Independent measurements of 129I content in environmental reference materials using accelerator and thermal ionization mass spectrometry

被引:6
作者
Olson, J. E. [1 ]
Adamic, M. L. [1 ]
Snyder, D. C. [1 ]
Brookhart, J. L. [1 ]
Hahn, P. A. [1 ]
Watrous, M. G. [1 ]
机构
[1] Idaho Natl Lab, POB 1625, Idaho Falls, ID 83415 USA
关键词
Accelerator mass spectrometry; AMS; Reference materials; TIMS; Iodine; 129I; Standards; SOIL SAMPLES; IODINE; AMS; IAEA-375;
D O I
10.1016/j.nimb.2018.07.018
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Iodine environmental measurement programs are in need of new materials with certified I-129 activity. Frequently I-129 measurements are validated in the literature using the standard material IAEA-375, Chernobyl soil, which is the only soil/sediment material with a recommended I-129 activity. IAEA-375 has not been available for purchase since 2010. This study is an extension of previous work at INL to include four additional standard materials that are commercially available (NIST materials: RM 8704, Buffalo River sediment, SRM 2710a, Montana I soil, and IAEA materials: SL-1, lake sediment, IAEA-385, Irish Sea sediment). These materials have certified or recommended activities for a variety of radionuclides but not for I-129. This paper reports a comparison of accelerator mass spectrometry (AMS) and thermal ionization mass spectrometry (TIMS) data for I-129 activity, as well as the I-129/127 ratios for these standards to assist in identifying a suitable alternative for IAEA-375. Two independent chemical separation and mass spectrometric analysis techniques have been applied in an effort to corroborate the data. Both methods were validated via analyses of IAEA-375 for I-129 and show good agreement with the recommended activity of 1.7 x 10(-3) Bq kg(-1) for I-129; (1.6 x 10(-3) Bq kg(-1) by AMS and 1.8 x 10(-3) Bq kg(-1) by TIMS) with both sets of results within the 95% confidence interval of the recommended value.
引用
收藏
页码:84 / 88
页数:5
相关论文
共 50 条
  • [31] Accelerator Mass Spectrometry Analysis of Ultra-Low-Level 129I in Carrier-Free AgI-AgCl Sputter Targets
    Liu, Qi
    Hou, Xiaolin
    Zhou, Weijian
    Fu, Yunchong
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2015, 26 (05) : 725 - 733
  • [32] Impurity measurements in semiconductor materials using trace element accelerator mass spectrometry
    McDaniel, FD
    Datar, SA
    Nigam, M
    Prasad, GVR
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2002, 190 : 826 - 830
  • [33] On the analysis of iodine-129 and iodine-127 in environmental materials by accelerator mass spectrometry and ion chromatography
    Schmidt, A
    Schnabel, C
    Handl, J
    Jakob, D
    Michel, R
    Synal, HA
    Lopez, JM
    Suter, M
    SCIENCE OF THE TOTAL ENVIRONMENT, 1998, 223 (2-3) : 131 - 156
  • [34] The French accelerator mass spectrometry facility ASTER after 4 years: Status and recent developments on 36Cl and 129I
    Arnold, Maurice
    Aumaitre, Georges
    Bourles, Didier L.
    Keddadouche, Karim
    Braucher, Regis
    Finkel, Robert C.
    Nottoli, Emmanuelle
    Benedetti, Lucilla
    Merchel, Silke
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 294 : 24 - 28
  • [35] Improved techniques for high accuracy isotope ratio measurements of nuclear materials using thermal ionization mass spectrometry
    Richter, S
    Goldberg, SA
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2003, 229 (03) : 181 - 197
  • [36] Determination of Low Level 129I in Soil Samples Using Coprecipitation Separation of Carrier Free Iodine and Accelerator Mass Spectrometric Measurement
    Luo Mao-Yi
    Zhou Wei-Jian
    Hou Xiao-Lin
    Chen Ning
    Zhang Lu-Yuan
    Liu Qi
    He Chao-Hui
    Fan Yu-Kun
    Liang Wang-Guo
    Wang Zhi-Wen
    Fu Yun-Chong
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 39 (02) : 193 - 197
  • [37] Matrix effects in plutonium isotope ratio measurements using thermal ionization mass spectrometry
    Mathew, Kattathu Joseph
    Ottenfeld, Chelsea
    Levesque, Steven
    Keller, Russ
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2022, 331 (02) : 1005 - 1017
  • [38] Cadmium isotopic composition of biogenic certified reference materials determined by thermal ionization mass spectrometry with double spike correction
    Borovicka, Jan
    Ackerman, Lukas
    Rejsek, Jan
    TALANTA, 2021, 221
  • [39] Influence of cathode materials on BeO currents in cosmogenic 10 Be measurements using a SNICS ion source and accelerator mass spectrometry
    Nakamura, Atsunori
    Yamagata, Takeyasu
    Matsuzaki, Hiroyuki
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2025, 558
  • [40] Isotopic Analysis of Uranium and Plutonium at Ultra Trace Levels Using Thermal Ionization Mass Spectrometry with Simultaneous Measurements
    Park, Jong-Ho
    Jeong, Kahee
    Song, Kyuseok
    ASIAN JOURNAL OF CHEMISTRY, 2013, 25 (12) : 7061 - 7063