniel J. ;
Zhao, Weixiang ;
Dandekar, Abhaya M. ;
Fiehn, Oliver ;
Ehsani, Reza ;
Davis, Cristina E. .
ANALYTICAL CHEMISTRY, 2014, 86 (05) :2481-2488
[3]  
[Anonymous], ULLMANNS ENCY IND CH
[4]  
[Anonymous], ATMOS CHEM PHYS
[5]  
[Anonymous], IUPAC COMPENDIUM ANA
[6]   Signal-to-noise ratio, signal processing, and spectral information in the instrumental analysis laboratory [J].
Blitz, JP ;
Klarup, DG .
JOURNAL OF CHEMICAL EDUCATION, 2002, 79 (11) :1358-1360
[7]   Hollow waveguide infrared gas sensing for biomedical applications [J].
Charlton, C ;
Inberg, A ;
Croitoru, N ;
Mizaikoff, B .
OPTICAL FIBERS AND SENSORS FOR MEDICAL APPLICATIONS III, 2003, 4957 :116-123
[8]   Potential of ion mobility spectrometry versus FT-MIR and GC-MS to study the evolution of a heat transfer fluid after its heating process in a thermosolar plant [J].
Criado-Garcia, L. ;
Garrido-Delgado, R. ;
Arce, L. ;
Lopez, F. ;
Peon, R. ;
Valcarcel, M. .
MICROCHEMICAL JOURNAL, 2015, 121 :163-171
[9]   iCONVERT: An Integrated Device for the UV-Assisted Determination of H2S via Mid-Infrared Gas Sensors [J].
da Silveira Petruci, Joao Flavio ;
Cardoso, Arnaldo Alves ;
Wilk, Andreas ;
Kokoric, Vjekoslav ;
Mizaikoff, Boris .
ANALYTICAL CHEMISTRY, 2015, 87 (19) :9580-9583
[10]   Monitoring of hydrogen sulfide via substrate-integrated hollow waveguide mid-infrared sensors in real-time [J].
da Silveira Petruci, Joao Flavio ;
Fortes, Paula Regina ;
Kokoric, Vjekoslav ;
Wilk, Andreas ;
Raimundo, Ivo Milton, Jr. ;
Cardoso, Arnaldo Alves ;
Mizaikoff, Boris .
ANALYST, 2014, 139 (01) :198-203