Aerosol deposition and clearance measurement: a novel technique using dynamic SPET

被引:30
作者
Eberl, S
Chan, HK
Daviskas, E
Constable, C
Young, I
机构
[1] Royal Prince Alfred Hosp, Dept PET & Nucl Med, Camperdown, NSW 2050, Australia
[2] Univ Sydney, Fac Pharm, Sydney, NSW 2006, Australia
[3] Royal Prince Alfred Hosp, Dept Resp Med, Camperdown, NSW 2050, Australia
关键词
dynamic SPET; radioaerosol deposition; radioaerosol clearance; penetration index;
D O I
10.1007/s002590100586
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Planar gamma camera scintigraphy is a well-established technique for characterising the deposition and clearance of radiolabelled aerosols. While single-photon emission tomography (SPET) can offer superior assessment of radioaerosol deposition and better differentiation between peripheral and central deposition, the long acquisition times of single-headed SPET have largely prevented its use for measuring clearance or deposition of fast-clearing radioaerosols. This study investigated the feasibility of fast dynamic SPET imaging (1 min/frame) using a three-headed gamma camera to assess the regional and total deposition and clearance of different radioaerosols over a period of 26 min. Six subjects inhaled nebulised technetium-99m diethylene triamine penta-acetic acid radiolabelled aerosols with small and large droplet sizes [mass median aerodynamic diameter (MMAD) 3.2 +/-0.2 and 6.5 +/-0.2 mum, span 1.8 and 1.7, respectively] and in normal (0.9%) or hypertonic (7%) saline with controlled breathing on four separate occasions. The penetration indices (Pls) calculated from the SPET data for normal saline were 0.50 +/-0.04 and 0.36 +/-0.02 for the small and large droplet sizes, respectively. Consistent with the hygroscopic growth of the hypertonic aerosols, the PIs for hypertonic saline were lower, at 0.43 +/-0.02 and 0.34 +/-0.02 for the small and large droplet sizes, respectively. PIs calculated from the planar data showed similar trends, but failed to detect the significant difference seen with SPET between small normal and small hypertonic saline radioaerosols. In conclusion, the feasibility of using fast dynamic SPET for imaging radioaerosol deposition and associated radiolabel clearance in the lung has been successfully demonstrated. The fast SPET was able to reveal important differences in aerosol deposition that were not detected by planar imaging.
引用
收藏
页码:1365 / 1372
页数:8
相关论文
共 19 条
  • [1] Regional distribution and kinetics of inhaled pharmaceuticals
    Berridge, MS
    Lee, ZH
    Heald, DL
    [J]. CURRENT PHARMACEUTICAL DESIGN, 2000, 6 (16) : 1631 - 1651
  • [2] REGIONAL DEPOSITION OF NEBULIZED HYPODENSE NONISOTONIC SOLUTIONS IN THE HUMAN RESPIRATORY-TRACT
    CHAN, HK
    PHIPPS, PR
    GONDA, I
    COOK, P
    FULTON, R
    YOUNG, I
    BAUTOVICH, G
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1994, 7 (08) : 1483 - 1489
  • [3] Deposition of aqueous aerosol of technetium-99m diethylene triamine penta-acetic acid generated and delivered by a novel system (AERx) in healthy subjects
    Chan, HK
    Daviskas, E
    Eberl, S
    Robinson, M
    Bautovich, G
    Young, I
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (04) : 320 - 327
  • [4] USE OF SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY IN AEROSOL STUDIES
    CHAN, HK
    [J]. JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1993, 6 (01): : 23 - 36
  • [5] CHAN HK, 1999, J AEROSOL MED, V12, P135
  • [6] Eberl S, 1996, J NUCL MED, V37, P137
  • [7] Regional deposition of inhaled hygroscopic aerosols: In vivo SPECT compared with mathematical modeling
    Finlay, WH
    Stapleton, KW
    Chan, HK
    Zuberbuhler, P
    Gonda, I
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1996, 81 (01) : 374 - 383
  • [8] Evaluation of the accuracy and precision of lung aerosol deposition measurements from single-photon emission computed tomography using simulation
    Fleming, JS
    Sauret, V
    Conway, JH
    Holgate, ST
    Bailey, AG
    Martonen, TB
    [J]. JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2000, 13 (03): : 187 - 198
  • [9] A technique for simulating radionuclide images from the aerosol deposition pattern in the airway tree
    Fleming, JS
    Hashish, AH
    Conway, JH
    HartleyDavies, R
    Nassim, MA
    Guy, MJ
    Coupe, J
    Holgate, ST
    Moore, E
    Bailey, AG
    Martonen, TB
    [J]. JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1997, 10 (03): : 199 - 212
  • [10] GULLBERG GT, 1995, NUCL CARDIOLOGY STAT, P137