Dual-plane ultrafast limited-angle electron beam x-ray tomography

被引:27
作者
Bieberle, M. [1 ]
Schleicher, E. [1 ]
Fischer, F. [1 ]
Koch, D. [2 ]
Menz, H. -J. [2 ]
Mayer, H. -G. [2 ]
Hampel, U. [1 ]
机构
[1] Forschungszentrum Dresden Rossendorf eV, Inst Safety Res, D-01314 Dresden, Germany
[2] Univ Stuttgart, Inst Nucl Energy Technol & Energy Syst, D-70569 Stuttgart, Germany
关键词
Electron beam x-ray CT; Limited-angle CT; Wire-mesh sensor; Two-phase flow; COMPUTED-TOMOGRAPHY; FLOW; SENSOR; PIPE;
D O I
10.1016/j.flowmeasinst.2009.12.001
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Electron beam x-ray tomography is an imaging technique, which can provide cross-sectional images of an object of interest with about 1 mm spatial resolution at frame rates of up to 10,000 frames per second. As a non-intrusive method it is especially suited for studying multiphase flows. For this purpose we devised an experimental limited-angle scan setup which utilizes linear beam deflection to generate radiographic projections. This setup was employed in the study of gas-liquid flow in an experimental flow loop operated at different liquid and gas flow rates. Electron-beam tomography images were compared with image data of a wire-mesh sensor. The latter is a fast but intrusive imaging device which is commonly used in gas-liquid flow imaging and achieves comparable frame rates but at lower spatial resolution. As a novelty we implemented a dual-plane limited-angle electron beam x-ray tomography which allows us to gain information about the phase velocities using cross-correlation data analysis. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:233 / 239
页数:7
相关论文
共 13 条
[1]   Evaluation of a limited angle scanned electron beam x-ray CT approach for two-phase pipe flows [J].
Bieberle, M. ;
Hampel, U. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2006, 17 (08) :2057-2065
[2]   Ultrafast limited-angle-type x-ray tomography [J].
Bieberle, M. ;
Fischer, F. ;
Schleicher, E. ;
Hampel, U. ;
Koch, D. ;
Aktay, K. S. D. C. ;
Menz, H.-J. ;
Mayer, H.-G. .
APPLIED PHYSICS LETTERS, 2007, 91 (12)
[3]   Simulation study on electron beam x-ray CT arrangements for two-phase flow measurements [J].
Bieberle, M. ;
Schleicher, E. ;
Hampel, U. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2008, 19 (09)
[4]  
BIEBERLE M, P INT C MULT FLOW S7
[5]   CARDIAC COMPUTED-TOMOGRAPHY [J].
BOYD, DP ;
LIPTON, MJ .
PROCEEDINGS OF THE IEEE, 1983, 71 (03) :298-307
[6]   ALGEBRAIC RECONSTRUCTION TECHNIQUES (ART) FOR 3-DIMENSIONAL ELECTRON MICROSCOPY AND X-RAY PHOTOGRAPHY [J].
GORDON, R ;
BENDER, R ;
HERMAN, GT .
JOURNAL OF THEORETICAL BIOLOGY, 1970, 29 (03) :471-&
[7]   Experimental ultra fast X-ray computed tomography with a linearly scanned electron beam source [J].
Hampel, U ;
Speck, M ;
Koch, D ;
Menz, HJ ;
Mayer, HG ;
Fietz, J ;
Hoppe, D ;
Schleicher, E ;
Zippe, C ;
Prasser, HM .
FLOW MEASUREMENT AND INSTRUMENTATION, 2005, 16 (2-3) :65-72
[8]  
Hori K., 2000, Heat Transfer-Asian Research, V29, P155, DOI 10.1002/(SICI)1523-1496(200005)29:3<155::AID-HTJ1>3.0.CO
[9]  
2-M
[10]   A dual sensor flow imaging tomographic system [J].
Johansen, GA ;
Froystein, T ;
Hjertaker, BT ;
Olsen, O .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (03) :297-307