Laboratory characterization of obligate carriers of type 3 von Willebrand disease with a potential role for Platelet Function Analyzer (PFA-200)

被引:2
|
作者
Geevar, Tulasi [1 ]
Dave, Rutvi Gautam [1 ]
Mathews, Nitty Skariah [1 ]
Kumar, Snehil [1 ]
Mammen, Joy John [1 ]
Vijayan, Ramya [1 ]
Aboobacker, Fouzia N. [2 ]
Abraham, Aby [2 ]
Srivastava, Alok [2 ]
Nair, Sukesh Chandran [1 ]
机构
[1] Christian Med Coll & Hosp, Dept Transfus Med & Immunohematol, Vellore 632004, Tamil Nadu, India
[2] Christian Med Coll & Hosp, Dept Clin Hematol, Vellore, Tamil Nadu, India
关键词
carrier; PFA-100; 200; Type 3 von Willebrand disease; von Willebrand disease; von Willebrand factor; HETEROZYGOUS CARRIERS; HEMORRHAGIC SYMPTOMS; FACTOR DEFICIENCY; CLINICAL UTILITY; BLEEDING HISTORY; BLOOD-GROUP; DIAGNOSIS; MANAGEMENT; RISK; ABO;
D O I
10.1111/ijlh.13787
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Type 3 von Willebrand disease (VWD) is a rare autosomal recessive disorder characterized by undetectable von Willebrand Antigen (VWF:Ag). Carriers of type 3 VWD carry one null allele and have von Willebrand factor (VWF) at about 50% of normal. The aim of this study was to characterize type 3 VWD carriers and to study the role of Platelet Function Analyzer (PFA-200) in this cohort. Methods This was a cross-sectional study where data were collected from carriers (parents/offspring) of type 3 VWD patients and evaluated with activated partial thromboplastin time, factor VIII, blood group, ristocetin cofactor assay (VWF:RCo), VWF:Ag, and closure time on PFA-200 with collagen/epinephrine (COL/EPI), and collagen/ADP (COL/ADP). Results One hundred carriers were included in the study of which 85 were included for PFA-200 analysis. The mean (SD) of VWF:Ag (IU/ml) and VWF:RCo (IU/ml) was 0.63 (0.24) and 0.61 (0.26), respectively. Among the 100 carriers, based on VWF levels (VWF:Ag and/or VWF:RCo) and bleeding history, there were 7 type 1 VWD, 10 type 2 VWD, 25 borderline VWF (0.30-0.50 IU/ml and no bleeding), and 58 normal VWF (>0.50 IU/ml). PFA-200 was prolonged in 71% of the carriers, all carriers with type 1 and type 2 VWD phenotype, 80% carriers with borderline VWF, and 59% with normal VWF. COL/EPI was more sensitive than COL/ADP and showed better correlation with VWF parameters than COL/ADP. Conclusion Carriers of type 3 VWD can have a variable laboratory phenotype. PFA-200 showed good sensitivity among the carriers at VWF levels <0.50 IU/ml.
引用
收藏
页码:603 / 609
页数:7
相关论文
共 33 条
  • [1] Utility of the platelet function analyser (PFA-100/200) for exclusion or detection of von Willebrand disease: A study 22 years in the making
    Favaloro, Emmanuel J.
    THROMBOSIS RESEARCH, 2020, 188 : 17 - 24
  • [2] Description of an In Vitro Platelet Function Analyzer (PFA-100/PFA-200) 30 Years in the Making
    Favaloro, Emmanuel J.
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2024, 50 (02) : 320 - 324
  • [3] The Platelet Function Analyzer (PFA-100) may not be suitable for monitoring the therapeutic efficiency of von Willebrand concentrate in type III von Willebrand disease
    A. Meskal
    F. Vertessen
    M. Van der Planken
    Z. N. Berneman
    Annals of Hematology, 1999, 78 : 426 - 430
  • [4] The Platelet Function Analyzer (PFA-100) may not be suitable for monitoring the therapeutic efficiency of von Willebrand concentrate in type III von Willebrand disease
    Meskal, A
    Vertessen, F
    Van der Planken, M
    Berneman, ZN
    ANNALS OF HEMATOLOGY, 1999, 78 (09) : 426 - 430
  • [5] Potential supplementary utility of combined PFA-100 and functional von Willebrand factor testing for the laboratory assessment of desmopressin and factor concentrate therapy in von Willebrand disease
    Favaloro, Emmanuel J.
    Thom, Jim
    Patterson, David
    Just, Sarah
    Baccala, Maria
    Dixon, Tracy
    Meiring, Muriel
    Koutts, Jerry
    Rowell, John
    Baker, Ross
    BLOOD COAGULATION & FIBRINOLYSIS, 2009, 20 (06) : 475 - 483
  • [6] Use of a novel platelet function analyzer (PFA-100™) with high sensitivity to disturbances in von willebrand factor to screen for von Willebrand's disease and other disorders
    Favaloro, EJ
    Facey, D
    Henniker, A
    AMERICAN JOURNAL OF HEMATOLOGY, 1999, 62 (03) : 165 - 174
  • [7] Comparison at the O'Brien filter test and the PFA-100 platelet analyzer in the laboratory diagnosis of von Willebrand's disease
    Schlammadinger, A
    Kerenyi, A
    Muszbek, L
    Boda, Z
    THROMBOSIS AND HAEMOSTASIS, 2000, 84 (01) : 88 - 92
  • [8] The role of the platelet function analyser (PFA-100™) in the characterization of patients with von Willebrand's disease and its relationships with von Willebrand factor and the ABO blood group
    Nitu-Whalley, IC
    Lee, CA
    Brown, SA
    Riddell, A
    Hermans, C
    HAEMOPHILIA, 2003, 9 (03) : 298 - 302
  • [9] Hemostasis Laboratory: Assays for Platelet Function and Von Willebrand Disease
    Anupama Kohli
    Fauzia Khan
    L. Michael Snyder
    L. Pechet
    Journal of Thrombosis and Thrombolysis, 1998, 6 : 159 - 167
  • [10] High Resolution Multimer Analysis and the PFA-100 Platelet Function Analyser Can Detect von Willebrand Disease Type 2A without a Pathological Ratio of Ristocetin Cofactor Activity and von Willebrand Antigen Level
    Weiss, Dominik R.
    Strasser, Erwin F.
    Ringwald, Juergen
    Zimmermann, Robert
    Eckstein, Reinhold
    CLINICAL LABORATORY, 2012, 58 (11-12) : 1203 - 1209