共 50 条
Slurry analysis after lead collection on a sorbent and its determination by electrothermal atomic absorption spectrometry
被引:7
|作者:
Baysal, Ash
[1
]
Tokman, Nigun
[1
]
Akman, Suleyman
[1
]
Ozeroglu, Cemal
[2
]
机构:
[1] Tech Univ Istanbul, Fac Sci & Letters, Dept Chem, TR-34469 Istanbul, Turkey
[2] Istanbul Univ, Fac Engn, Dept Chem, Istanbul, Turkey
关键词:
separation;
preconcentration;
lead;
electrothermal atomic absorption spectrometry;
slurry sampling;
D O I:
10.1016/j.jhazmat.2007.05.033
中图分类号:
X [环境科学、安全科学];
学科分类号:
08 ;
0830 ;
摘要:
In this study, in order to eliminate the drawbacks of elution step and to reach higher enrichment factors, a novel preconcentration/separation technique for the slurry analysis of sorbent loaded with lead prior to its determination by electrothermal atomic absorption spectrometry was described. For this purpose, at first, lead was collected on ethylene glycol dimethacrylate methacrylic acid copolymer (EGDMA-MA) treated with ammonium pyrolidine dithiocarbamate (APDQ by conventional batch technique. After separation of liquid phase, slurry of the sorbent was prepared and directly pipetted into graphite furnace of atomic absorption spectrophotometer. Optimum conditions for quantitative sorption and preparation of the slurry were investigated. A 100-fold enrichment factor could be easily reached. The analyte element in certified sea-water and Bovine-liver samples was determined in the range of 95% confidence level. The proposed technique was fast and simple and the risks of contamination and analyte loss were low. Detection limit (33) for Pb was 1.67 mu g l(-1). (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:804 / 808
页数:5
相关论文