Comparison of Accuracy of Intraocular Lens Calculations Using Automated Keratometry, a Placido-Based Corneal Topographer, and a Combined Placido-Based and Dual Scheimpflug Corneal Topographer

被引:2
作者
Shirayama, Mariko [1 ,2 ]
Wang, Li [1 ]
Koch, Douglas D. [1 ]
Weikert, Mitchell P. [1 ]
机构
[1] Baylor Coll Med, Dept Ophthalmol, Houston, TX 77030 USA
[2] Univ Tokyo, Dept Ophthalmol, Tokyo 113, Japan
关键词
IOL power calculation; Galilei dual Scheimpflug analyzer; corneal anterior and posterior corneal surface; corneal power; POWER; ERROR;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the accuracy of intraocular lens (IOL) power calculations with automated keratometry, a Placido-based corneal topographer, and a combined Placido-based and dual Scheimpflug corneal topographer. Methods: Retrospectively, 75 eyes of 62 patients who had phacoemulsification with implantation of the SN60WF IOL were analyzed. Corneal powers were measured using 5 techniques: (1) automated keratometry (AutoK, IOL Master), (2) simulated keratometry from the Placido-based corneal topographer (SimK(P), Atlas), (3) simulated keratometry from the combined Placido-based and dual Scheimpflug corneal topographer (SimK(P+DS), Galilei), (4) total corneal power of the steep and flat meridians over a central 1.0-4.0 mm zone using ray tracing (TCPMeridian, Galilei), and (5) TCP over the central 4.0-mm zone by ray tracing (TCPCentral, Galilei). For each IOL implanted, prediction errors were determined by comparing the predicted refractions calculated with the Holladay 1 formula to the actual refraction that was obtained 3-4 weeks postoperatively. The surgeon factor was optimized for each method of corneal power measurement. The accuracy of IOL power calculation using these corneal powers was determined by calculating the mean absolute prediction error (MAE) and the percentage of eyes with prediction errors within 0.5, 1.0, and 1.5 diopters (D). Result: The MAEs were 0.37, 0.39, 0.39, 0.41, and 0.42 D for the AutoK, SimK(P), SimK(P+DS), TCPMeridian, and TCPCentral methods, respectively. There were no significant differences among groups. Over 93% of eyes had MAEs within 1.0 D for all methods of corneal power measurement. Conclusions: The accuracy of IOL power calculation was comparable with AutoK, the Placido-based corneal topographer, and the combined Placido-based and dual Scheimpflug corneal topographer.
引用
收藏
页码:1136 / 1138
页数:3
相关论文
共 10 条
  • [1] Changes in corneal asphericity after laser in situ keratomileusis
    Anera, RG
    Jiménez, JR
    del Barco, LJ
    Bermúdez, J
    Hita, E
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (04) : 762 - 768
  • [2] COMPARISON OF THE ACCURACY OF COMPUTERIZED VIDEOKERATOGRAPHY AND KERATOMETRY FOR USE IN INTRAOCULAR-LENS CALCULATIONS
    CUAYCONG, MJ
    GAY, CA
    EMERY, J
    HAFT, EA
    KOCH, DD
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 : 178 - 181
  • [3] Computerized videokeratography and keratometry in determining intraocular lens calculations
    Husain, SE
    Kohnen, T
    Maturi, R
    Er, H
    Koch, DD
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (03) : 362 - 366
  • [4] Consideration of the posterior corneal curvature for assessment of corneal power after myopic LASIK
    Langenbucher, A
    Torres, F
    Behrens, A
    Suarez, E
    Haigis, W
    Seitz, B
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2004, 82 (03): : 264 - 269
  • [5] Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems
    Menassa, Nardine
    Kaufmann, Claude
    Goggin, Michael
    Job, Oliver M.
    Bachmann, Lucas M.
    Thiel, Michael A.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (10) : 1742 - 1747
  • [6] Sources of error in intraocular lens power calculation
    Norrby, Sverker
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (03) : 368 - 376
  • [7] SOURCES OF ERROR IN INTRAOCULAR-LENS POWER CALCULATION
    OLSEN, T
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1992, 18 (02) : 125 - 129
  • [8] Scheimpflug photography keratometry readings for routine intraocular lens power calculation
    Shammas, H. John
    Hoffer, Kenneth J.
    Shammas, Maya C.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (02) : 330 - 334
  • [9] Comparison of Corneal Powers Obtained from 4 Different Devices
    Shirayama, Mariko
    Wang, Li
    Weikert, Mitchell P.
    Koch, Douglas D.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (04) : 528 - 535
  • [10] Response of the posterior in situ keratomileusis for corneal surface to laser myopia
    Twa, MD
    Roberts, C
    Mahmoud, AM
    Chang, JS
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) : 61 - 71