The circadian variation of atrial defibrillation thresholds

被引:2
作者
Mitchell, ARJ
Warman, E
Schaaf, K
Sulke, N
机构
[1] Eastbourne Gen Hosp, Dept Cardiol, Eastbourne, England
[2] Medtronic Inc, Minneapolis, MN USA
关键词
defibrillation threshold; atrial fibrillation; circadian rhythm; atrial defibrillator;
D O I
10.1023/B:JICE.0000011489.99159.25
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A circadian variation exists for ventricular defibrillation thresholds (DFTs) with a morning peak and a corresponding decrease in therapy success rates from implantable cardioverter defibrillators. Such a variation in atrial DFTs may have implications for the timing of internal cardioversion of atrial arrhythmias. The aim of this study was therefore to determine the circadian variation of atrial DFTs in patents with recurrent atrial fibrillation (AF). Methods and results: Data were collected as part of the worldwide Jewel AF-only study. Patients had recurrent persistent AF and no history of ventricular arrhythmias. The atrial DFT was assessed at device implantation using a step-up protocol and was recorded for 100 patients (age 63.0 +/- 11.7, 74% male, ejection fraction 49.6 +/- 17.8%, left atrial diameter 46 +/- 9 mm). The mean atrial DFT was 6.3 +/- 4.3 J. For the most commonly tested lead configuration (right atrium to coronary sinus in 56 patients), the atrial DFT for patients implanted in the morning (3.3 +/- 1.5 J) was significantly lower than for both the DFT measured in the afternoon (5.8 +/- 3.4 J, p < 0.01) and the DFT measured in the evening (7.4 +/- 5.9 J, p = 0.01). Conclusion: There may be a significant variation in measured atrial DFT for the right atrium to coronary sinus configuration, with a nadir in the morning. This is the converse to measurements of ventricular DFTs suggesting different regulatory electrophysiological mechanisms. Further investigation of this possible variation is warranted.
引用
收藏
页码:79 / 83
页数:5
相关论文
共 29 条
  • [1] Pain threshold for low energy intracardiac cardioversion of atrial fibrillation with low or no sedation
    Ammer, R
    Alt, E
    Ayers, G
    Schmitt, C
    Pasquantonio, J
    Schmidt, M
    Putter, K
    Schomig, A
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 230 - 236
  • [2] Effect of increased parasympathetic and sympathetic tone on internal atrial defibrillation thresholds in humans
    Barold, HS
    Shander, G
    Tomassoni, G
    Simons, GR
    Wharton, JM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (01): : 238 - 242
  • [3] Favorable effects of flecainide in transvenous internal cardioversion of atrial fibrillation
    Boriani, G
    Biffi, M
    Capucci, A
    Bronzetti, G
    Ayers, GM
    Zannoli, R
    Branzi, A
    Magnani, B
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (02) : 333 - 341
  • [4] Atrial signal analysis and defibrillation threshold assessment in chronic persistent and reinduced atrial fibrillation
    Boriani, G
    Bartolini, P
    Biffi, M
    Calcagnini, G
    Camanini, C
    Corazza, I
    Zannoli, R
    Barbaro, V
    Branzi, A
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2002, 13 (05) : 449 - 454
  • [5] Boriani G, 2000, PACE, V23, P1898
  • [6] CINCA J, 1990, ANN NY ACAD SCI, V601, P222
  • [7] AT WHAT TIME ARE IMPLANTABLE DEFIBRILLATOR SHOCKS IS DELIVERED - EVIDENCE FOR INDIVIDUAL CIRCADIAN VARIANCE IN SUDDEN CARDIAC DEATH
    DAVILA, A
    WELLENS, F
    ANDRIES, E
    BRUGADA, P
    [J]. EUROPEAN HEART JOURNAL, 1995, 16 (09) : 1231 - 1233
  • [8] The effects of atrial electrical remodeling on atrial defibrillation thresholds
    Everett, TH
    Li, H
    Mangrum, JM
    Mastrangelo, C
    Haines, DE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (08): : 1208 - 1215
  • [9] Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias
    Gold, MR
    Sulke, N
    Schwartzman, DS
    Mehra, R
    Euler, DE
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (11) : 1247 - 1253
  • [10] Rounded biphasic waveform reduces energy requirements for transvenous catheter cardioversion of atrial fibrillation and flutter
    Harbinson, MT
    Allen, JD
    Imam, Z
    Dempsey, G
    Anderson, JM
    Ayers, GM
    Adgey, AAJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 226 - 229