Strength analysis of clavicle fracture fixation devices and fixation techniques using finite element analysis with musculoskeletal force input

被引:24
作者
Marie, Cronskar [1 ]
机构
[1] Mid Sweden Univ, Dept Qual Mech & Math, S-83125 Ostersund, Sweden
关键词
Clavicle; Finite element analysis; Multibody simulation; Bone plates; Fixation techniques; PLATE FIXATION; BIOMECHANICAL EVALUATION; SIMULATION; STABILITY; NONUNIONS; SUPERIOR; SHOULDER;
D O I
10.1007/s11517-015-1288-5
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In the cases, when clavicle fractures are treated with a fixation plate, opinions are divided about the best position of the plate, type of plate and type of screw units. Results from biomechanical studies of clavicle fixation devices are contradictory, probably partly because of simplified and varying load cases used in different studies. The anatomy of the shoulder region is complex, which makes it difficult and expensive to perform realistic experimental tests; hence, reliable simulation is an important complement to experimental tests. In this study, a method for finite element simulations of stresses in the clavicle plate and bone is used, in which muscle and ligament force data are imported from a multibody musculoskeletal model. The stress distribution in two different commercial plates, superior and anterior plating position and fixation including using a lag screw in the fracture gap or not, was compared. Looking at the clavicle fixation from a mechanical point of view, the results indicate that it is a major benefit to use a lag screw to fixate the fracture. The anterior plating position resulted in lower stresses in the plate, and the anatomically shaped plate is more stress resistant and stable than a regular reconstruction plate.
引用
收藏
页码:759 / 769
页数:11
相关论文
共 46 条
[1]  
Ahmad M, 2006, J BONE JOINT SURG S3, V88, P401
[2]   A Parametric Study of Hard Tissue Injury Prediction Using Finite Elements: Consideration of Geometric Complexity, Subfailure Material Properties, CT-Thresholding, and Element Characteristics [J].
Arregui-Dalmases, Carlos ;
Del Pozo, Eduardo ;
Duprey, Sonia ;
Lopez-Valdes, Francisco J. ;
Lau, Anthony ;
Subit, Damien ;
Kent, Richard .
TRAFFIC INJURY PREVENTION, 2010, 11 (03) :286-293
[3]   Development of a finite element model of the shoulder: application during a side impact [J].
Astier, V. ;
Thollon, L. ;
Arnoux, P. J. ;
Mouret, F. ;
Brunet, Pr. C. .
INTERNATIONAL JOURNAL OF CRASHWORTHINESS, 2008, 13 (03) :301-312
[4]   Comparison of biomechanical behaviour of maxilla following Le Fort 1 osteotomy with 2-versus 4-plate fixation using 3D-FEA Part 2: Impaction surgery [J].
Atac, M. S. ;
Erkmen, E. ;
Yucel, E. ;
Kurt, A. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 38 (01) :58-63
[5]  
Bolte J.H., 2000, SAE Technical Paper
[6]   Complications of plate fixation of fresh displaced midclavicular fractures [J].
Bostman, O ;
Manninen, M ;
Pihlajamaki, H .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1997, 43 (05) :778-783
[7]   Biomechanical evaluation of clavicle fracture plating techniques: Does a locking plate provide improved stability? [J].
Celestre, Paul ;
Roberston, Claire ;
Mahar, Andrew ;
Oka, Richard ;
Meunier, Matthew ;
Schwartz, Alexandra .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (04) :241-247
[8]   Anterior-inferior plating of middle-third fractures of the clavicle [J].
Chen, Chin-En ;
Juhn, Rei-Jahn ;
Ko, Jih-Yang .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (04) :507-511
[9]   Anterior-inferior plate fixation of middle-third fractures and nonunions of the clavicle [J].
Collinge, Cory ;
Devinney, Scott ;
Herscovici, Dolfi ;
DiPasquale, Thomas ;
Sanders, Roy .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (10) :680-686
[10]   Modeling of fractured clavicles and reconstruction plates using CAD, finite element analysis and real musculoskeletal forces input [J].
Cronskar, M. ;
Backstrom, M. .