Acquisition of 3He ventilation images, ADC, T2*and B1 maps in a single scan with compressed sensing

被引:9
作者
Ajraoui, S. [1 ]
Parra-Robles, J. [1 ]
Marshall, H. [1 ]
Deppe, M. H. [1 ]
Clemence, M. [2 ]
Wild, J. M. [1 ]
机构
[1] Univ Sheffield, Sect Acad Radiol, Sheffield S10 2JF, S Yorkshire, England
[2] Philips Healthcare, Philips Clin Sci, Guildford, Surrey, England
基金
英国工程与自然科学研究理事会;
关键词
hyperpolarised; helium-3; lung; compressed sensing; T2*; ADC; HYPERPOLARIZED HE-3; DIFFUSION MEASUREMENTS; PARTIAL-PRESSURE; IN-VIVO; LUNG; MRI; TIME; GAS; COEFFICIENT; RESOLUTION;
D O I
10.1002/nbm.1710
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In imaging of human lungs with hyperpolarised noble gases, measurements of apparent diffusion coefficient (ADC) and relaxation time provide valuable information for the assessment of lung microstructure. In this work, a sequence was developed for interleaved acquisition of ventilation images, ADC, T2* and flip angle maps in a single scan from the human lungs with a single dose of inhaled 3He at 3T. Spatially registered ventilation images with parametric maps were obtained. The total acquisition time was reduced by random undersampling of the k-space and reconstruction using compressed sensing (CS). The gain in speed was used for an increase in spatial resolution. Mean ADC values from the fully sampled and undersampled CS data exhibit no statistically significant difference in a given subject. The mean T2* values, however, were found to differ significantly, which is attributed to the combined effect of low signal-to-noise ratio (SNR) of the fully sampled data and the smoothing effect inherent in CS reconstruction. Copyright (c) 2011 John Wiley & Sons, Ltd.
引用
收藏
页码:44 / 51
页数:8
相关论文
共 35 条
[1]  
Ajraoui S, 2008, P INT SOC MAG RESON, V16, P2652
[2]  
Ajraoui S, 2009, P INT SOC MAG RESON, V17, P2175
[3]   Compressed Sensing in Hyperpolarized 3He Lung MRI [J].
Ajraoui, Salma ;
Lee, Kuan J. ;
Deppe, Martin H. ;
Parnell, Steven R. ;
Parra-Robles, Juan ;
Wild, Jim M. .
MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (04) :1059-1069
[4]   Simultaneous T-2* and diffusion measurements with He-3 [J].
Bock, M .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) :890-895
[5]   Robust uncertainty principles:: Exact signal reconstruction from highly incomplete frequency information [J].
Candès, EJ ;
Romberg, J ;
Tao, T .
IEEE TRANSACTIONS ON INFORMATION THEORY, 2006, 52 (02) :489-509
[6]  
Chen XJ, 1999, MAGN RESON MED, V42, P729, DOI 10.1002/(SICI)1522-2594(199910)42:4<729::AID-MRM15>3.0.CO
[7]  
2-2
[8]  
Cieslar K, 2009, P INL SOC MAG RESON, V17, P4
[9]   Quantification of regional intrapulmonary oxygen partial pressure evolution during apnea by 3He MRI [J].
Deninger, AJ ;
Eberle, B ;
Ebert, M ;
Grossmann, T ;
Heil, W ;
Kauczor, HU ;
Lauer, L ;
Markstaller, K ;
Otten, E ;
Schmiedeskamp, J ;
Schreiber, W ;
Surkau, R ;
Thelen, M ;
Weiler, N .
JOURNAL OF MAGNETIC RESONANCE, 1999, 141 (02) :207-216
[10]   Susceptibility Effects in Hyperpolarized 3He Lung MRI at 1.5T and 3T [J].
Deppe, Martin H. ;
Parra-Robles, Juan ;
Ajraoui, Salma ;
Parnell, Steven R. ;
Clemence, Matthew ;
Schulte, Rolf F. ;
Wild, Jim M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (02) :418-423