Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0

被引:15
|
作者
Hatzakis, Angelos. [1 ,2 ,4 ]
Papachristou, Helen [1 ]
Nair, Sangeetha J. [3 ]
Fortunko, Jacqueline [3 ]
Foote, Tracy [3 ]
Kim, HeeCheol [3 ]
Peling, Tashi L. [3 ]
Worlock, Andrew J. [3 ]
机构
[1] Univ Athens, Sch Med, Dept Hyg Epidemiol & Med Stat, Mikras Asias 75, GR-11527 Athens, Greece
[2] Hellen Sci Soc, Study AIDS & Sexually Transmitted Dis, Athens, Greece
[3] Hologic Inc, 10210 Genet Ctr Dr, San Diego, CA 92121 USA
[4] Univ Athens, Sch Med, Natl Retrovirus Reference Ctr, Dept Hyg Epidemiol & Med Stat, Mikras Asias 75, GR-11527 Athens, Greece
来源
VIROLOGY JOURNAL | 2016年 / 13卷
关键词
IMMUNODEFICIENCY-VIRUS TYPE-1; PLASMA VIRAL LOAD; REALTIME HIV-1; RNA LEVELS; PERFORMANCE; INFECTION; QUANTIFICATION; SUBTYPE; TIME; QUANTITATION;
D O I
10.1186/s12985-016-0627-y
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Quantitation of HIV-RNA is critically important for diagnosis, prognosis, treatment, monitoring and assessment of infectivity in HIV-1 infection. The objective of this study was to assess performance characteristics of the Aptima HIV-1 Quant Dx assay (Aptima), a new transcription mediated amplification (TMA), fully integrated and automated assay from Hologic Inc., San Diego, CA, USA. The analytical sensitivity, analytical specificity, precision and detection of HIV-1 subtypes were tested based on commercially available international standards or panels. A selected group of 244 anti-HIV-1 (+) plasma samples was used for comparison with Roche COBAS Ampliprep/COBAS TaqMan HIV-1 test v2.0 (Roche CAP/CTM), (Roche Molecular Systems, Pleasanton, CA). Results: The 50 and 95 % limit of detection were estimated at 4.9 (95 % CI 3.9-5.7) and 17.6 (15.2-21.2) IU/mL respectively. The specificity was found 99.83 (99.06-99.97) %. The standard deviations and coefficient of variations for panels with 50 and 100 copies/mL (1.7 and 2 log copies/mL) were 0.14 log copies/mL (8.67 % CV) and 0.18 log copies/mL (9.91 % CV) respectively. The detection rate for Aptima and Roche assays was 220/244 (90.2 %) and 217/244 (88.9 %) respectively. Conclusion: The Aptima assay is a sensitive, specific, precise and accurate test for measuring HIV-1 viral loads and for the detection of HIV-1 infections.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0
    Angelos. Hatzakis
    Helen Papachristou
    Sangeetha J. Nair
    Jacqueline Fortunko
    Tracy Foote
    HeeCheol Kim
    Tashi L. Peling
    Andrew J. Worlock
    Virology Journal, 13
  • [2] COMPARISON OF THE APTIMA HIV-1 QUANT ASSAY TO THE COBAS AMPLIPREP/COBAS TAQMAN HIV-1 TEST, V2.0
    Hatzakis, A.
    Papachristou, H.
    Kantzanou, M.
    SEXUALLY TRANSMITTED INFECTIONS, 2013, 89 : A353 - A353
  • [3] Comparison of the Hologic Aptima HIV-1 Quant Dx Assay to the Roche COBAS Ampliprep/COBAS TaqMan HIV-1 Test v2.0 for the quantification of HIV-1 RNA in plasma samples
    Schonning, Kristian
    Johansen, Kim
    Landt, Bodil
    Benfield, Thomas
    Westh, Henrik
    JOURNAL OF CLINICAL VIROLOGY, 2017, 92 : 14 - 19
  • [4] Comparison of the Hologic Aptima HIV-1 Quant Dx assay to the Roche COBAS Ampliprep/COBAS TaqMan HIV-1 Test v2.0 for the quantification of HIV-RNA in plasma samples
    Schonning, K.
    Johansen, K.
    Landt, B.
    Benfield, T.
    Westh, H.
    JOURNAL OF CLINICAL VIROLOGY, 2016, 82 : S53 - S53
  • [5] Comparison of the Aptima HIV-1 Quant Dx assay with the COBAS AmpliPrep/COBAS TaqMan HIV-1 v2.0 Test for HIV-1 viral load quantification in plasma samples from HIV-1-infected patients
    Longo, Serena
    Bon, Isabella
    Musumeci, Giuseppina
    Bertoldi, Alessia
    D'Urbano, Vanessa
    Calza, Leonardo
    Re, Maria Carla
    HEALTH SCIENCE REPORTS, 2018, 1 (04)
  • [6] Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma
    Schalasta, Gunnar
    Boerner, Anna
    Speicher, Andrea
    Enders, Martin
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (03) : 493 - 499
  • [7] Performance evaluation of the Aptima® HIV-1 Quant Dx and Aptima® HBV Quant assays on the fully automated Panther in comparison to COBAS® AmpliPrep/COBAS® TaqMan® HIV-1 and HBV tests
    Ebel, A.
    Bali, C.
    Guis, L.
    Merlin, S.
    JOURNAL OF CLINICAL VIROLOGY, 2016, 82 : S35 - S36
  • [8] Cobas Ampliprep/Cobas TaqMan HIV-1 v2.0 Assay: Consequences at the Cohort Level
    Taylor, Ninon
    Grabmeier-Pfistershammer, Katharina
    Egle, Alexander
    Greil, Richard
    Rieger, Armin
    Ledergerber, Bruno
    Oberkofler, Hannes
    PLOS ONE, 2013, 8 (08):
  • [9] Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime HCV assay and Roche COBAS AmpliPrep/COBAS TaqMan HCV quantitative test v2.0
    Worlock, A.
    Blair, D.
    Hunsicker, M.
    Le-Nguyen, T.
    Motta, C.
    Nguyen, C.
    Papachristou, E.
    Pham, J.
    Williams, A.
    Vi, M.
    Vinluan, B.
    Hatzakis, A.
    VIROLOGY JOURNAL, 2017, 14
  • [10] Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime HCV assay and Roche COBAS AmpliPrep/COBAS TaqMan HCV quantitative test v2.0
    A. Worlock
    D. Blair
    M. Hunsicker
    T. Le-Nguyen
    C. Motta
    C. Nguyen
    E. Papachristou
    J. Pham
    A. Williams
    M. Vi
    B. Vinluan
    A. Hatzakis
    Virology Journal, 14