Assessing the impact of non-dilating the eye on full-field electroretinogram and standard flash response

被引:21
作者
Gagne, A-M [1 ]
Lavoie, J. [1 ]
Lavoie, M-P [1 ]
Sasseville, A. [1 ]
Charron, M-C [1 ]
Hebert, M. [1 ]
机构
[1] Ctr Rech Univ Laval Robert Giffard, Quebec City, PQ G1J 2G3, Canada
关键词
Eye dilatation; Electroretinogram; Photopic; Scotopic; Standard flash; ISCEV; PHOTOPIC HILL; SENSITIVITY; RETINA;
D O I
10.1007/s10633-010-9242-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We investigated the possibility of performing electroretinography (ERG) in non-pharmacologically dilated eyes using brighter flash (time-integrated) luminance Photopic (N = 26, background 25 5 cd m(-2), white LED flashes) and scotopic ERG (N = 23, green LED flashes) luminance response functions were obtained simultaneously in a dilated (DE) and non-dilated eye (NDE) In the NDE, photopic V-max b-wave amplitude was reduced by 14% (P < 0 0001), implicit time prolonged (P < 0 0001), and retinal sensitivity (log K) decreased by 0 38 log units (P < 0 0001) with no effect on a-wave Using a xenon strobe light (N = 6) to increase flash luminance, Vmax remained lower by about 12% in the NDE (P = 0 02) V-max with LED and xenon was achieved at 3 9 +/- 1 0 cd S m(-2) and 3 3 +/- 0 81 cd S m(-2) in the DE and 10 6 +/- 1 2 cd s m(-2) and 12 3 +/- 1 90 cd s m(-2) in the NDE, that is an increase of 0 43 and 0 57 log unit (P <0 0001), respectively Increasing background luminance by 0 50 log units (80 cd m-2, N = 4) resulted in implicit time normalization but not V-max amplitude Rod V-max was decreased by 7% in NDE (P < 0 05) and sensitivity reduced by 0 40 log units (P <0 0001), but our data suggest that the luminance may have not been sufficient to reach V-max in all participants in the NDE and that the sensitivity change may have been due to an inadequate inter-stimulus interval For the photopic ERG, increasing flash luminance is not sufficient to compensate for the smaller pupil size, whereas for the scotopic ERG, more data are needed to establish proper inter-stimulus interval to perform recordings in a non-pharmacologically dilated
引用
收藏
页码:167 / 175
页数:9
相关论文
共 19 条
[11]  
LACHAPELLE P, 1987, CAN J OPHTHALMOL, V22, P354
[12]   ISCEV Standard for full-field clinical electroretinography (2008 update) [J].
Marmor, M. F. ;
Fulton, A. B. ;
Holder, G. E. ;
Miyake, Y. ;
Brigell, M. ;
Bach, M. .
DOCUMENTA OPHTHALMOLOGICA, 2009, 118 (01) :69-77
[13]   Standard for clinical electroretinography (2004 update) [J].
Marmor, MF ;
Holder, GE ;
Seeliger, MW ;
Yamamoto, S .
DOCUMENTA OPHTHALMOLOGICA, 2004, 108 (02) :107-114
[14]   PROPERTIES OF ELECTRORETINOGRAPHIC INTENSITY-RESPONSE FUNCTIONS IN RETINITIS PIGMENTOSA [J].
MASSOF, RW ;
WU, L ;
FINKELSTEIN, D ;
PERRY, C ;
STARR, SJ ;
JOHNSON, MA .
DOCUMENTA OPHTHALMOLOGICA, 1984, 57 (03) :279-296
[15]   PROPERTIES OF THE HUMAN CONE SYSTEM ELECTRORETINOGRAM DURING LIGHT ADAPTATION [J].
PEACHEY, NS ;
ALEXANDER, KR ;
FISHMAN, GA ;
DERLACKI, DJ .
APPLIED OPTICS, 1989, 28 (06) :1145-1150
[16]   The photopic ERG luminance-response function (photopic hill): method of analysis and clinical application [J].
Rufiange, M ;
Dassa, J ;
Dembinska, O ;
Koenekoop, RK ;
Little, JM ;
Polomeno, RC ;
Dumont, M ;
Chemtob, S ;
Lachapelle, P .
VISION RESEARCH, 2003, 43 (12) :1405-1412
[17]   Cone-dominated ERG luminance-response function: The Photopic Hill revisited [J].
Rufiange M. ;
Rousseau S. ;
Dembinska O. ;
Lachapelle P. .
Documenta Ophthalmologica, 2002, 104 (3) :231-248
[18]   THE PHOTOPIC HILL - A NEW PHENOMENON OF THE LIGHT ADAPTED ELECTRORETINOGRAM [J].
WALI, N ;
LEGUIRE, LE .
DOCUMENTA OPHTHALMOLOGICA, 1992, 80 (04) :335-342
[19]   Directional sensitivity of the retina: 75 years of Stiles-Crawford effect [J].
Westheimer, Gerald .
PROCEEDINGS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 2008, 275 (1653) :2777-2786