A particle-into-liquid collector for rapid measurement of aerosol bulk chemical composition

被引:357
作者
Weber, RJ
Orsini, D
Daun, Y
Lee, YN
Klotz, PJ
Brechtel, F
机构
[1] Georgia Inst Technol, Dept Earth & Atmospher Sci, Atlanta, GA 30332 USA
[2] Brookhaven Natl Lab, Dept Environm Sci, New York, NY USA
关键词
D O I
10.1080/02786820152546761
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
We report on a new instrument developed for rapid automated on-line and continuous measurement of ambient aerosol bulk composition. The general approach is based on earlier devices (Khlystov et al. 1995; Simon and Dasgupta 1995) in which ambient particles are mixed with saturated water vapor to produce droplets easily collected by inertial techniques. The resulting liquid stream is an ah-zed with an ion chromatograph to quantitatively measure the bulk aerosol ionic components. In this instrument, a modified version of a particle size magnifier (Okuyama et al. 1984) is employed to activate and grow particles comprising the fine aerosol mass. A single jet inertial impactor is used to collect the droplets onto a vertical glass plate that is continually washed with a constant water diluent flow of nominally 0.10 ml min(-1). The flow is divided and then analyzed by a dual channel ion chromatograph. In its current form, 4.3 min integrated samples were measured every 7 min. The instrument provides bulk composition measurements with a detection limit of approximately 0.1 mug m(-3) for chloride, nitrate, sulfate, sodium, ammonium, calcium, and potassium. Results obtained during its deployment in the August 1999 Atlanta EPA Supersite study are presented.
引用
收藏
页码:718 / 727
页数:10
相关论文
共 26 条
[1]   DEVELOPMENT OF A SEMICONTINUOUS METHOD FOR THE MEASUREMENT OF NITRIC-ACID VAPOR AND PARTICULATE NITRATE AND SULFATE [J].
BUHR, SM ;
BUHR, MP ;
FEHSENFELD, FC ;
HOLLOWAY, JS ;
KARST, U ;
NORTON, RB ;
PARRISH, DD ;
SIEVERS, RE .
ATMOSPHERIC ENVIRONMENT, 1995, 29 (19) :2609-2624
[2]  
CARSON PG, 1995, J AEROSOL SCI, V26, P535, DOI 10.1016/0021-8502(94)00133-J
[3]   MEASUREMENT METHODS TO DETERMINE COMPLIANCE WITH AMBIENT AIR-QUALITY STANDARDS FOR SUSPENDED PARTICLES [J].
CHOW, JC .
JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 1995, 45 (05) :320-382
[4]  
DAHNEKE B, 1973, AEROSOL SCI, V4, P139
[5]  
HINZ KP, 1994, ANAL CHEM, V66, P2017
[6]   A continuous monitoring system for strong acidity in aerosols [J].
Ito, K ;
Chasteen, CC ;
Chung, HK ;
Poruthoor, SK ;
Zhang, GF ;
Dasgupta, PK .
ANALYTICAL CHEMISTRY, 1998, 70 (14) :2839-2847
[7]   Development of an aerosol mass spectrometer for size and composition analysis of submicron particles [J].
Jayne, JT ;
Leard, DC ;
Zhang, XF ;
Davidovits, P ;
Smith, KA ;
Kolb, CE ;
Worsnop, DR .
AEROSOL SCIENCE AND TECHNOLOGY, 2000, 33 (1-2) :49-70
[8]   Single-stage flowing liquid film impactor for continuous on-line particle analysis [J].
Karlsson, A ;
Irgum, K ;
Hansson, HC .
JOURNAL OF AEROSOL SCIENCE, 1997, 28 (08) :1539-1551
[9]   THE STEAM-JET AEROSOL COLLECTOR [J].
KHLYSTOV, A ;
WYERS, GP ;
SLANINA, J .
ATMOSPHERIC ENVIRONMENT, 1995, 29 (17) :2229-2234
[10]  
Knutson E. O., 1975, Journal of Aerosol Science, V6, P443, DOI 10.1016/0021-8502(75)90060-9