In situ calibration of rotating sensor coils for magnet testing

被引:10
作者
Arpaia, P. [1 ,2 ]
Buzio, M. [2 ]
Golluccio, G. [1 ,2 ]
Walckiers, L. [2 ]
机构
[1] Univ Sannio, Dept Engn, I-82100 Benevento, Italy
[2] CERN, European Org Nucl Res, CH-1211 Geneva 23, Switzerland
关键词
FIELD MEASUREMENTS;
D O I
10.1063/1.3675578
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An in situ procedure for calibrating equivalent magnetic area and rotation radius of rotating coils is proposed for testing accelerator magnets shorter than the measuring coil. The procedure exploits measurements of magnetic field and mechanical displacement inside a reference quadrupole magnet. In a quadrupole field, an offset between the magnet and coil rotation axes gives rise to a dipole component in the field series expansion. The measurements of the focusing strength, the displacement, and the resulting dipole term allow the equivalent area and radius of the coil to be determined analytically. The procedure improves the accuracy of coils with large geometrical irregularities in the winding. This is essential for short magnets where the coil dimensions constrain the measurement accuracy. Experimental results on different coils measuring small-aperture permanent magnets are shown. (C) 2012 American Institute of Physics. [doi:10.1063/1.3675578]
引用
收藏
页数:7
相关论文
共 23 条
[1]  
[Anonymous], 2002008 JCGM
[2]   Digital integrator for fast accurate measurement of magnetic flux by rotating coils [J].
Arpaia, Pasquale ;
Masi, Alessandro ;
Spiezia, Giovanni .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (02) :216-220
[3]  
BILLAN J, 1985, ATOMKERNENERG/KERNT, V46, P193
[4]  
Buzio M., 2009, P 2009 CAS CERN ACC, P387, DOI [10.5170/CERN-2010-004.387, DOI 10.5170/CERN-2010-004.387]
[5]  
Cobb J., 1970, P 3 INT C MAGN TECHN
[6]  
Dunkel O., 2005, P 14 INT MAGN MEAS W
[7]   Design, construction, and calibration of a three-axis, high-frequency magnetic probe (B-dot probe) as a diagnostic for exploding plasmas [J].
Everson, E. T. ;
Pribyl, P. ;
Constantin, C. G. ;
Zylstra, A. ;
Schaeffer, D. ;
Kugland, N. L. ;
Niemann, C. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (11)
[8]  
Garoby R., 2006, P EUR PART ACC C ED
[9]   Calibration of an industrial pulsed field magnetometer [J].
Grössinger, R ;
Taraba, M ;
Wimmer, A ;
Dudding, J ;
Cornelius, R ;
Knell, R ;
Bissel, P ;
Enzberg-Mahlke, B ;
Fernengel, W ;
Toussaint, JC ;
Edwards, D .
IEEE TRANSACTIONS ON MAGNETICS, 2002, 38 (05) :2982-2984
[10]  
Henrichsen K. N., 2002, ENCY IMAGING SCI TEC, P970