VDDR pacing after His-bundle ablation for paroxysmal atrial fibrillation: A pilot study

被引:4
作者
Buys, EM [1 ]
van Hemel, NM [1 ]
Jessurun, ER [1 ]
Kelder, JC [1 ]
Bakema, L [1 ]
Kingma, JH [1 ]
机构
[1] St Antonius Hosp, Dept Cardiol, NL-3430 EM Nieuwegein, Netherlands
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 10期
关键词
paroxysmal atrial fibrillation; VDDR pacing; intraatrial electrogram; His-bundle ablation;
D O I
10.1111/j.1540-8159.1998.tb00004.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
His-bundle ablation followed by pacemaker implantation is today a widely accepted therapeutic choice when drug refractoriness of symptomatic AF is evident. The selection of pacing mode in patients suffering from paroxysmal AF is still controversial. Preservation of AV synchrony is an attractive option in patients with paroxysmal AF who undergo His-bundle ablation. The purpose of this study was to examine prospectively the contribution of VDDR pacing for preservation of AV synchrony. After His-bundle ablation a VDDR pacing system was implanted in 17 patients with paroxysmal AF and all antiarrhythmic drugs were withdrawn. The endpoint of the study was defined as the onset of chronic AF, To document the onset of chronic AF 48-hour Holter recordings were made every 6-8 weeks. After a mean followup of 18.2 (range 14-21) months, VDDR pacing is still operative in 13 patients (77%). Four patients developed chronic AF after a mean follow-up of 6 months. Of several baseline characteristics, only the intraatrial P wave at implantation was significantly smaller in patients developing chronic AF than in patients in whom the VDDR mode is still operative. This pilot study suggests that VDDR pacing is an attractive pacing method for patients with paroxysmal AF after His-bundle ablation. A low intraatrial P wave electrogram at implant was associated with a higher risk for the development of chronic AF.
引用
收藏
页码:1869 / 1872
页数:4
相关论文
共 31 条
  • [1] ANSINI L, 1994, PACE, V17, P1865
  • [2] INFLUENCE OF ATRIOVENTRICULAR JUNCTION RADIOFREQUENCY ABLATION IN PATIENTS WITH CHRONIC ATRIAL-FIBRILLATION AND FLUTTER ON QUALITY-OF-LIFE AND CARDIAC-PERFORMANCE
    BRIGNOLE, M
    GIANFRANCHI, L
    MENOZZI, C
    BOTTONI, N
    BOLLINI, R
    LOLLI, G
    ODDONE, D
    GAGGIOLI, G
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (03) : 242 - 246
  • [3] A NEW PACEMAKER FOR PAROXYSMAL ATRIAL-FIBRILLATION TREATED WITH RADIOFREQUENCY ABLATION OF THE AV JUNCTION
    BRIGNOLE, M
    GIANFRANCHI, L
    MENOZZI, C
    BOTTONI, N
    BOLLINI, R
    LOLLI, G
    ODDONE, D
    GAGGIOLI, G
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 1889 - 1894
  • [4] UNSATISFYING RESULTS IN LONG-TERM ATRIAL-PACING WITH A BIPOLAR ACTIVE FIXATION ATRIAL LEAD
    BUYS, EM
    VANHEMEL, NM
    KELDER, JC
    VANDESSEL, PFHM
    BAKEMA, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (12): : 2292 - 2296
  • [5] Exercise capacity after His bundle ablation and rate response ventricular pacing for drug refractory chronic atrial fibrillation
    Buys, EM
    vanHemel, NM
    Keider, JC
    Ascoop, CAPL
    vanDessel, PFHM
    Bakema, L
    Kingma, JH
    [J]. HEART, 1997, 77 (03) : 238 - 241
  • [6] CAPUCCI A, 1997, PACE, V20, P1228
  • [7] EDNER M, 1995, BRIT HEART J, V74, P261
  • [8] GARRIGUE S, 1997, PACE, V20, P1092
  • [9] CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD
    GLIKSON, M
    VONFELDT, LK
    SUMAN, VJ
    HAYES, DL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08): : 1399 - 1404
  • [10] GROSS JN, 1994, PACE, V17, P769