Electric field distribution analysis for the design of an electrode system in a 3D neuromuscular junction microfluidic device

被引:0
|
作者
Forconi, Flavia [1 ]
Apa, Ludovica [2 ]
D'Alvia, Livio [2 ]
Cosentino, Marianna [1 ]
Rizzuto, Emanuele [2 ]
Del Prete, Zaccaria [2 ]
机构
[1] Sapienza Univ Rome, DAHFMO Unit Histol & Med Embriol, Rome, Italy
[2] Sapienza Univ Rome, Dept Mech & Aerosp Engn, Rome, Italy
来源
2021 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (IEEE MEMEA 2021) | 2021年
关键词
electrical stimulation; electrodes; electric field; microfluidic device; neuromuscular junction; STIMULATION; CELLS;
D O I
10.1109/MeMeA52024.2021.9478775
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electrical stimulation (ES) highly influences the cellular microenvironment, affecting cell migration, proliferation and differentiation. It also plays a crucial role in tissue engineering to improve the biomechanical properties of the constructs and regenerate the damaged tissues. However, the effects of the ES on the neuromuscular junction (NMJ) are still not fully analyzed. In this context, the development of a specialized microfluidic device combined with an ad-hoc electrical stimulation can allow a better investigation of the NMJ functionality. To this aim, we performed an analysis of the electric field distribution in a 3D neuromuscular junction microfluidic device for the design of several electrode systems. At first, we designed and modeled the 3D microfluidic device in order to promote the formation of the NMJ between neuronal cells and the muscle engineered tissue. Subsequently, with the aim of identifying the optimal electrode configuration able to properly stimulate the neurites, thus enhancing the formation of the NMJ, we performed different simulation tests of the electric field distribution, by varying the electrode type, size, position and applied voltage. Our results revealed that all the tested configurations did not induce an electric field dangerous for the cell vitality. Among these configurations, the one with cylindrical pin of 0.3 mm of radius, placed in the internal position of the neuronal chambers, allowed to obtain the highest electrical field in the zone comprising the neurites.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] A microfluidic device for 2D to 3D and 3D to 3D cell navigation
    Shamloo, Amir
    Amirifar, Leyla
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2016, 26 (01)
  • [2] Numerical design of a 3D microfluidic systems for bioparticle analysis: electrostatic and dielectrophoretic simulations
    Corman, Ramona
    Nedelcu, Oana
    Ravariu, Cristian
    2016 39TH INTERNATIONAL SEMICONDUCTOR CONFERENCE (CAS), 2016, : 187 - 190
  • [3] A new 3D mass diffusion–reaction model in the neuromuscular junction
    Abdul Khaliq
    Frank Jenkins
    Mark DeCoster
    Weizhong Dai
    Journal of Computational Neuroscience, 2011, 30 : 729 - 745
  • [4] Development of a 3D Graphene Electrode Dielectrophoretic Device
    Xie, Hongyu
    Tewari, Radheshyam
    Fukushima, Hiroyuki
    Narendra, Jeffri
    Heldt, Caryn
    King, Julia
    Minerick, Adrienne R.
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2014, (88):
  • [5] Novel 3D integrated microfluidic device: Design, construction, and application to the preparation of liposomes for vaccine delivery
    Xiu, Xueliang
    Ma, Shichao
    Li, Zhiyong
    Li, Zhipeng
    Li, Xuemei
    Ruan, Yangtao
    Li, Dezhao
    Wang, Jianwei
    Ma, Fengsen
    JOURNAL OF DRUG DELIVERY SCIENCE AND TECHNOLOGY, 2024, 91
  • [6] A new 3D mass diffusion-reaction model in the neuromuscular junction
    Khaliq, Abdul
    Jenkins, Frank
    DeCoster, Mark
    Dai, Weizhong
    JOURNAL OF COMPUTATIONAL NEUROSCIENCE, 2011, 30 (03) : 729 - 745
  • [7] Spatiotemporal Image Correlation Analysis for 3D Flow Field Mapping in Microfluidic Devices
    Ceffa, Nicolo G.
    Bouzin, Margaux
    D'Alfonso, Laura
    Sironi, Laura
    Marquezin, Cassia A.
    Auricchio, Ferdinando
    Marconi, Stefania
    Chirico, Giuseppe
    Collini, Maddalena
    ANALYTICAL CHEMISTRY, 2018, 90 (03) : 2277 - 2284
  • [8] Development of 2D and 3D functional platforms for integrated neuromuscular junction studies
    Prime, Stuart
    Springe, Signe
    Paredes, Amaia
    Wallbank, David
    JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 2023, 123
  • [9] 3D electric field analysis of needleless electrospinning from a ring coil
    Wang, Xin
    Wang, Xungai
    Lin, Tong
    JOURNAL OF INDUSTRIAL TEXTILES, 2014, 44 (03) : 463 - 476
  • [10] Design, Fabrication and Characterization of a Low-Impedance 3D Electrode Array System for Neuro-Electrophysiology
    Kusko, Mihaela
    Craciunoiu, Florea
    Amuzescu, Bogdan
    Halitzchi, Ferdinand
    Selescu, Tudor
    Radoi, Antonio
    Popescu, Marian
    Simion, Monica
    Bragaru, Adina
    Ignat, Teodora
    SENSORS, 2012, 12 (12): : 16571 - 16590