Traceability validation of six enzyme measurements on the Abbott Alinity c analytical system

被引:15
作者
Aloisio, Elena [1 ]
Frusciante, Erika [1 ]
Pasqualetti, Sara [1 ]
Infusino, Ilenia [1 ]
Krintus, Magdalena [2 ]
Sypniewska, Grazyna [2 ]
Panteghini, Mauro [1 ]
机构
[1] Univ Milan, Res Ctr Metrol Traceabil Lab Med CIRME, Via GB Grassi 74, I-20157 Milan, Italy
[2] Nicolaus Copernicus Univ, Coll Med, Dept Lab Med, Torun, Poland
关键词
enzymes; standardization; traceability; trueness; uncertainty; CATALYTIC ACTIVITY CONCENTRATIONS; METROLOGICAL TRACEABILITY; BIOLOGICAL VARIATION; REFERENCE INTERVALS; ALKALINE-PHOSPHATASE; UNCERTAINTY BUDGET; SERUM; 37-DEGREES-C; ALT; RESPONSIBILITIES;
D O I
10.1515/cclm-2020-0015
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Laboratory professionals should independently verify the correct implementation of metrological traceability of commercial measuring systems and determine if their performance is fit for purpose. We evaluated the trueness, uncertainty of measurements, and transferability of six clinically important enzyme measurements (alanine aminotransferase [ALT], alkaline phosphatase [ALP], aspartate aminotransferase [AST], creatine kinase [CK], gamma-glutamyltransferase [gamma GT], and lactate dehydrogenase [LDH]) performed on the Abbott Alinity c analytical system. Methods: Target values and associated uncertainties were assigned to three pools for each enzyme by using the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference measurement procedures (RMPs) and the pools were then measured on the Alinity system. Bias estimation and regression studies were performed, and the uncertainty associated with Alinity measurements was also estimated, using analytical performance specifications (APS) derived from biological variability of measurands as goals. Finally, to validate the transferability of the obtained results, a comparison study between two Alinity systems located in Milan, Italy, and Bydgoszcz, Poland, was carried out. Results: Correct implementation of traceability to the IFCC RMPs and acceptable measurement uncertainty fulfilling desirable (ALP, AST, LDH) or optimal APS (ALT, CK, gamma GT) was verified for all evaluated enzymes. An optimal alignment between the two Alinity systems located in Milan and Bydgoszcz was also found for all enzyme measurements. Conclusions: We confirmed that measurements of ALT, ALP, AST, CK, gamma GT, and LDH performed on the Alinity c analytical system are correctly standardized to the IFCC reference measurement systems and the system alignment is consistent between different platforms.
引用
收藏
页码:1250 / 1256
页数:7
相关论文
共 40 条
[1]   Traceability of alkaline phosphatase measurement may also vary considerably using the same analytical system: the case of Abbott Architect [J].
Aloisio, Elena ;
Frusciante, Erika ;
Pasqualetti, Sara ;
Quercioli, Massimo ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (06) :E135-E137
[2]  
[Anonymous], 2019, Medical laboratories -practical guidance for the estimation of measurement uncertainty, V1st
[3]  
Bais R, 2018, TIETZ TXB CLIN CHEM, P328
[4]   Defining acceptable limits for the metrological traceability of specific measurands [J].
Bais, Renze ;
Armbruster, Dave ;
Jansen, Rob T. P. ;
Klee, George ;
Panteghini, Mauro ;
Passarelli, Joseph ;
Sikaris, Ken A. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (05) :973-979
[5]   Defining permissible limits for the combined uncertainty budget in the implementation of metrological traceability [J].
Braga, Federica ;
Panteghini, Mauro .
CLINICAL BIOCHEMISTRY, 2018, 57 :7-11
[6]   Evaluation of the trueness of serum alkaline phosphatase measurement in a group of Italian laboratories [J].
Braga, Federica ;
Frusciante, Erika ;
Infusino, Ilenia ;
Aloisio, Elena ;
Guerra, Elena ;
Ceriotti, Ferruccio ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (03) :E47-E50
[7]   ROLE AND RESPONSIBILITIES OF LABORATORY MEDICINE SPECIALISTS IN THE VERIFICATION OF METROLOGICAL TRACEABILITY OF IN VITRO MEDICAL DIAGNOSTICS [J].
Braga, Federica ;
Infusino, Ilenia ;
Panteghini, Mauro .
JOURNAL OF MEDICAL BIOCHEMISTRY, 2015, 34 (03) :282-287
[8]   Performance criteria for combined uncertainty budget in the implementation of metrological traceability [J].
Braga, Federica ;
Infusino, Ilenia ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (06) :905-912
[9]   Verification of in vitro medical diagnostics (IVD) metrological traceability: Responsibilities and strategies [J].
Braga, Federica ;
Panteghini, Mauro .
CLINICA CHIMICA ACTA, 2014, 432 :55-61
[10]   Biological Variation Estimates Obtained from 91 Healthy Study Participants for 9 Enzymes in Serum [J].
Carobene, Anna ;
Roraas, Thomas ;
Solvik, Una Orvim ;
Sylte, Marit Sverresdotter ;
Sandberg, Sverre ;
Guerra, Elena ;
Marino, Irene ;
Jonker, Niels ;
Barla, Gerhard ;
Bartlett, William A. ;
Fernandez-Calle, Pilar ;
Diaz-Garzon, Jorge ;
Tosato, Francesca ;
Plebani, Mario ;
Coskun, Abdurrahman ;
Serteser, Mustafa ;
Unsal, Ibrahim ;
Ceriottil, Ferruccio .
CLINICAL CHEMISTRY, 2017, 63 (06) :1141-1150