The biomechanics of a validated finite element model of stress shielding in a novel hybrid total knee replacement

被引:25
作者
Bougherara, H. [2 ]
Zdero, R. [1 ,2 ]
Mahboob, Z. [3 ]
Dubov, A. [1 ,2 ]
Shah, S. [1 ,2 ]
Schemitsch, E. H. [1 ,4 ]
机构
[1] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[2] Ryerson Univ, Dept Mech & Ind Engn, Toronto, ON, Canada
[3] Ryerson Univ, Dept Aerosp Engn, Toronto, ON, Canada
[4] Univ Toronto, Dept Surg, Toronto, ON, Canada
关键词
biomechanics; stress shielding; finite element; Duracon knee; hybrid; BONE-MINERAL DENSITY; DISTAL FEMUR; FOLLOW-UP; MECHANICAL VALIDATION; SYNTHETIC FEMURS; FEMORAL STEMS; ARTHROPLASTY; PROSTHESIS; DESIGN; OSTEOLYSIS;
D O I
10.1243/09544119JEIM691
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study proposes a novel hybrid total knee replacement (TKR) design to improve stress transfer to bone in the distal femur and, thereby, reduce stress shielding and consequent bone loss. Three-dimensional finite element (FE) models were developed for a standard and a hybrid TKR and validated experimentally. The Duracon knee system (Stryker Canada) was the standard TKR used for the FE models and for the experimental tests. The FE hybrid device was identical to the standard TKR, except that it had an interposing layer of carbon fibre-reinforced polyamide 12 lining the back of the metallic femoral component. A series of experimental surface strain measurements were then taken to validate the FE model of the standard TKR at 3000 N of axial compression and at 0 degrees of knee flexion. Comparison of surface strain values from FE analysis with experiments demonstrated good agreement, yielding a high Pearson correlation coefficient of R-2 = 0.94. Under a 3000 N axial load and knee flexion angles simulating full stance (0 degrees), heel strike (20 degrees), and toe off (60 degrees) during normal walking gait, the FE model showed considerable changes in maximum Von Mises stress in the region most susceptible to stress shielding (i.e. the anterior region, just behind the flange of the femoral implant). Specifically, going from a standard to a hybrid TKR caused an increase in maximum stress of 87.4 per cent (0 degrees; from 0.15 to 0.28 MPa), 68.3 per cent (20 degrees; from 1.02 to 1.71 MPa), and 12.6 per cent (60 degrees; from 2.96 to 3.33 MPa). This can potentially decrease stress shielding and subsequent bone loss and knee implant loosening. This is the first report to propose and biomechanically to assess a novel hybrid TKR design that uses a layer of carbon fibre-reinforced polyamide 12 to reduce stress shielding.
引用
收藏
页码:1209 / 1219
页数:11
相关论文
共 57 条
[1]  
ANGELIDES M, 1988, T ORTHOP RES SOC, V13, P475
[2]   Contribution of loading conditions and material properties to stress shielding near the tibial component of total knee replacements [J].
Au, Anthony G. ;
Raso, V. James ;
Liggins, A. B. ;
Amirfazli, A. .
JOURNAL OF BIOMECHANICS, 2007, 40 (06) :1410-1416
[3]   Corrosion aspects of metallic implants - An overview [J].
Balamurugan, A. ;
Rajeswari, S. ;
Balossier, G. ;
Rebelo, A. H. S. ;
Ferreira, J. M. F. .
MATERIALS AND CORROSION-WERKSTOFFE UND KORROSION, 2008, 59 (11) :855-869
[4]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[5]  
Bartel DonaldL., 2006, Orthopaedic Biomechanics: Mechanics and Design in Musculoskeletal Systems, V1
[6]  
BOMBELLI R, 1982, J ROY SOC MED, V75, P588
[7]   The biomechanics of the T2 femoral nailing system: a comparison of synthetic femurs with finite element analysis [J].
Bougherara, H. ;
Zdero, R. ;
Miric, M. ;
Shah, S. ;
Hardisty, M. ;
Zalzal, P. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H3) :303-314
[8]   Design of a biomimetic polymer-composite hip prosthesis [J].
Bougherara, Habiba ;
Bureau, Martin ;
Campbell, Melissa ;
Vadean, Aurelian ;
Yahia, L'Hocine .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2007, 82A (01) :27-40
[9]  
Cadambi A, 1994, J Arthroplasty, V9, P579, DOI 10.1016/0883-5403(94)90111-2
[10]  
CAMPBELL M, 2006, MED DEVICE MAT, V3, P49